The Seebeck Effect
Learn how temperature differences between two conducting materials produce a thermoelectric voltage. Explore the physics behind the Seebeck effect, how thermocouples measure temperature, and how to solve thermoelectric voltage problems.
Touch a metal probe to a hot surface, leave its other end somewhere cooler, and a sensitive voltmeter can register a voltage even though there is no battery connected. Where does that voltage come from?
Now predict what happens if the entire probe and its connections are heated evenly. Will the voltage increase simply because everything is hotter? No. If the whole circuit is at one uniform temperature, there is no Seebeck voltage. The important ingredient is a temperature difference across the conductors.
This phenomenon is called the Seebeck effect. It is the basis of thermocouples, which convert temperature differences into measurable electrical voltages.
The Seebeck effect goes beyond the required content of the NSW Physics 11-12 Syllabus (2025). The Year 11 Electricity and magnetism focus area provides the starting knowledge, particularly the factors affecting electrical resistance. Studying this extension is not an assessment requirement. Year 11 is the prerequisite entry point for the explanation.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
01Start with electrical resistance
Before looking at thermoelectric voltage, it helps to revisit something familiar: electrical resistance.
Resistance describes how strongly a conductor opposes the flow of electric current. For an ohmic conductor, the relationship between potential difference, current, and resistance is:
\[ V=IR \]
Here:
- \(V\) is potential difference, measured in volts (V).
- \(I\) is current, measured in amperes (A).
- \(R\) is resistance, measured in ohms (\(\Omega\)).
But resistance is not determined by the applied voltage alone. It also depends on the physical properties of the conductor.
For a uniform conductor at a fixed temperature, resistance can be modelled using:
\[ R=\frac{\rho L}{A} \]
Here:
- \(\rho\) is the resistivity of the material, measured in \(\Omega\,\mathrm{m}\).
- \(L\) is the conductor’s length, measured in metres (m).
- \(A\) is its cross-sectional area, measured in square metres (\(\mathrm{m^2}\)).
This relationship helps explain four important factors.
Material
Different materials have different resistivities.
Copper allows electrical current to flow relatively easily, while materials such as nichrome offer much greater resistance.
This difference comes from the availability and behaviour of mobile charge carriers, along with the way they scatter as they move through the material.
Resistivity is a property of the material itself. Resistance is a property of the particular piece of material, including its dimensions.
Length
Imagine two wires made from the same material, with identical thickness.
The longer wire has greater resistance because charge carriers encounter more opportunities for scattering as they travel through it.
Therefore:
\[ R\propto L \]
Doubling the length doubles the resistance, assuming the material and cross-sectional area remain unchanged.
Cross-sectional area
Now compare a thin wire with a much thicker wire of the same material and length.
The thicker wire provides more available pathways for charge to flow. It can carry more current for the same applied potential difference.
Therefore:
\[ R\propto\frac{1}{A} \]
Doubling the cross-sectional area halves the resistance.
Temperature
Temperature also affects resistance, although the direction of the change depends on the material.
In many metals, increasing temperature increases resistance.
Why?
Metal atoms vibrate around their equilibrium positions. When the metal becomes hotter, these vibrations generally become stronger. Moving electrons are more likely to scatter from the vibrating atomic lattice, making electrical conduction more difficult.
Semiconductors can behave differently. Increasing temperature can release additional mobile charge carriers, allowing current to flow more easily and often decreasing resistance.
The important point is that resistivity depends on both the material and its temperature.
A useful distinction
We now have a foundation for understanding the Seebeck effect, but there is an important difference between the two phenomena.
Heating a metal uniformly can change its resistance. However, a temperature difference along a conductor can produce an electrical potential difference even when no external battery is connected.
These are different effects.
The Seebeck effect is not simply resistance increasing because a wire becomes hotter. It involves the interaction between temperature, charge carriers, and the electrical behaviour of the material.
02What happens when one end of a conductor is hotter?
Imagine a metal rod with one end maintained at a high temperature and the other end kept cool.
There is now a temperature gradient along the rod. A temperature gradient describes how temperature changes with position.
What happens to the mobile charge carriers inside the metal?
At the microscopic level, heating changes the distribution of carrier energies. The carriers that transport thermal energy also transport electrical charge.
This creates a tendency for charge carriers to diffuse along the temperature gradient. Their net movement is not perfectly balanced by movement in the opposite direction.
As charge redistributes, one region develops a slight excess of charge relative to another. This produces an internal electric field.
That electric field opposes further net charge diffusion.
Eventually, under open-circuit conditions, the electric field becomes strong enough to balance the diffusion process. There is no net electrical current, but an electric potential difference remains between the hot and cold ends.
This is the central physical idea behind the Seebeck effect.
A useful analogy: a crowded room
Imagine a crowded room connected to a much less crowded hallway.
People tend to move from the crowded room towards the available space. As they move, the distribution of people changes.
Eventually, the resulting crowding and movement constraints can balance the tendency for further movement.
Charge carriers behave differently from people, of course. Their behaviour is governed by the electronic structure of the material and the way they transport energy.
However, the analogy helps picture the central idea: movement produces a redistribution, and that redistribution creates an opposing influence.
In a conductor, that opposing influence is an electric field.
Why does the voltage remain if there is no current?
This is an important distinction.
An open circuit is a circuit in which there is no complete conducting path for appreciable current to flow.
A voltmeter has a very high input resistance, so it measures the potential difference while drawing very little current.
The temperature gradient creates the thermoelectric potential difference, and the charge redistribution establishes an equilibrium in which the net current is zero.
Therefore:
- A temperature gradient can produce a voltage.
- A voltage does not necessarily mean that current is flowing.
- Connecting an external load allows current to flow and electrical energy to be transferred.
The voltage generated by the temperature difference is called a thermoelectric voltage or Seebeck voltage.
03Why do thermocouples use two different materials?
A single conductor can develop a thermoelectric potential difference between its hot and cold ends.
However, measuring this voltage introduces another conductor: the wires connecting the measuring instrument.
That second conductor matters.
Different materials respond differently to the same temperature gradient. Their thermoelectric responses are described by a property called the Seebeck coefficient.
A practical thermocouple uses two different conducting materials, usually labelled A and B.
The two conductors are joined at one end to form the measuring junction. Their other ends connect to a measuring instrument in a reference-temperature region.
The measuring junction is placed where the temperature is to be measured. The reference junction provides the second temperature needed to interpret the voltage.
Why does the material difference matter?
Suppose both conductors were made from the same material.
Their Seebeck responses would be identical. The thermoelectric contributions from the two conductors would cancel around the measurement circuit, giving an ideal net voltage of zero.
Now replace one conductor with a different material.
Its response to the temperature gradient will generally be different. The two contributions no longer cancel, producing a measurable net voltage.
This is why a thermocouple requires a pair of dissimilar materials.
The resulting voltage depends on:
- The two materials used.
- The temperature of the measuring junction.
- The temperature of the reference junction.
- The temperature dependence of the materials’ thermoelectric properties.
An important detail: the useful thermoelectric voltage is not generated solely at the physical point where the two wires are welded together.
It develops along the sections of the conductors exposed to temperature gradients. The junction provides an electrical connection between the materials.
This distinction is important when understanding how thermocouples work in real measurement systems.
The Seebeck coefficient
The Seebeck coefficient describes how much thermoelectric voltage develops for a given temperature difference.
For a pair of materials A and B, we define the pair coefficient as:
\[ S_{AB}=S_A-S_B \]
Here:
- \(S_A\) is the Seebeck coefficient of material A.
- \(S_B\) is the Seebeck coefficient of material B.
- \(S_{AB}\) is the Seebeck coefficient of the material pair.
The coefficient is measured in volts per kelvin (V/K), although microvolts per kelvin (\(\mu\mathrm{V/K}\)) are commonly used.
The sign depends on the chosen voltage polarity and the order of the materials.
For this article, we use the convention that a positive \(S_{AB}\) produces a positive measured voltage when the measuring junction is hotter than the reference junction.
The general relationship is:
\[ V_{AB}= \int_{T_r}^{T_m}S_{AB}(T)\,dT \]
Here:
- \(V_{AB}\) is the open-circuit thermoelectric voltage.
- \(T_m\) is the measuring-junction temperature.
- \(T_r\) is the reference-junction temperature.
- \(S_{AB}(T)\) is the temperature-dependent Seebeck coefficient of the pair.
This equation accounts for the fact that the Seebeck coefficient can change with temperature.
For a limited temperature range where the coefficient is approximately constant, the equation simplifies to:
\[ \boxed{V_{AB}\approx S_{AB}\Delta T} \]
where:
\[ \Delta T=T_m-T_r \]
This is the main equation to use for introductory Seebeck effect calculations.
An important unit detail: a temperature difference of \(1\,^\circ\mathrm C\) is numerically equal to a temperature difference of \(1\,\mathrm K\). Therefore, you can use either unit for \(\Delta T\), provided the coefficient uses the corresponding temperature-difference unit.
What happens if the reference temperature changes?
Suppose the measuring junction remains at \(150\,^\circ\mathrm C\), but the reference junction warms from \(20\,^\circ\mathrm C\) to \(40\,^\circ\mathrm C\).
The temperature difference has decreased.
Therefore, assuming the same positive, approximately constant Seebeck coefficient, the measured voltage decreases.
This means a thermocouple cannot determine an absolute temperature from its voltage alone. The reference temperature must also be known.
Professional thermocouple instruments account for this using a process called cold-junction compensation. The instrument measures the reference-junction temperature and corrects the voltage reading accordingly.
Real thermocouples also use calibration tables because their Seebeck coefficients are not perfectly constant over large temperature ranges.
04Worked example: Calculate a thermoelectric voltage
A thermocouple consists of two different metals with a calibrated Seebeck coefficient of \(+24\,\mu\mathrm{V/K}\).
The measuring junction is maintained at \(145\,^\circ\mathrm C\), while the reference junction is at \(25\,^\circ\mathrm C\).
Calculate the open-circuit thermoelectric voltage.
Step 1
The measuring junction is hotter than the reference junction.
\[ \Delta T=T_m-T_r \]
\[ \Delta T=145-25 \]
\[ \Delta T=120\,\mathrm K \]
Step 2
The Seebeck coefficient is approximately constant over the specified range, so we use:
\[ V_{AB}=S_{AB}\Delta T \]
Step 3
Convert the coefficient into volts per kelvin:
\[ S_{AB}=24\times10^{-6}\,\mathrm{V/K} \]
Therefore:
\[ \begin{aligned} V_{AB} &=(24\times10^{-6})(120)\\ &=2.88\times10^{-3}\,\mathrm V \end{aligned} \]
Step 4
\[ \boxed{V_{AB}=+2.88\,\mathrm{mV}} \]
The thermocouple produces an open-circuit voltage of \(2.88\,\mathrm{mV}\).
Notice that the voltage is relatively small, even though the temperature difference is \(120\,\mathrm K\). This is why thermocouple instruments need to measure small voltages accurately.
05Worked example: Account for a changing Seebeck coefficient
A thermocouple has been calibrated over a temperature range of \(20\,^\circ\mathrm C\) to \(140\,^\circ\mathrm C\).
Its pair coefficient is approximately:
| Temperature interval | Seebeck coefficient |
|---|---|
| \(20\) to \(80\,^\circ\mathrm C\) | \(+20\,\mu\mathrm{V/K}\) |
| \(80\) to \(140\,^\circ\mathrm C\) | \(+25\,\mu\mathrm{V/K}\) |
The reference junction is at \(20\,^\circ\mathrm C\), and the voltmeter measures an open-circuit voltage of \(+2.20\,\mathrm{mV}\).
Determine the temperature of the measuring junction.
The calibration values are hypothetical and are supplied specifically for this example.
Step 1
\[ 2.20\,\mathrm{mV}=2200\,\mu\mathrm V \]
Step 2
The temperature difference across this first interval is:
\[ \Delta T_1=80-20=60\,\mathrm K \]
Therefore:
\[ \begin{aligned} V_1 &=S_1\Delta T_1\\ &=20\times60\\ &=1200\,\mu\mathrm V \end{aligned} \]
Step 3
The total measured voltage is \(2200\,\mu\mathrm V\).
The second interval must therefore contribute:
\[ V_2=2200-1200 \]
\[ V_2=1000\,\mu\mathrm V \]
Using the second Seebeck coefficient:
\[ V_2=S_2\Delta T_2 \]
Rearrange:
\[ \Delta T_2=\frac{V_2}{S_2} \]
Substitute:
\[ \Delta T_2=\frac{1000}{25} \]
\[ \Delta T_2=40\,\mathrm K \]
Step 4
The second interval begins at \(80\,^\circ\mathrm C\).
Therefore:
\[ T_m=80+40 \]
\[ \boxed{T_m=120\,^\circ\mathrm C} \]
The measuring junction is at \(120\,^\circ\mathrm C\).
The important idea is that the Seebeck coefficient changes with temperature, so each temperature interval contributes a different amount of voltage.
Using a single coefficient across the entire range would give an inaccurate result.
06Common misconceptions about the Seebeck effect
There are several tempting mistakes worth avoiding.
| Misconception | What actually happens |
|---|---|
| A hotter conductor automatically produces a voltage. | A temperature gradient is required. Uniform temperature produces no Seebeck voltage. |
| The voltage is generated at the welded junction. | The thermoelectric voltage develops along conductor sections exposed to temperature gradients. |
| A higher-resistance material produces a larger Seebeck voltage. | Electrical resistance and the Seebeck coefficient are different material properties. |
| A thermocouple measures absolute temperature directly. | Its voltage depends on the temperature difference between measuring and reference junctions. |
| A third metal always changes the thermocouple reading. | If its two junctions are maintained at the same temperature, their additional contributions cancel in the ideal case. |
That last point is particularly interesting.
Imagine connecting copper wires to a thermocouple to attach a voltmeter. You have introduced a third material into the circuit.
Does this automatically invalidate the measurement?
No. If the two connections between the thermocouple wires and copper leads are maintained at the same temperature, the additional thermoelectric contributions cancel.
However, if those connections are at different temperatures, the copper leads can introduce an additional thermoelectric voltage.
The temperature of the connections matters just as much as the materials themselves.
This is one reason practical thermocouple installations pay close attention to where wires are joined and where temperature gradients occur.
07Where is the Seebeck effect used?
The Seebeck effect has two particularly useful applications.
Thermocouples: Measuring temperature
Thermocouples are widely used in temperature measurement.
They are useful in environments where other temperature sensors may be difficult to use, including:
- Industrial furnaces.
- Engines and exhaust systems.
- Industrial heating equipment.
- Food-processing equipment.
- Scientific temperature measurements.
The basic process is straightforward.
A temperature difference develops between the measuring junction and the reference junction. The thermocouple produces a small voltage, and an instrument uses the calibrated voltage-temperature relationship to determine the measuring-junction temperature.
Thermoelectric generators: Producing electrical power
The Seebeck effect can also be used to generate electrical power directly from heat.
A thermoelectric generator uses specially selected semiconductor materials, commonly arranged as alternating p-type and n-type legs.
The legs are connected electrically in series but thermally in parallel between two plates.
One plate is maintained at a higher temperature than the other.
The temperature difference creates thermoelectric voltages across the semiconductor legs. Connecting multiple legs in series allows their individual voltages to add together.
An external electrical load can then receive power.
There is an important energy consideration here.
A thermoelectric generator does not create energy from nothing. It converts some of the thermal energy transferred from the hot side towards the cold side into electrical energy.
At open circuit, it can develop a voltage without delivering electrical power to a load. Connecting a load allows current to flow and electrical power to be transferred.
The Seebeck effect is therefore an example of how a temperature difference can produce an electrical effect, rather than requiring an external electrical source.
08Questions and solutions
Question 1
A thermocouple consists of two different metal wires with a constant pair Seebeck coefficient of \(+22\,\mu\mathrm{V/K}\).
The measuring junction is at \(160\,^\circ\mathrm C\), and the reference junction is at \(25\,^\circ\mathrm C\).
Calculate the open-circuit thermoelectric voltage.
Solution 1
The open-circuit thermoelectric voltage is \(+2.97\,\mathrm{mV}\).
The Seebeck effect produces a voltage proportional to the temperature difference when the pair coefficient is approximately constant.
First, calculate the temperature difference:
\[ \Delta T=160-25 \]
\[ \Delta T=135\,\mathrm K \]
Use:
\[ V_{AB}=S_{AB}\Delta T \]
Substitute:
\[ \begin{aligned} V_{AB} &=(22\times10^{-6})(135)\\ &=2970\times10^{-6}\,\mathrm V \end{aligned} \]
Therefore:
\[ \boxed{V_{AB}=+2.97\,\mathrm{mV}} \]
The positive sign follows from the specified polarity convention and the positive Seebeck coefficient.
The result demonstrates that the thermocouple responds to the temperature difference, not the absolute temperature of the measuring junction.
Question 2
A hypothetical thermocouple has the following calibrated pair Seebeck coefficients:
| Temperature interval | \(S_{AB}\) |
|---|---|
| \(20\) to \(60\,^\circ\mathrm C\) | \(+18\,\mu\mathrm{V/K}\) |
| \(60\) to \(100\,^\circ\mathrm C\) | \(-10\,\mu\mathrm{V/K}\) |
The reference junction is maintained at \(20\,^\circ\mathrm C\), and the measuring junction is at \(100\,^\circ\mathrm C\).
Calculate the total signed open-circuit voltage.
Explain why the change in sign of the Seebeck coefficient does not necessarily make the total voltage zero.
Solution 2
The total signed thermoelectric voltage is \(+320\,\mu\mathrm V\), or \(+0.320\,\mathrm{mV}\).
The key is to calculate the voltage contribution from each temperature interval separately.
The general relationship is:
\[ V_{AB}=\int_{T_r}^{T_m}S_{AB}(T)\,dT \]
Because the coefficient is constant within each specified interval, we can use:
\[ V_{AB}=S_1\Delta T_1+S_2\Delta T_2 \]
Step 1: Calculate the voltage contribution from the first interval.
The first interval is \(20\) to \(60\,^\circ\mathrm C\).
\[ \Delta T_1=60-20=40\,\mathrm K \]
Therefore:
\[ V_1=18\times40 \]
\[ V_1=+720\,\mu\mathrm V \]
Step 2: Calculate the voltage contribution from the second interval.
The second interval is \(60\) to \(100\,^\circ\mathrm C\).
\[ \Delta T_2=100-60=40\,\mathrm K \]
Therefore:
\[ V_2=-10\times40 \]
\[ V_2=-400\,\mu\mathrm V \]
Step 3: Add the signed contributions.
\[ V_{AB}=V_1+V_2 \]
\[ V_{AB}=720-400 \]
\[ \boxed{V_{AB}=+320\,\mu\mathrm V} \]
The voltage contributions partially cancel, but they do not cancel completely.
A negative Seebeck coefficient in one temperature interval does not mean the total thermoelectric voltage must be negative or zero.
The final result depends on the accumulated voltage across the entire temperature range.
This is why using the coefficient at just one temperature can give a misleading result.
Question 3
Two thermocouple devices have the same wire lengths, cross-sectional areas, and total electrical resistance.
- Device 1 uses copper for both conducting legs.
- Device 2 uses two different materials, A and B, with different Seebeck coefficients.
Both devices have measuring junctions at \(150\,^\circ\mathrm C\) and reference junctions at \(25\,^\circ\mathrm C\).
A student argues that both devices must produce the same thermoelectric voltage because their electrical resistances are equal.
Evaluate this prediction.
Your explanation must address:
- Whether the two devices produce the same voltage.
- Why the choice of materials matters.
- Where the thermoelectric voltage is generated.
- What happens if both junctions of Device 2 are brought to \(150\,^\circ\mathrm C\).
Solution 3
The two devices do not necessarily produce the same voltage. In the ideal case, Device 1 produces zero net Seebeck voltage, while Device 2 produces a non-zero voltage.
Equal electrical resistance does not imply equal Seebeck coefficients.
Step 1: Consider Device 1.
Both conducting legs are made from copper.
Their Seebeck coefficients are identical:
\[ S_A=S_B \]
Therefore:
\[ S_{AB}=S_A-S_B=0 \]
The thermoelectric contributions from the two legs cancel.
Although the circuit has a temperature difference of:
\[ \Delta T=150-25=125\,\mathrm K \]
its ideal net thermoelectric voltage is:
\[ \boxed{V_{AB}=0\,\mathrm V} \]
Step 2: Consider Device 2.
The two materials have different Seebeck coefficients.
Therefore:
\[ S_{AB}=S_A-S_B\ne0 \]
The thermoelectric contributions do not cancel, producing a measurable voltage.
Its magnitude depends on the pair coefficient and the temperature difference.
Step 3: Explain where the voltage originates.
The voltage develops along the sections of the conductors exposed to temperature gradients.
It is not generated solely at the physical junction between the two metals.
The different thermoelectric responses of materials A and B produce the net measured voltage.
Step 4: Consider what happens when both junctions reach \(150\,^\circ\mathrm C\).
Now:
\[ \Delta T=150-150=0\,\mathrm K \]
Therefore:
\[ V_{AB}=\int_{T_r}^{T_m}S_{AB}(T)\,dT=0 \]
There is no net Seebeck voltage in the ideal isothermal circuit.
The student’s original prediction fails because resistance and Seebeck coefficient describe different electrical properties.
Matching resistance does not make two materials respond identically to a temperature gradient.
Question 4
A thermocouple has a constant pair Seebeck coefficient of \(+30\,\mu\mathrm{V/K}\).
Initially:
- Measuring-junction temperature: \(200\,^\circ\mathrm C\).
- Reference-junction temperature: \(20\,^\circ\mathrm C\).
The reference junction then warms to \(50\,^\circ\mathrm C\), while the measuring junction remains at \(200\,^\circ\mathrm C\).
Answer the following:
(a) Calculate the initial thermoelectric voltage.
(b) Calculate the new thermoelectric voltage.
(c) If the measuring instrument incorrectly assumes the reference junction is still at \(20\,^\circ\mathrm C\), what temperature will it report?
(d) If both junctions had instead increased in temperature by \(30\,^\circ\mathrm C\), what would happen to the original thermoelectric voltage?
Solution 4
The initial voltage is \(+5.40\,\mathrm{mV}\), and the new voltage is \(+4.50\,\mathrm{mV}\). If the instrument incorrectly assumes the reference remains at \(20\,^\circ\mathrm C\), it reports \(170\,^\circ\mathrm C\).
The reason is that the thermocouple responds to the difference between its junction temperatures.
(a) Initial voltage
The initial temperature difference is:
\[ \Delta T_1=200-20 \]
\[ \Delta T_1=180\,\mathrm K \]
Using:
\[ V_{AB}=S_{AB}\Delta T \]
\[ V_1=30\times180 \]
\[ V_1=5400\,\mu\mathrm V \]
Therefore:
\[ \boxed{V_1=+5.40\,\mathrm{mV}} \]
(b) New voltage
The measuring junction remains at \(200\,^\circ\mathrm C\), but the reference junction is now at \(50\,^\circ\mathrm C\).
The new temperature difference is:
\[ \Delta T_2=200-50 \]
\[ \Delta T_2=150\,\mathrm K \]
Therefore:
\[ V_2=30\times150 \]
\[ V_2=4500\,\mu\mathrm V \]
Hence:
\[ \boxed{V_2=+4.50\,\mathrm{mV}} \]
The voltage decreases because the temperature difference has decreased.
(c) Incorrect temperature reading
The instrument assumes that the reference junction is still at \(20\,^\circ\mathrm C\).
It interprets the measured voltage as corresponding to a temperature difference of:
\[ \Delta T_{\mathrm{indicated}} =\frac{4500}{30} \]
\[ \Delta T_{\mathrm{indicated}}=150\,\mathrm K \]
It then adds its assumed reference temperature:
\[ T_{\mathrm{indicated}}=20+150 \]
\[ \boxed{T_{\mathrm{indicated}}=170\,^\circ\mathrm C} \]
The actual measuring-junction temperature is \(200\,^\circ\mathrm C\).
The instrument underestimates the temperature by:
\[ 200-170=30\,^\circ\mathrm C \]
This is why reference-junction compensation is essential for accurate thermocouple measurements.
(d) Both junctions increase by \(30\,^\circ\mathrm C\)
The new measuring-junction temperature would be:
\[ T_m=230\,^\circ\mathrm C \]
The new reference-junction temperature would be:
\[ T_r=50\,^\circ\mathrm C \]
The temperature difference is:
\[ \Delta T=230-50=180\,\mathrm K \]
This is unchanged from the original temperature difference.
Therefore:
\[ \boxed{V_{AB}=+5.40\,\mathrm{mV}} \]
The voltage remains unchanged under the constant-coefficient assumption.
This demonstrates that a thermocouple measures a temperature difference, not an absolute temperature.
Question 5
An A-B thermocouple has a constant pair Seebeck coefficient of:
\[ S_{AB}=+32\,\mu\mathrm{V/K} \]
Its measuring junction is maintained at \(180\,^\circ\mathrm C\), and its reference junction is at \(20\,^\circ\mathrm C\).
A section of the A conductor is replaced with a third material, C.
The arrangement is as follows:
- The A-C connection on the hotter side is at \(140\,^\circ\mathrm C\).
- The C-A connection on the cooler side is at \(50\,^\circ\mathrm C\).
- The B conductor remains unchanged throughout.
- The pair coefficient for C relative to B is \(S_{CB}=+6\,\mu\mathrm{V/K}\).
- All conductors are homogeneous within their respective sections, and the voltmeter operates under open-circuit conditions.
Assume the supplied coefficients are constant across their relevant temperature intervals.
(a) Calculate the original thermoelectric voltage before material C was introduced.
(b) Calculate the new thermoelectric voltage after the A-to-C-to-A replacement.
(c) Explain why the two A-C connections do not simply cancel each other.
(d) State a condition under which introducing the third material would have no net effect on the ideal measured voltage.
Solution 5
The original thermoelectric voltage is \(+5.12\,\mathrm{mV}\), but after introducing material C into the temperature-gradient region, the new voltage is only \(+2.78\,\mathrm{mV}\).
The tempting prediction is that the two A-C connections should cancel because one connection introduces material C and the other removes it.
That reasoning overlooks an important condition: the two connections are at different temperatures.
(a) Original thermoelectric voltage
Before introducing C, the entire thermocouple consists of the A-B pair.
The total temperature difference is:
\[ \Delta T=180-20=160\,\mathrm K \]
Therefore:
\[ V_{AB}=S_{AB}\Delta T \]
\[ V_{AB}=32\times160 \]
\[ V_{AB}=5120\,\mu\mathrm V \]
Hence:
\[ \boxed{V_{AB}=+5.12\,\mathrm{mV}} \]
(b) New thermoelectric voltage
We must divide the temperature range into three sections because the material pair changes within the circuit.
The sections are:
| Temperature interval | Material pair | Seebeck coefficient |
|---|---|---|
| \(180\) to \(140\,^\circ\mathrm C\) | A-B | \(+32\,\mu\mathrm{V/K}\) |
| \(140\) to \(50\,^\circ\mathrm C\) | C-B | \(+6\,\mu\mathrm{V/K}\) |
| \(50\) to \(20\,^\circ\mathrm C\) | A-B | \(+32\,\mu\mathrm{V/K}\) |
The voltage contributions must be calculated separately.
For the first section:
\[ \Delta T_1=180-140=40\,\mathrm K \]
\[ V_1=32\times40=1280\,\mu\mathrm V \]
For the second section:
\[ \Delta T_2=140-50=90\,\mathrm K \]
\[ V_2=6\times90=540\,\mu\mathrm V \]
For the third section:
\[ \Delta T_3=50-20=30\,\mathrm K \]
\[ V_3=32\times30=960\,\mu\mathrm V \]
Add the contributions:
\[ \begin{aligned} V_{\mathrm{new}} &=V_1+V_2+V_3\\ &=1280+540+960\\ &=2780\,\mu\mathrm V \end{aligned} \]
Therefore:
\[ \boxed{V_{\mathrm{new}}=+2.78\,\mathrm{mV}} \]
The new voltage is smaller than the original voltage because the C-B section has a smaller Seebeck coefficient than the original A-B section.
(c) Why don’t the two A-C connections cancel?
The two connections occur at different temperatures.
The thermoelectric contribution depends on both the material properties and the temperature interval across which those properties act.
The C section spans:
\[ \Delta T_C=140-50=90\,\mathrm K \]
Consequently, it contributes a thermoelectric voltage that differs from the voltage that would have been generated by the original A section.
The physical contacts themselves are not the source of the net thermoelectric voltage. The important change is the material response across a region with a temperature gradient.
The two connections therefore cannot be treated as equal-and-opposite contributions simply because the same two materials meet at each connection.
(d) When would the third material have no net effect?
One important condition is that both new A-C connections are maintained at the same temperature.
In that case, there is no temperature difference across the inserted C section, and its additional thermoelectric contribution is zero.
The two connections also introduce no net change to the ideal thermocouple voltage.
Another possibility is that the C-B Seebeck coefficient exactly matches the original A-B coefficient throughout the relevant temperature interval.
This question demonstrates an important feature of thermoelectric circuits: the location of temperature gradients matters whenever the material properties change along a conductor.
09Where to go next
The Seebeck effect explains how a temperature difference can generate an electrical potential difference. It also provides the foundation for understanding thermocouples and thermoelectric generators.
The closely related Peltier effect considers the opposite direction of energy conversion: when an electrical current passes through a junction between different materials, heat can be absorbed or released at that junction.
Understanding both effects gives you a broader picture of how heat flow and electrical behaviour are connected in conducting materials.
10Further reading
- NSW Education Standards Authority – Physics 11-12 Syllabus (2025). The official syllabus and its Year 11 Electricity and magnetism focus area.
- Bureau International des Poids et Mesures – Guide on Secondary Thermometry: Thermocouple Thermometry, Part 1. An authoritative explanation of thermocouple operation, reference junctions, calibration, and measurement errors.
- European Thermoelectric Society – Thermoelectric Effects. An introduction to the Seebeck, Peltier, and Thomson effects.
- MIT – Solid State Physics: Transport Properties of Solids. University-level teaching notes covering thermoelectric phenomena and the physical basis of thermopower.
- OpenStax – University Physics, Resistivity and Resistance. A useful reference for the relationship between resistance, resistivity, conductor dimensions, and temperature.