Cyclic Voltammetry: Understanding Redox Peaks and Reversibility
Learn how cyclic voltammetry works by sweeping electrode potential and measuring current. Explore how redox peaks reveal reaction reversibility, diffusion, scan-rate effects, and chemical reactions following electron transfer. An accessible extension beyond the Year 12 Chemistry syllabus.
A solution can contain a redox-active substance without showing an obvious colour change. You might know from its redox half-equation that it can gain or lose electrons, but that doesn’t tell you exactly when the reaction occurs at an electrode, how quickly it proceeds, or whether the product can be converted back into the original substance.
There is a way to investigate all three. Instead of holding an electrode at one potential, you continuously change its potential, reverse the direction, and measure the resulting current. The resulting graph can reveal surprisingly detailed information about the redox behaviour of a substance.
This technique is called cyclic voltammetry (CV).
A note about the syllabus: Cyclic voltammetry is an extension beyond the required Year 12 Chemistry content. It is not an HSC assessment requirement. The 2025 Chemistry 11-12 syllabus identifies Applying chemical ideas as a Year 12 focus area, which provides a useful starting point through chemical analysis and identification. NESA schedules implementation of the new syllabus from 2028, with the first HSC examination in 2029. The syllabus reference here is a learning entry point, not an assessment requirement for cyclic voltammetry. NESA syllabus information.
01The redox chemistry behind cyclic voltammetry
Before looking at the instrument, recall what happens during a redox reaction.
Oxidation involves the loss of electrons, while reduction involves the gain of electrons.
Consider a simple one-electron redox couple:
\[ \ce{Ox + e^- <=> Red} \]
Here:
- \(\ce{Ox}\) represents the oxidised form of a substance.
- \(\ce{Red}\) represents its reduced form.
- \(e^-\) represents an electron.
The forward reaction is reduction: \(\ce{Ox}\) gains an electron to become \(\ce{Red}\).
The reverse reaction is oxidation: \(\ce{Red}\) loses an electron to become \(\ce{Ox}\).
In a conventional electrochemical experiment, these reactions can occur at an electrode surface.
The important idea is that the electrode potential influences which direction is favoured.
- Making the working electrode potential more negative encourages reduction.
- Making it more positive encourages oxidation.
Think of the electrode potential as controlling how favourable electron transfer is at the electrode surface.
This is a simplified model. Actual reactions also depend on the concentrations of the oxidised and reduced species, the electrode material, the solution composition, and the reaction kinetics.
Why measure current?
When electrons transfer between an electrode and a dissolved chemical species, an electrical current flows through the external circuit.
Current measures the rate of charge transfer:
\[ I=\frac{\Delta Q}{\Delta t} \]
where:
- \(I\) is current, measured in amperes (A).
- \(\Delta Q\) is the change in electrical charge, measured in coulombs (C).
- \(\Delta t\) is the time interval, measured in seconds (s).
A larger current generally means that charge is being transferred more rapidly.
This gives us two useful measurements:
- Electrode potential tells us how strongly the conditions favour oxidation or reduction.
- Current tells us how rapidly electron transfer is occurring.
Cyclic voltammetry combines these measurements to investigate a substance’s electrochemical behaviour.
02How the experiment works
A typical cyclic voltammetry experiment uses an electrochemical cell containing a solution of the substance being investigated and a supporting electrolyte.
The cell usually contains three electrodes.
Working electrode
This is where the redox reaction being investigated occurs.
It is commonly made from an electrically conductive material such as platinum, gold, or glassy carbon.
Reference electrode
This provides a stable reference potential against which the working electrode potential is measured.
It allows the instrument to control the working electrode potential accurately.
Counter electrode
This completes the electrical circuit and carries the current required to balance the reaction at the working electrode.
A device called a potentiostat controls the working electrode potential relative to the reference electrode and measures the resulting current between the working and counter electrodes.
The supporting electrolyte helps conduct electricity through the solution and reduces unwanted effects from the movement of analyte ions under the electric field. It is selected to avoid reacting with the substance being investigated within the potential range used.
Sweeping the electrode potential
Here is the defining feature of cyclic voltammetry.
Rather than maintaining a constant electrode potential, the potentiostat changes the working electrode potential continuously.
Suppose we begin at a potential where no significant reaction occurs.
The instrument then:
- Changes the potential steadily in one direction.
- Reaches a predetermined limit, called the switching potential.
- Reverses the direction of the potential sweep.
- Returns towards the starting potential.
This produces a triangular potential-versus-time pattern.
The speed at which the potential changes is called the scan rate.
It is represented by \(\nu\) (the Greek letter nu).
\[ \nu=\frac{|\Delta E|}{\Delta t} \]
Here:
- \(\nu\) is the scan rate, measured in volts per second (V s\(^{-1}\)).
- \(\Delta E\) is the change in electrode potential, measured in volts (V).
- \(\Delta t\) is the time taken, measured in seconds (s).
For example, a scan rate of \(\pu{0.10 V s-1}\) means the electrode potential changes by 0.10 V every second during each straight section of the sweep.
The potential is then reversed at the switching potential.
The experiment records the current continuously as the potential changes. A graph of current against electrode potential is called a cyclic voltammogram.
IUPAC defines cyclic voltammetry as recording current while the electrode potential is varied cyclically between two potential limits, normally at a constant scan rate. IUPAC Gold Book.
03Understanding a cyclic voltammogram
Let’s work through what happens during the experiment.
Assume the solution initially contains the oxidised species \(\ce{Ox}\), and the working electrode starts at a potential where little redox activity occurs.
We gradually make the working electrode potential more negative.
1. The reduction peak
At sufficiently negative potentials, reduction becomes favourable:
\[ \ce{Ox + e^- -> Red} \]
The oxidised species gains electrons at the working electrode.
As the reduction reaction becomes more favourable, the magnitude of the reduction current increases.
Eventually, the current reaches a maximum in the reduction direction.
This is called the cathodic peak.
Its peak potential is represented by \(E_{pc}\), and its peak current is represented by \(i_{pc}\).
Under the conventional sign convention used in this article, cathodic current is negative.
But why does the current decrease after reaching its peak?
You might predict that making the potential increasingly negative should continuously increase the reduction current.
That prediction seems reasonable because the reaction is becoming more favourable.
However, it overlooks the supply of reactant.
As \(\ce{Ox}\) is reduced, its concentration near the electrode surface decreases. More \(\ce{Ox}\) must travel from the bulk solution towards the electrode to replace it.
This movement is called diffusion.
As the reaction continues, the region affected by the reaction spreads further into the solution. The supply of fresh \(\ce{Ox}\) becomes less able to keep up with electron transfer.
Consequently, the magnitude of the current decreases, even though the potential continues becoming more negative.
The peak therefore reflects both the electrode reaction and the rate at which reactant reaches the electrode.
2. Reversing the potential
Once the switching potential is reached, the instrument reverses the potential sweep.
Instead of continuing towards more negative potentials, it gradually becomes more positive.
Remember that the forward reduction reaction has produced \(\ce{Red}\) near the electrode.
What happens when the potential becomes more positive?
The reduced species can now lose electrons:
\[ \ce{Red -> Ox + e^-} \]
This is oxidation.
The oxidation current increases until it reaches a maximum. This maximum is called the anodic peak.
Its peak potential is represented by \(E_{pa}\), and its peak current is represented by \(i_{pa}\).
Under our conventional sign convention, anodic current is positive.
As the scan continues, the concentration of \(\ce{Red}\) near the electrode decreases because it is being oxidised. The anodic current eventually falls again.
The result is a second peak on the reverse sweep.
3. Why are there two peaks?
The two peaks represent opposite reactions of the same redox couple.
| Feature | Reaction | Current direction |
|---|---|---|
| Cathodic peak | Reduction: \(\ce{Ox + e^- -> Red}\) | Negative |
| Anodic peak | Oxidation: \(\ce{Red -> Ox + e^-}\) | Positive |
The first peak shows the reduction of the original substance.
The second peak shows the oxidation of the product generated during the first sweep.
This is an important distinction: the reverse peak does not necessarily represent a second substance that was originally present in the solution.
It can represent the same redox-active substance returning to its original oxidation state.
The appearance of both peaks provides evidence that the redox process can operate in both directions under the experimental conditions.
However, the two peaks will not always be equally prominent. We need to understand what their positions and magnitudes tell us.
04What can we learn from the peaks?
Cyclic voltammetry provides several useful pieces of information.
1. The formal potential
For a simple, reversible redox couple, the average of the anodic and cathodic peak potentials provides an estimate of its formal potential.
It is represented by \(E^{0'}\).
\[ E^{0'}\approx\frac{E_{pa}+E_{pc}}{2} \]
The formal potential describes the characteristic potential of a redox couple under the particular solution conditions.
It is useful for comparing redox behaviour and assessing how readily a substance undergoes oxidation or reduction.
Importantly, it is not automatically identical to a tabulated standard reduction potential, \(E^\circ\). The solvent, electrolyte, pH, complex formation, and reference electrode can all affect the measured value.
2. Peak separation
The difference between the two peak potentials is called the peak-to-peak separation.
It is represented by \(\Delta E_p\).
\[ \Delta E_p=E_{pa}-E_{pc} \]
For an ideal, reversible, diffusion-controlled, one-electron redox couple at approximately \(25^\circ\mathrm{C}\):
\[ \Delta E_p\approx57\ \mathrm{mV} \]
This value is a useful reference when interpreting experimental data.
A peak separation close to the theoretical value suggests that electron transfer at the electrode is sufficiently rapid to maintain an approximately reversible response.
A much larger separation can indicate slower electron-transfer kinetics.
However, it is not proof by itself. Resistance within the solution, for example, can also increase the apparent peak separation.
3. Comparing the peak currents
For a simple reversible redox couple, the anodic and cathodic peak currents should have similar magnitudes.
In other words:
\[ i_{pa}\approx |i_{pc}| \]
This makes physical sense.
If the reduced species produced during the forward scan remains stable, it should be available for oxidation during the reverse scan.
If much less material is available for the reverse reaction, the anodic peak may be smaller.
This could indicate that the reduced species has undergone another chemical reaction before it can be oxidised.
It is important to distinguish two different ideas:
- Electrochemical reversibility: How rapidly electrons transfer between the electrode and the chemical species.
- Chemical reversibility: Whether the product of the electrode reaction remains sufficiently stable to undergo the reverse chemical reaction.
A system can have fast electron transfer but poor chemical reversibility.
That distinction becomes particularly useful when investigating more complicated redox reactions.
4. Background current
Not all measured current comes from the oxidation or reduction of the analyte.
Changing the electrode potential also rearranges charged particles near the electrode surface. This produces a small charging current, even when no analyte is undergoing a redox reaction.
This is called capacitive current, or non-Faradaic current.
The current arising from actual electron-transfer reactions is called Faradaic current.
A background measurement using the supporting electrolyte without the analyte helps identify this contribution.
Without accounting for background current, a student or researcher might mistake an electrical charging effect for a chemical reaction.
This becomes particularly important when comparing small peaks or experiments conducted at different scan rates. Journal of Chemical Education: A Practical Beginner’s Guide to Cyclic Voltammetry.
A useful interpretation guide
| Observation | Possible interpretation |
|---|---|
| Two peaks with similar magnitudes | Both oxidation and reduction products remain available |
| Peak separation close to the ideal value | Fast electron transfer may be occurring |
| Large peak separation | Slower electron-transfer kinetics or experimental resistance |
| Small or missing reverse peak | Product instability, insufficient potential excursion, or other limitations |
| Peak current increases with concentration | More electroactive material is available |
| Peak current increases with scan rate | Faster potential sweeps are producing a larger current response |
These are diagnostic clues, not independent proofs. The experimental conditions must be considered before drawing a conclusion.
05How scan rate changes the voltammogram
So far, we have considered what happens when the potential changes at a particular speed.
But what if we repeat the experiment at a faster scan rate?
Imagine sweeping the potential across the same range in one second rather than four seconds.
The electrode potential changes more rapidly, leaving less time for the diffusion layer to spread during the sweep.
For a freely diffusing redox species undergoing an ideal reversible electrode reaction, the peak current increases in proportion to the square root of the scan rate:
\[ i_p\propto\sqrt{\nu} \]
Here:
- \(i_p\) is the peak current magnitude.
- \(\nu\) is the scan rate in V s\(^{-1}\).
This is the qualitative relationship described by the Randles-Ševčík model.
It applies under particular assumptions, including diffusion-controlled transport, a suitable planar electrode, and sufficiently rapid electron transfer. It is not a universal relationship for every cyclic voltammogram. IUPAC: Randles-Ševčík equations.
There is another useful comparison.
For an ideal redox species confined to the electrode surface, peak current is instead proportional to scan rate:
\[ i_p\propto\nu \]
The distinction is useful:
- Freely diffusing species: peak current is proportional to \(\sqrt{\nu}\).
- Surface-confined species: peak current is proportional to \(\nu\).
This comparison requires care because capacitive current also depends on scan rate, and resistance or electrode structure can complicate the interpretation. Electrochemistry, Cyclic Voltammetry Part 2.
06Worked examples
Worked example: Interpreting a reversible redox couple
A solution containing an oxidised species is analysed using cyclic voltammetry at \(25^\circ\mathrm{C}\).
The following peak values are obtained:
| Measurement | Value |
|---|---|
| Cathodic peak potential, \(E_{pc}\) | \(-0.184\ \mathrm{V}\) |
| Anodic peak potential, \(E_{pa}\) | \(-0.126\ \mathrm{V}\) |
| Cathodic peak current, \(i_{pc}\) | \(-12.0\ \mathrm{\mu A}\) |
| Anodic peak current, \(i_{pa}\) | \(+11.4\ \mathrm{\mu A}\) |
All potentials are measured against the same Ag/AgCl reference electrode.
Calculate the peak separation and estimate the formal potential. Comment on the reversibility of the redox couple.
Step 1
The cathodic peak occurs at \(-0.184\ \mathrm{V}\), while the anodic peak occurs at \(-0.126\ \mathrm{V}\).
The anodic peak occurs at the more positive potential, as expected for the reverse oxidation reaction.
Step 2
\[ \begin{aligned} \Delta E_p&=E_{pa}-E_{pc}\\ &=-0.126-(-0.184)\\ &=0.058\ \mathrm{V} \end{aligned} \]
Therefore:
\[ \boxed{\Delta E_p=58\ \mathrm{mV}} \]
Step 3
Use the average of the two peak potentials:
\[ \begin{aligned} E^{0'}&\approx\frac{E_{pa}+E_{pc}}{2}\\ &=\frac{-0.126-0.184}{2}\\ &=-0.155\ \mathrm{V} \end{aligned} \]
The estimated formal potential is:
\[ \boxed{E^{0'}\approx-0.155\ \mathrm{V}} \]
This value is relative to the stated reference electrode.
Step 4
Compare the magnitudes of the two currents:
\[ \frac{i_{pa}}{|i_{pc}|} = \frac{11.4}{12.0} =0.95 \]
The ratio is close to 1, meaning the two peak currents have similar magnitudes.
Step 5
The measured peak separation is 58 mV, which is close to the ideal value of approximately 57 mV for a reversible one-electron reaction at \(25^\circ\mathrm{C}\).
The peak currents are also similar.
Together, these observations are consistent with a reversible redox couple under the experimental conditions.
The important point is that the potential values and current magnitudes provide different information. The peak separation helps assess electron-transfer behaviour, while the current ratio provides evidence about whether the products remain available for the reverse reaction.
Worked example: Comparing two scan rates
A researcher investigates a freely diffusing redox species at a constant concentration.
The background-corrected cathodic peak currents are measured at three scan rates.
The following values are illustrative data:
| Scan rate, \(\nu\) (mV s\(^{-1}\)) | Peak current magnitude (\(\mathrm{\mu A}\)) |
|---|---|
| 25 | 8.0 |
| 100 | 16.0 |
| 225 | 24.0 |
Determine whether the results are consistent with diffusion-controlled behaviour.
Step 1
For an ideal reversible redox species undergoing diffusion-controlled electron transfer:
\[ i_p\propto\sqrt{\nu} \]
This means that multiplying the scan rate by four should multiply the peak current by two.
Step 2
The scan rate increases from 25 to 100 mV s\(^{-1}\).
\[ \frac{100}{25}=4 \]
The expected current ratio is:
\[ \sqrt{4}=2 \]
The measured current ratio is:
\[ \frac{16.0}{8.0}=2 \]
The measurements agree with the predicted relationship.
Step 3
The scan rate increases from 25 to 225 mV s\(^{-1}\).
\[ \frac{225}{25}=9 \]
The expected current ratio is:
\[ \sqrt{9}=3 \]
The measured current ratio is:
\[ \frac{24.0}{8.0}=3 \]
Again, the measurements agree.
Step 4
The peak current increases in proportion to the square root of the scan rate.
The results are therefore consistent with diffusion-controlled behaviour for a freely dissolved redox species.
A faster potential sweep produces a steeper concentration gradient near the electrode, allowing a larger current to flow.
This is a useful diagnostic relationship, although further experiments would be needed to establish the mechanism conclusively.
07Questions and solutions
Question 1
A solution initially contains the oxidised species \(\ce{Ox}\).
The working electrode potential is initially positive and is gradually swept towards more negative values. The scan is then reversed towards more positive values.
The redox reaction is:
\[ \ce{Ox + e^- <=> Red} \]
Using the conventional current sign convention, a negative current peak appears during the first sweep, followed by a positive current peak during the return sweep.
Identify the reaction occurring at each peak and explain why the second peak appears.
Solution 1
The negative peak represents reduction, while the positive peak represents oxidation.
During the negative-going sweep, the working electrode supplies electrons to \(\ce{Ox}\):
\[ \ce{Ox + e^- -> Red} \]
This produces \(\ce{Red}\) near the electrode surface and generates a cathodic current.
When the potential sweep reverses and becomes more positive, \(\ce{Red}\) can lose electrons:
\[ \ce{Red -> Ox + e^-} \]
This generates an anodic current.
The second peak therefore occurs because the first reaction has produced a substance that can undergo the reverse reaction.
The two peaks represent opposite directions of the same redox couple, rather than necessarily representing two different substances originally present in the solution.
Question 2
Two redox-active substances, A and B, are investigated at the same concentration using the same electrode area and solution conditions.
Their background-corrected peak currents are measured at two scan rates.
| Scan rate (mV s\(^{-1}\)) | Substance A (\(\mathrm{\mu A}\)) | Substance B (\(\mathrm{\mu A}\)) |
|---|---|---|
| 25 | 6.0 | 6.0 |
| 100 | 12.0 | 24.0 |
Assume that uncompensated resistance is negligible and that both responses are dominated by a single redox process.
(a) Determine which substance is more consistent with a freely diffusing redox species.
(b) Determine which substance is more consistent with a surface-confined redox process.
(c) Suggest an additional experiment to investigate whether substance B is adsorbed onto the electrode surface.
Solution 2
(a) Substance A is consistent with a freely diffusing redox species.
The scan rate increases by a factor of four:
\[ \frac{100}{25}=4 \]
For a diffusion-controlled process, the peak current should increase by:
\[ \sqrt{4}=2 \]
Substance A increases from 6.0 to 12.0 \(\mathrm{\mu A}\), which is exactly double.
This supports diffusion-controlled transport.
(b) Substance B is consistent with a surface-confined redox process.
Substance B increases from 6.0 to 24.0 \(\mathrm{\mu A}\).
Its peak current increases by a factor of four when the scan rate increases by a factor of four.
This is consistent with:
\[ i_p\propto\nu \]
For a surface-confined species, the electroactive material is already associated with the electrode surface. Its current therefore responds differently to changes in scan rate compared with a freely diffusing species.
(c) Rinse and transfer the electrode into a clean supporting electrolyte solution.
If substance B is adsorbed onto the electrode, some of it may remain attached after the surrounding solution is replaced.
A redox peak that persists in the clean electrolyte would provide evidence of surface-bound material.
If the peak disappears, the original response may have depended on dissolved material rather than a strongly adsorbed species.
The scan-rate relationship is a useful starting point, but it does not independently prove adsorption. Thin-layer effects and residual background currents can also complicate the interpretation.
Question 3
A student investigates a redox-active substance using two different potential ranges.
In the first experiment, the potential is swept from \(+0.30\ \mathrm{V}\) to \(-0.10\ \mathrm{V}\), then reversed. The voltammogram shows only a small cathodic shoulder and no distinct anodic peak.
In the second experiment, the switching potential is changed to \(-0.35\ \mathrm{V}\). All other experimental conditions remain unchanged.
This time, the voltammogram shows a clear cathodic peak followed by a clear anodic peak.
The student initially claims that the substance is chemically irreversible because the first experiment did not produce an oxidation peak.
Explain why this conclusion is unjustified.
Solution 3
The first experiment most likely failed to extend the potential far enough into the reduction region to produce a substantial amount of the reduced species.
The original substance must undergo reduction before enough \(\ce{Red}\) can be generated near the electrode for a clear oxidation peak to appear on the return sweep.
In the first experiment, the scan reverses at \(-0.10\ \mathrm{V}\). It does not reach sufficiently negative potentials to complete the reduction response.
In the second experiment, the scan extends to \(-0.35\ \mathrm{V}\). This allows substantially more reduction to occur, producing enough \(\ce{Red}\) for a measurable anodic peak during the reverse scan.
The tempting interpretation is that the missing reverse peak demonstrates an irreversible reaction.
However, the crucial experimental condition is the switching potential.
A missing oxidation peak does not automatically demonstrate chemical irreversibility. It can also occur because the initial scan did not generate enough reduced product.
To investigate chemical irreversibility, the potential range must first be sufficient to produce the relevant redox product.
Question 4
A one-electron redox couple is investigated at \(25^\circ\mathrm{C}\).
The same reference electrode, supporting electrolyte, solution composition, and working electrode are used throughout. Background current is corrected, and a separate resistance check confirms that uncompensated resistance is negligible.
The following measurements are obtained:
| Scan rate | \(E_{pc}\) | \(E_{pa}\) | \(i_{pa}/|i_{pc}|\) |
|---|---|---|---|
| 25 mV s\(^{-1}\) | \(-0.210\ \mathrm{V}\) | \(-0.153\ \mathrm{V}\) | 0.98 |
| 400 mV s\(^{-1}\) | \(-0.270\ \mathrm{V}\) | \(-0.080\ \mathrm{V}\) | 1.02 |
The student suggests that the reduced species becomes chemically unstable at higher scan rates.
Evaluate this explanation and identify the more likely change in the electrochemical system.
Solution 4
The results are more consistent with increasingly limited electron-transfer kinetics at the faster scan rate, rather than chemical instability of the reduced species.
First, calculate the peak separation at each scan rate.
At 25 mV s\(^{-1}\):
\[ \begin{aligned} \Delta E_p &=E_{pa}-E_{pc}\\ &=-0.153-(-0.210)\\ &=0.057\ \mathrm{V} \end{aligned} \]
Therefore:
\[ \Delta E_p=57\ \mathrm{mV} \]
At 400 mV s\(^{-1}\):
\[ \begin{aligned} \Delta E_p &=-0.080-(-0.270)\\ &=0.190\ \mathrm{V} \end{aligned} \]
Therefore:
\[ \Delta E_p=190\ \mathrm{mV} \]
The peak separation has increased substantially.
At the slower scan rate, the separation is close to the ideal value for a reversible one-electron couple at \(25^\circ\mathrm{C}\).
At the faster scan rate, the electrode reaction cannot maintain the same near-equilibrium response as the potential changes.
Greater driving force is required to achieve electron transfer in each direction. The peaks consequently move further apart.
The peak-current ratios provide another important clue.
At both scan rates, the anodic and cathodic peak currents have similar magnitudes.
This suggests that the reduced species remains available for oxidation during the reverse sweep.
If substantial chemical decomposition were consuming the reduced species, we would generally expect the return oxidation peak to become relatively smaller.
The most likely explanation is therefore quasi-reversible electron transfer: electron transfer is sufficiently rapid at the slower scan rate but becomes kinetically limited as the potential is swept more quickly.
The key distinction is that the electron-transfer reaction is becoming too slow to keep up with the changing potential. This does not necessarily mean that the chemical product is unstable.
Question 5
A researcher investigates a redox-active substance using cyclic voltammetry.
The simplified electrode reaction is:
\[ \ce{Ox + e^- <=> Red} \]
Two experiments are conducted using the same solution, electrode, potential limits, and reference electrode.
The only changed experimental variable is the scan rate.
The following background-corrected peak currents are obtained:
| Measurement | Slow scan | Fast scan |
|---|---|---|
| Scan rate | 20 mV s\(^{-1}\) | 200 mV s\(^{-1}\) |
| Cathodic peak current | \(-10\ \mathrm{\mu A}\) | \(-32\ \mathrm{\mu A}\) |
| Anodic peak current | \(+2\ \mathrm{\mu A}\) | \(+25\ \mathrm{\mu A}\) |
| Peak separation | 58 mV | 58 mV |
Assume:
- Electron transfer at the working electrode is sufficiently fast.
- Uncompensated resistance is negligible.
- The product of any subsequent chemical reaction is electroinactive within the potential range used.
- No significant adsorption occurs.
- The same switching potential is used in both experiments.
(a) Calculate the ratio of anodic to cathodic peak-current magnitudes for each experiment.
(b) Explain why the anodic peak changes much more dramatically than would be expected from the square-root relationship alone.
(c) Propose a reaction mechanism that explains the observations.
(d) Predict what would happen to the relative size of the anodic peak if the experiment were conducted at an even slower scan rate.
Solution 5
(a) The anodic-to-cathodic peak-current ratios are 0.20 and 0.78, respectively.
For the slow scan:
\[ \frac{i_{pa}}{|i_{pc}|} = \frac{2}{10} =0.20 \]
For the fast scan:
\[ \frac{i_{pa}}{|i_{pc}|} = \frac{25}{32} \approx0.78 \]
The ratio increases considerably when the scan rate increases.
(b) The return oxidation peak is recovering because more reduced species survives at the faster scan rate.
First, consider the cathodic peak.
The scan rate increases by a factor of ten:
\[ \frac{200}{20}=10 \]
For a freely diffusing species under diffusion-controlled conditions, the expected peak-current increase is:
\[ \sqrt{10}\approx3.16 \]
The measured cathodic peak increases from 10 to 32 \(\mathrm{\mu A}\), a factor of 3.2.
This is consistent with the expected diffusion-controlled response.
However, the anodic peak increases from 2 to 25 \(\mathrm{\mu A}\), a factor of 12.5.
This is far greater than the expected square-root increase.
Something else must be affecting the amount of reduced species available during the reverse scan.
(c) The most likely mechanism is an electrochemical reaction followed by a chemical reaction, known as an EC mechanism.
The initial electrode reaction is:
\[ \ce{Ox + e^- -> Red} \]
However, the reduced species does not remain completely stable.
It undergoes a subsequent chemical reaction:
\[ \ce{Red -> P} \]
Here, \(\ce{P}\) is a product that cannot undergo another electrode reaction within the potential range being investigated.
The sequence is therefore:
- E: an electrochemical reduction generates \(\ce{Red}\).
- C: a subsequent chemical reaction consumes some of the \(\ce{Red}\).
Consider what happens during the slow experiment.
The electrode produces \(\ce{Red}\), but the longer experimental timescale gives it more opportunity to react chemically and become \(\ce{P}\).
By the time the potential reverses, relatively little \(\ce{Red}\) remains available for oxidation.
This explains the small anodic peak.
Now consider the fast experiment.
The potential reverses more quickly, giving the chemical reaction less time to consume \(\ce{Red}\).
A greater proportion of the reduced species survives and can undergo oxidation.
Consequently, the anodic peak becomes much larger relative to the cathodic peak.
(d) An even slower scan would generally produce a smaller relative anodic peak.
A slower scan gives the reduced species more time to undergo the subsequent chemical reaction.
More \(\ce{Red}\) is converted into \(\ce{P}\) before the reverse oxidation can occur.
Therefore, the ratio:
\[ \frac{i_{pa}}{|i_{pc}|} \]
would generally decrease.
The important distinction is between the speed of electron transfer and the stability of the product after electron transfer.
A small reverse peak might initially look like evidence of an irreversible electrode reaction. However, the nearly unchanged peak separation and recovery of the anodic peak at higher scan rates support a different explanation: electron transfer is fast, but the reduced product undergoes a competing chemical reaction.
This is one of the most useful applications of cyclic voltammetry. It allows researchers to investigate not only whether a substance gains or loses electrons, but also what happens to the products afterwards.
08Where to take this next
Cyclic voltammetry connects redox chemistry with analytical chemistry by turning electron transfer into a measurable electrical response.
You can use peak potentials to investigate redox behaviour, peak currents to investigate the amount of electroactive material, and changes in peak shape to investigate reaction mechanisms.
The next useful step is quantitative electroanalysis: learning how to use calibration standards and background correction to estimate the concentration of an unknown substance from its voltammetric response.
That requires one final piece of judgement: establishing whether the measured current is actually controlled by the dissolved analyte, rather than electrode adsorption, background charging, or a subsequent chemical reaction.
09Further reading
The following scientific resources provide the foundations for the additional material in this guide.
- IUPAC Gold Book – Cyclic Voltammetry: The formal definition and terminology of cyclic voltammetry.
- IUPAC Gold Book – Randles-Ševčík Equations: The relationship between peak current and scan rate.
- A Practical Beginner’s Guide to Cyclic Voltammetry – Journal of Chemical Education: An accessible scientific teaching resource covering experimental setup, voltammogram interpretation, diffusion, and scan-rate effects.
- Cyclic Voltammetry Part 1: Fundamentals – Electrochemistry: Further explanation of current responses and electrochemical mechanisms.
- Cyclic Voltammetry and Chronoamperometry: Mechanistic Tools for Organic Electrosynthesis – Chemical Society Reviews: A more advanced discussion of EC mechanisms and how chemical reactions following electron transfer influence voltammograms.