Photon Counting: Detection Efficiency and Count Statistics
Explore how photons become measurable detection events, why detectors miss some photons, and how statistical fluctuations affect photon counts. This extension guide introduces detection efficiency, Poisson statistics, dark counts, detector dead time, and photon-number resolution.
Send one photon towards a detector in each of 100 separate trials. If the detector has a 70% detection efficiency, should it record exactly 70 clicks?
Not necessarily. Seventy is the expected number of detections across many repetitions. A particular batch might produce 66, 72, or 75 clicks. Some photons will arrive without producing a registered event.
This raises an interesting problem. If photons have discrete energies and detectors can respond to individual photons, why can’t we simply count every photon that arrives?
The answer involves three ideas: detection is probabilistic, detectors have limitations, and even a perfectly stable light source produces statistical fluctuations in its counts.
This article goes beyond the required Year 12 Physics syllabus content. It uses the particle model from the Year 12 Nature of light focus area in the NSW Physics 11-12 Syllabus (2025) as its starting point, then introduces photon-counting efficiency and statistics as additional physics. These extension concepts are not required HSC assessment content.
01The photon model: energy comes in discrete amounts
The particle model of light describes electromagnetic radiation in terms of photons. A photon is a quantum of electromagnetic radiation, carrying a discrete amount of energy.
The energy of an individual photon depends on the frequency of the radiation:
\[ E_\gamma=hf \]
Where:
- \(E_\gamma\) is the energy of one photon, measured in joules (J).
- \(h\) is Planck’s constant, \(6.626\times10^{-34}\,\mathrm{J\,s}\).
- \(f\) is the frequency of the radiation, measured in hertz (Hz).
Using the relationship between frequency and wavelength, we can also write:
\[ E_\gamma=\frac{hc}{\lambda} \]
Here, \(c\) is the speed of light in a vacuum, approximately \(3.00\times10^8\,\mathrm{m\,s^{-1}}\), and \(\lambda\) is the wavelength in metres.
The important idea is that photon energy is quantised, and its value is determined by frequency. Frequency itself is not restricted to a discrete set of values.
For monochromatic light (light with one frequency), every photon has the same energy.
Why wavelength matters when counting photons
Imagine two light sources producing the same optical power. One emits red light, and the other emits blue light.
Which source produces more photons each second?
Blue light has a higher frequency, so each blue photon carries more energy. If both sources deliver the same energy per second, the red source must produce more photons to deliver that energy.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
This gives us an important distinction:
- Increasing the power of monochromatic light increases the number of photons arriving per second, assuming the frequency stays constant.
- Increasing the frequency increases the energy carried by each photon.
- At the same optical power, increasing the frequency decreases the number of photons arriving per second.
This distinction becomes useful when interpreting photon-counting measurements. A detector measuring light at two different wavelengths cannot automatically assume that the same optical power means the same photon arrival rate.
02From photons to detector counts
A photon-counting detector converts interactions with light into measurable electrical signals.
For example, a photon absorbed by a semiconductor detector can produce an electrical charge that contributes to an output pulse. Electronic circuitry then identifies the pulse as a detection event.
Instead of measuring a continuous signal proportional to light intensity, the system records individual events.
However, not every photon produces a count.
A photon might:
- Be lost through imperfect optical coupling before reaching the detector.
- Pass through or reflect from the detector without being absorbed.
- Be absorbed without producing the required measurable electrical signal.
- Produce an internal response that is too weak to pass the electronic detection threshold.
A detector’s performance is therefore described using its detection efficiency.
Detection efficiency
Detection efficiency is the probability that a photon incident on a detector produces a registered detection event.
We represent this probability using \(\eta\) (the Greek letter eta).
For this article, we use the system detection efficiency: the probability that a photon entering the complete detection system produces a measurable output event.
\[ \eta=\frac{\text{expected detected signal photons}} {\text{incident photons}} \]
This simplified ratio is useful when the detector operates under stable conditions and individual detection probabilities are independent.
For example, suppose 1,000 photons enter a system with a detection efficiency of 0.70.
The expected number of registered detections is:
\[ \mu=1000\times0.70 \]
\[ \mu=700 \]
The detector is expected to register 700 events.
It does not follow that exactly 700 events will occur in every measurement. The actual result fluctuates.
Detection efficiency is also not necessarily constant under all conditions. It can depend on wavelength, optical losses, and the detector’s operating conditions, including its count rate.
The National Institute of Standards and Technology (NIST) distinguishes the efficiency of individual detector components from the efficiency of the complete detection system.
Connecting optical power to expected counts
We can now combine the photon energy model with detection efficiency.
Suppose a monochromatic light source has optical power \(P\), measured in watts.
Since each photon carries energy \(E_\gamma\), the incident photon rate is:
\[ \Phi=\frac{P}{E_\gamma} \]
Here, \(\Phi\) is the photon arrival rate, measured in photons per second.
Over an observation time \(t\), measured in seconds, the expected number of incident photons is:
\[ N_{\mathrm{in}}=\Phi t \]
If the system detection efficiency is \(\eta\), the expected number of registered signal counts is:
\[ \boxed{\mu=\eta\Phi t} \]
Combining the equations gives:
\[ \boxed{\mu=\frac{\eta Pt}{E_\gamma}} \]
This is one of the most useful relationships in photon counting.
It connects the energy of individual photons, the intensity of the light source, the detector’s efficiency, and the observation time.
Worked example: calculating expected photon counts
A monochromatic light source emits light with a wavelength of 650 nm and an optical power of \(2.0\times10^{-15}\,\mathrm W\).
The light enters a detector with a system detection efficiency of 0.30.
Calculate the expected number of registered signal counts during a 0.50 s measurement.
Assume the detector has negligible dark counts and dead time at this operating rate.
Step 1
Using:
\[ E_\gamma=\frac{hc}{\lambda} \]
Substitute the known values:
\[ E_\gamma= \frac{(6.626\times10^{-34})(3.00\times10^8)} {650\times10^{-9}} \]
Therefore:
\[ E_\gamma=3.06\times10^{-19}\,\mathrm J \]
Each photon carries approximately \(3.06\times10^{-19}\) J.
Step 2
The photon rate is the optical power divided by the energy per photon:
\[ \Phi=\frac{P}{E_\gamma} \]
\[ \Phi=\frac{2.0\times10^{-15}} {3.06\times10^{-19}} \]
\[ \Phi=6.54\times10^3\, \mathrm{photons\,s^{-1}} \]
Approximately 6,540 photons enter the system each second.
Step 3
\[ N_{\mathrm{in}}=\Phi t \]
\[ N_{\mathrm{in}}=6.54\times10^3\times0.50 \]
\[ N_{\mathrm{in}}=3.27\times10^3 \]
Approximately 3,270 photons enter during the measurement.
Step 4
\[ \mu=\eta N_{\mathrm{in}} \]
\[ \mu=0.30\times3270 \]
\[ \boxed{\mu\approx982\text{ counts}} \]
The detector is expected to register approximately 982 signal counts.
Notice that the incident photon number and the detected count are different quantities. The detector does not register every photon.
03Why photon counts fluctuate
Suppose a light source is operating at a constant average photon arrival rate.
You measure its output for one second, then repeat the measurement several times.
Would you get exactly the same number of counts every time?
No. Even when the average source intensity remains constant, the arrival times of individual photons can be random.
Some measurement intervals contain slightly more photons than others. This randomness produces fluctuations in the measured count.
For many ordinary photon-counting measurements, we can model independent photon arrivals at a constant average rate using a Poisson distribution.
This model is particularly useful for understanding photon shot noise.
The Poisson distribution
The probability of detecting exactly \(k\) events during a measurement interval is:
\[ \boxed{ \Pr(C=k)=\frac{e^{-\mu}\mu^k}{k!} } \]
Where:
- \(C\) is the number of detected events.
- \(k\) is the particular integer number of events being considered.
- \(\mu\) is the expected number of detected events during the interval.
- \(e\) is the mathematical constant approximately equal to 2.718.
- \(k!\) is the factorial of \(k\).
The mean of a Poisson distribution is \(\mu\).
Its variance is also \(\mu\), meaning that its standard deviation is:
\[ \boxed{\sigma_C=\sqrt{\mu}} \]
Here, \(\sigma_C\) represents the typical statistical spread in the number of counts across repeated measurements.
This relationship is important because it tells us how much variation to expect even when the detector and light source are operating normally.
- μ = 1, σ = 1
- μ = 4, σ = 2
- μ = 16, σ = 4
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
What does the square root relationship mean?
Consider a detector with an expected count of 100.
Its standard deviation is:
\[ \sigma_C=\sqrt{100}=10 \]
The expected count is 100, but individual measurements naturally fluctuate.
The relative statistical uncertainty is:
\[ \frac{\sigma_C}{\mu} =\frac{10}{100} =0.10 \]
This corresponds to 10%.
Now imagine increasing the measurement time so that the expected count becomes 10,000.
The new standard deviation is:
\[ \sigma_C=\sqrt{10000}=100 \]
The absolute spread has increased from 10 to 100 counts.
However, the relative uncertainty has decreased:
\[ \frac{100}{10000}=0.01 \]
This is just 1%.
So, collecting more photons improves the relative precision of a measurement, even though the absolute fluctuations become larger.
The general relationship is:
\[ \boxed{ \text{Relative uncertainty}=\frac{1}{\sqrt{\mu}} } \]
This is the counting-statistics limit for the idealised Poisson model.
Worked example: comparing measurement precision
A detector records an average of 400 signal counts during a 1.0 s measurement.
Assuming Poisson statistics, calculate the standard deviation and relative uncertainty.
Then determine what happens if the measurement time increases to 4.0 s while the source and detector remain unchanged.
Step 1
For a Poisson distribution:
\[ \sigma_C=\sqrt{\mu} \]
\[ \sigma_C=\sqrt{400} \]
\[ \boxed{\sigma_C=20\text{ counts}} \]
The expected count is 400, with a typical statistical spread of 20 counts.
Step 2
\[ \text{Relative uncertainty} =\frac{\sigma_C}{\mu} \]
\[ =\frac{20}{400} \]
\[ \boxed{5\%} \]
Step 3
The observation time increases by a factor of four.
Since the source’s average count rate is unchanged, the expected number of counts also increases by a factor of four.
\[ \mu_{\mathrm{new}}=4\times400 \]
\[ \mu_{\mathrm{new}}=1600 \]
Step 4
\[ \sigma_{C,\mathrm{new}}=\sqrt{1600} \]
\[ \boxed{\sigma_{C,\mathrm{new}}=40\text{ counts}} \]
Step 5
\[ \frac{40}{1600}\times100 \]
\[ \boxed{2.5\%} \]
The longer measurement has twice the absolute standard deviation but half the relative uncertainty.
This is why longer observation times are useful when measuring weak light. They reduce the importance of random counting fluctuations relative to the measured signal.
The Poisson model assumes independent events occurring at a constant average rate. It is not a universal description of every possible light source. The source’s statistical properties can change the distribution of measured counts.
University teaching materials from the University of Florida and the University of Sheffield provide further explanations of photon-counting statistics.
04A click is not necessarily one photon
There is another important limitation.
Not every photon-counting detector can distinguish between one detected photon and several detected photons arriving within the same measurement window.
Some detectors are called non-photon-number-resolving detectors, or threshold detectors.
They operate using a simple rule:
- If no photon is detected, there is no click.
- If one or more photons are detected, there is a click.
The detector can identify the presence of light without determining exactly how many photons were detected.
Other detectors are photon-number-resolving detectors. These can distinguish between different numbers of detected photons, within their operating range.
Calculating the probability of a click
Consider a pulsed light source.
Each pulse has an average of \(\mu_{\mathrm{in}}\) incident photons. Assume the number of photons per pulse follows a Poisson distribution.
A detector with efficiency \(\eta\) independently detects each incident photon with probability \(\eta\).
The average number of detected photons per pulse is:
\[ \mu_{\mathrm{det}}=\eta\mu_{\mathrm{in}} \]
For Poisson-distributed arrivals, the probability of detecting zero photons is:
\[ \Pr(0)=e^{-\eta\mu_{\mathrm{in}}} \]
Therefore, the probability of detecting at least one photon, and hence producing a click, is:
\[ \boxed{ p_{\mathrm{click}}=1-e^{-\eta\mu_{\mathrm{in}}} } \]
Here, \(p_{\mathrm{click}}\) is the probability of a registered click during one measurement gate.
A gate is simply the defined time window associated with one incoming light pulse.
Notice that this equation gives the probability of a click, not the expected number of photons detected.
That difference becomes important when multiple photons can arrive during the same gate.
Worked example: calculating a detector’s click probability
A pulsed light source produces an average of 2.0 incident photons per pulse.
The system detection efficiency is 0.40.
A threshold detector produces at most one click per pulse. It has enough time to recover between successive pulses.
Calculate:
- The probability of a click for one pulse.
- The expected number of clicks over 1,000 pulses.
Assume there are no dark counts.
Step 1
\[ \mu_{\mathrm{det}}=\eta\mu_{\mathrm{in}} \]
\[ =0.40\times2.0 \]
\[ \mu_{\mathrm{det}}=0.80 \]
The mean number of detected photons is 0.80 per pulse.
Step 2
For Poisson-distributed detected photons:
\[ \Pr(0)=e^{-0.80} \]
\[ \Pr(0)\approx0.449 \]
There is approximately a 44.9% probability of detecting no photons during a pulse.
Step 3
A click occurs whenever at least one photon is detected.
\[ p_{\mathrm{click}}=1-\Pr(0) \]
\[ p_{\mathrm{click}}=1-0.449 \]
\[ \boxed{p_{\mathrm{click}}=0.551} \]
There is approximately a 55.1% probability of a click per pulse.
Step 4
For 1,000 independent pulses:
\[ \mu_{\mathrm{click}}=1000\times0.551 \]
\[ \boxed{\mu_{\mathrm{click}}\approx551\text{ clicks}} \]
A threshold detector is therefore expected to produce approximately 551 clicks.
A photon-number-resolving detector operating with the same efficiency would have an expected detected photon number of 800 across the 1,000 pulses.
Why are these results different?
Some pulses contain multiple detected photons. The threshold detector registers only one click for those pulses, whereas a photon-number-resolving detector can distinguish the multiple detections.
The detector’s output is not necessarily a direct count of every photon.
05Where the simple counting model breaks down
The relationships developed so far are useful, but real detectors introduce additional effects.
Three limitations are particularly important.
1. Dark counts and background light
A detector can sometimes produce an output event even when no photon from the intended light source arrives.
These events are called dark counts.
They can arise from processes inside the detector, including thermally generated charge carriers or unwanted thermal radiation.
Background counts are slightly different. They can result from unwanted light entering the optical system, such as stray light.
Both can increase the measured count.
Suppose the expected signal count is \(\mu_{\mathrm{signal}}\), and the detector has a dark count rate \(D\), measured in counts per second.
During an observation lasting \(t\) seconds, the expected number of dark counts is:
\[ \mu_{\mathrm{dark}}=Dt \]
If the signal and dark counts are independent Poisson processes, the expected raw count becomes:
\[ \boxed{ \mu_{\mathrm{raw}} =\mu_{\mathrm{signal}}+Dt } \]
The standard deviation of the raw count is:
\[ \boxed{ \sigma_{\mathrm{raw}}= \sqrt{\mu_{\mathrm{signal}}+Dt} } \]
This is important because subtracting the expected dark count does not remove the statistical fluctuations associated with those events.
For very weak light, dark counts can make up a substantial proportion of the total measured events.
2. Detector dead time
After registering a photon, some detectors require a short recovery period before they can register another event.
This period is called the dead time.
During dead time, another arriving photon may not produce a measurable output.
Consider a detector receiving photons at a high rate.
If two photons arrive close together, the first may produce a click while the second arrives before the detector has recovered.
The second photon is then missed.
As the photon arrival rate increases, this effect can cause the measured count rate to become lower than the actual detection rate would suggest.
This means that the detector’s effective efficiency can depend on how quickly photons arrive.
A simple Poisson model without dead time will no longer describe the measured output accurately.
3. The photon source itself affects the statistics
So far, we have used Poisson statistics to describe random photon arrivals.
However, this is not the only possible distribution.
The statistical behaviour depends on the light source.
Consider two possible sources:
- A source that emits exactly one photon in every pulse.
- A source whose photon number varies randomly according to a Poisson distribution.
Both sources could have the same average number of photons per pulse, but their individual pulse distributions are different.
There is also a useful distinction between randomness in the source and randomness in the detector.
If exactly \(N\) photons arrive, and each has an independent probability \(\eta\) of being detected, the detected photon number follows a binomial distribution:
\[ \Pr(C=k\mid N) = \binom Nk\eta^k(1-\eta)^{N-k} \]
Here:
- \(N\) is the fixed number of incident photons.
- \(k\) is the number of detected photons.
- \(\eta\) is the detection efficiency.
- \(\binom Nk\) counts the possible ways of selecting which photons are detected.
The expected detected photon number is:
\[ \mu=N\eta \]
The variance is:
\[ \sigma_C^2=N\eta(1-\eta) \]
This differs from Poisson statistics.
For a Poisson-distributed incident photon number, independent detection losses preserve the Poisson distribution, but reduce its mean.
For a source that emits exactly one photon per pulse, the input distribution is not Poisson. The randomness in the detected output comes from the detector’s probability of registering that photon.
This distinction will matter in the final question.
06Questions and solutions
The following questions build from standard photon-counting calculations towards more demanding problems involving detector limitations and the statistical properties of light sources.
Question 1
A monochromatic light source emits photons with a wavelength of 500 nm and an optical power of \(1.0\times10^{-15}\,\mathrm W\).
The light enters a detector with a system detection efficiency of 0.20.
Calculate the expected number of registered signal counts during a 0.10 s measurement.
Assume there are no dark counts and detector dead time is negligible.
Solution 1
The detector is expected to register approximately 50 signal counts.
First, calculate the energy of each photon:
\[ E_\gamma=\frac{hc}{\lambda} \]
\[ E_\gamma= \frac{(6.626\times10^{-34})(3.00\times10^8)} {500\times10^{-9}} \]
\[ E_\gamma=3.98\times10^{-19}\,\mathrm J \]
Next, calculate the number of photons arriving during the measurement:
\[ N_{\mathrm{in}}=\frac{Pt}{E_\gamma} \]
\[ N_{\mathrm{in}}= \frac{(1.0\times10^{-15})(0.10)} {3.98\times10^{-19}} \]
\[ N_{\mathrm{in}}\approx251 \]
Only 20% of these photons are expected to produce registered events.
\[ \mu=\eta N_{\mathrm{in}} \]
\[ \mu=0.20\times251 \]
\[ \boxed{\mu\approx50\text{ counts}} \]
The result is an expected count, not a guaranteed result for one measurement.
Question 2
A photon-counting detector observes a stable light source. The detected signal follows Poisson statistics, with an average count rate of 900 counts per second.
Two measurements are performed:
- Measurement A lasts 1.0 s and records 930 counts.
- Measurement B lasts 4.0 s and records 3,540 counts.
Determine:
(a) The expected count and standard deviation for each measurement.
(b) The relative statistical uncertainty for each measurement.
(c) Whether the results provide convincing evidence that the source intensity changed between measurements.
Solution 2
Neither measurement provides convincing evidence of a change in source intensity. Both observed counts are one standard deviation away from their expected values.
(a) Expected counts and standard deviations
For measurement A:
\[ \mu_A=900\times1.0=900 \]
\[ \sigma_A=\sqrt{900} \]
\[ \boxed{\sigma_A=30\text{ counts}} \]
For measurement B:
\[ \mu_B=900\times4.0=3600 \]
\[ \sigma_B=\sqrt{3600} \]
\[ \boxed{\sigma_B=60\text{ counts}} \]
(b) Relative statistical uncertainty
For measurement A:
\[ \frac{\sigma_A}{\mu_A}\times100 = \frac{30}{900}\times100 \]
\[ \boxed{3.33\%} \]
For measurement B:
\[ \frac{\sigma_B}{\mu_B}\times100 = \frac{60}{3600}\times100 \]
\[ \boxed{1.67\%} \]
The longer measurement has a smaller relative uncertainty.
(c) Interpreting the observations
For measurement A, the difference between the observed and expected counts is:
\[ 930-900=30 \]
This is exactly one standard deviation above the expected count.
For measurement B:
\[ 3540-3600=-60 \]
This is exactly one standard deviation below the expected count.
Both observations are compatible with ordinary statistical fluctuations.
A tempting interpretation is that measurement B is worse because it differs from its expected count by 60 rather than 30. However, the absolute difference must be considered relative to the expected statistical spread.
Both deviations are equally large when measured in standard deviations.
There is therefore insufficient evidence to conclude that the source intensity changed.
Question 3
A laboratory is comparing two photon-counting detectors for measuring the same weak light source.
The source delivers an average of 500 incident photons per second. Each detector is used for 2.0 s.
Their specifications are:
| Property | Detector A | Detector B |
|---|---|---|
| System detection efficiency | 0.60 | 0.30 |
| Dark count rate | 20 counts/s | 2.0 counts/s |
Assume:
- Signal detections and dark counts are independent Poisson processes.
- Both detectors have negligible dead time.
- The stated dark count rates are known accurately.
- The signal-to-noise ratio is defined as the expected signal count divided by the standard deviation of the total raw count.
(a) Calculate the expected signal counts and dark counts for each detector.
(b) Calculate the standard deviation of the raw count for each detector.
(c) Determine which detector provides the better signal-to-noise ratio.
Explain why the detector with the lower dark count rate does not necessarily provide the better measurement.
Solution 3
Detector A provides the better signal-to-noise ratio, despite having a higher dark count rate.
The reason is that its greater detection efficiency produces a substantially stronger signal.
(a) Expected signal and dark counts
The expected signal count is:
\[ \mu_{\mathrm{signal}}=\eta\Phi t \]
For detector A:
\[ \mu_{\mathrm{signal,A}} =0.60\times500\times2.0 \]
\[ \boxed{\mu_{\mathrm{signal,A}}=600} \]
For detector B:
\[ \mu_{\mathrm{signal,B}} =0.30\times500\times2.0 \]
\[ \boxed{\mu_{\mathrm{signal,B}}=300} \]
Now calculate the dark counts using:
\[ \mu_{\mathrm{dark}}=Dt \]
For detector A:
\[ \mu_{\mathrm{dark,A}}=20\times2.0=40 \]
For detector B:
\[ \mu_{\mathrm{dark,B}}=2.0\times2.0=4 \]
The expected raw counts are therefore:
| Detector | Signal counts | Dark counts | Raw expected counts |
|---|---|---|---|
| A | 600 | 40 | 640 |
| B | 300 | 4 | 304 |
(b) Standard deviation of the raw count
Because the signal and dark counts are independent Poisson processes, their expected counts add.
The standard deviation is the square root of the total expected count.
For detector A:
\[ \sigma_A=\sqrt{600+40} \]
\[ \boxed{\sigma_A\approx25.3} \]
For detector B:
\[ \sigma_B=\sqrt{300+4} \]
\[ \boxed{\sigma_B\approx17.4} \]
(c) Comparing signal-to-noise ratios
The signal-to-noise ratio is:
\[ \mathrm{SNR} = \frac{\text{expected signal counts}} {\text{standard deviation of raw counts}} \]
For detector A:
\[ \mathrm{SNR}_A=\frac{600}{25.3} \]
\[ \boxed{\mathrm{SNR}_A\approx23.7} \]
For detector B:
\[ \mathrm{SNR}_B=\frac{300}{17.4} \]
\[ \boxed{\mathrm{SNR}_B\approx17.2} \]
Detector A has the higher signal-to-noise ratio.
Detector B produces fewer dark counts, but it also detects a much smaller proportion of the incoming signal photons.
The important lesson is that detector performance cannot be judged using dark count rate alone. Detection efficiency and statistical noise must be considered together.
Question 4
A detector receives two light pulses, separated by 4 microseconds (\(4\,\mathrm{\mu s}\)).
Each pulse contains exactly one photon.
The detector has:
- A detection efficiency of 0.80.
- A dead time of \(5\,\mathrm{\mu s}\) following each successful detection.
Assume the detector is fully recovered before the first pulse and that separate pulse pairs are sufficiently far apart for complete recovery.
(a) Calculate the probability of recording at least one click during a pulse pair.
(b) Calculate the expected number of registered clicks per pair.
(c) Explain why multiplying the number of incident photons by the detection efficiency gives an incorrect result if used without considering dead time.
Solution 4
The expected number of registered clicks is 0.96 per pulse pair.
The key is that the probability of detecting the second photon depends on whether the first photon was detected.
Step 1: Consider the first photon.
The probability of detecting it is:
\[ P(\text{first detected})=0.80 \]
The probability of missing it is:
\[ P(\text{first missed})=1-0.80 \]
\[ P(\text{first missed})=0.20 \]
Step 2: Consider what happens if the first photon is detected.
The first detection produces one click.
However, the detector then enters its \(5\,\mathrm{\mu s}\) dead time.
The second photon arrives after only \(4\,\mathrm{\mu s}\), so it arrives while the detector is unable to register another event.
Therefore, if the first photon is detected, the total number of clicks is exactly one.
Step 3: Consider what happens if the first photon is missed.
The detector has not produced a detection event, so it has not entered dead time.
It remains ready to detect the second photon.
The probability of detecting the second photon is therefore still 0.80.
The probability of detecting the second photon after missing the first is:
\[ 0.20\times0.80=0.16 \]
Step 4: Calculate the probability of at least one click.
There are two possible ways to obtain a click:
- Detect the first photon.
- Miss the first photon, then detect the second.
Therefore:
\[ P(\text{at least one click}) =0.80+0.16 \]
\[ \boxed{P(\text{at least one click})=0.96} \]
Step 5: Calculate the expected number of clicks.
Because the detector can register at most one click from this pulse pair, the expected number of clicks is equal to the probability of obtaining at least one click.
\[ \boxed{\mu_{\mathrm{click}}=0.96} \]
Why does the simple efficiency calculation fail?
A tempting calculation is:
\[ 2\times0.80=1.60\text{ clicks} \]
This would be correct if the detector could independently register both photons.
However, the detector cannot register the second photon if the first one has already triggered its dead time.
The two detection opportunities are therefore not independent.
If the pulses were separated by \(6\,\mathrm{\mu s}\) instead, the detector would have recovered before the second pulse. The expected number of clicks would then be 1.60.
This shows why the timing of photon arrivals matters, not just their total number.
Question 5
A laboratory is testing a pulsed light source. Its average incident photon number is one photon per pulse.
Two possible models are being considered:
Model A: The source emits exactly one photon in every pulse.
Model B: The number of photons in each pulse follows a Poisson distribution with a mean of one photon per pulse.
The laboratory uses a threshold detector with a system detection efficiency of 0.50.
The detector:
- Produces at most one click per pulse.
- Is fully recovered before each new pulse.
- Has no dark counts.
- Detects each incident photon independently with probability 0.50.
The laboratory records 10,000 pulses and observes 4,020 clicks.
(a) Calculate the probability of a click per pulse for Model A.
(b) Calculate the probability of a click per pulse for Model B.
(c) Calculate the expected total clicks and standard deviation for each model over 10,000 pulses.
(d) Determine which model is more consistent with the observed 4,020 clicks. Explain why the result does not prove that model is correct.
(e) The laboratory replaces the threshold detector with a photon-number-resolving detector of the same efficiency. Calculate the expected number of detected photons per pulse under both models. Explain why measuring the mean alone cannot distinguish the two sources, and identify another statistical property that could.
Solution 5
Model B is much more consistent with the observed 4,020 clicks, even though both sources have the same average incident photon number.
The important distinction is that the sources have different photon-number distributions. A threshold detector responds to the probability of at least one successful detection, not simply the average number of incident photons.
(a) Click probability for Model A
Model A produces exactly one photon per pulse.
The detector has a 50% probability of detecting that photon.
Therefore:
\[ \boxed{p_A=0.50} \]
There is a 50% probability of a click during each pulse.
(b) Click probability for Model B
Model B has a Poisson-distributed photon number with mean:
\[ \mu_{\mathrm{in}}=1 \]
The detection efficiency is:
\[ \eta=0.50 \]
The probability of no detected photons is:
\[ P(0)=e^{-\eta\mu_{\mathrm{in}}} \]
Substitute:
\[ P(0)=e^{-0.50} \]
\[ P(0)\approx0.6065 \]
Therefore, the probability of at least one detected photon is:
\[ p_B=1-0.6065 \]
\[ \boxed{p_B\approx0.3935} \]
Model B produces a click in approximately 39.35% of pulses.
Notice that the two models have the same mean incident photon number, but different click probabilities.
(c) Expected total clicks and standard deviations
The number of clicks over repeated independent pulses follows a binomial distribution because each pulse either produces a click or does not.
For \(M\) pulses, the expected number of clicks is:
\[ \mu_{\mathrm{click}}=Mp \]
The standard deviation is:
\[ \sigma_{\mathrm{click}} =\sqrt{Mp(1-p)} \]
For Model A:
\[ \mu_A=10000\times0.50 \]
\[ \boxed{\mu_A=5000\text{ clicks}} \]
Its standard deviation is:
\[ \sigma_A= \sqrt{10000\times0.50\times0.50} \]
\[ \boxed{\sigma_A=50\text{ clicks}} \]
For Model B:
\[ \mu_B=10000\times0.3935 \]
\[ \boxed{\mu_B\approx3935\text{ clicks}} \]
Its standard deviation is:
\[ \sigma_B= \sqrt{10000\times0.3935\times0.6065} \]
\[ \boxed{\sigma_B\approx48.9\text{ clicks}} \]
The predictions are substantially different.
| Quantity | Model A | Model B |
|---|---|---|
| Click probability | 0.5000 | 0.3935 |
| Expected clicks | 5,000 | 3,935 |
| Standard deviation | 50 | 48.9 |
(d) Interpreting the observed result
The laboratory records 4,020 clicks.
For Model A, the difference from the expected count is:
\[ 4020-5000=-980 \]
In standard deviations:
\[ \frac{-980}{50}=-19.6 \]
The observed result is 19.6 standard deviations below the prediction for Model A.
For Model B:
\[ 4020-3935=85 \]
In standard deviations:
\[ \frac{85}{48.9}\approx1.74 \]
The observed result is approximately 1.74 standard deviations above the prediction for Model B.
Therefore, Model B is much more consistent with the experimental result.
However, this does not prove Model B is correct. Statistical measurements involve uncertainty, and other source distributions or detector effects could potentially produce similar results.
The conclusion is conditional on the two proposed models and the assumptions given in the question.
(e) Replacing the detector with a photon-number-resolving detector
This is where the problem becomes particularly interesting.
A photon-number-resolving detector can distinguish between different numbers of detected photons in a pulse.
For Model A, exactly one photon arrives per pulse.
The detector has a 50% probability of detecting it.
Therefore, the expected number of detected photons per pulse is:
\[ \boxed{\mu_A=1\times0.50=0.50} \]
For Model B, the incident photon number follows a Poisson distribution with mean one.
Independent detection losses reduce the mean by the efficiency factor.
Therefore:
\[ \boxed{\mu_B=1\times0.50=0.50} \]
The expected detected photon number is identical for both models.
So measuring the mean count alone cannot distinguish them.
However, their variances are different.
For Model A, the detected number per pulse can only be zero or one.
It follows a binomial distribution with \(N=1\) and \(\eta=0.50\).
Its variance is:
\[ \sigma_A^2=N\eta(1-\eta) \]
\[ \sigma_A^2=1\times0.50\times0.50 \]
\[ \boxed{\sigma_A^2=0.25} \]
For Model B, the detected photon number follows a Poisson distribution with mean 0.50.
Its variance is equal to its mean:
\[ \boxed{\sigma_B^2=0.50} \]
The two sources therefore have:
| Statistical property | Model A | Model B |
|---|---|---|
| Mean detected photons per pulse | 0.50 | 0.50 |
| Variance per pulse | 0.25 | 0.50 |
A photon-number-resolving detector allows the laboratory to examine the distribution of detected photon numbers, rather than relying only on the average.
The variance provides additional information that distinguishes the two models.
This is the central lesson of photon counting: knowing the average number of detected photons does not necessarily tell you how those photons are distributed between measurement intervals.
07Where photon counting leads next
Photon counting connects the particle model of light to the practical problem of measuring individual quantum events.
You can now distinguish between:
- The energy carried by each photon.
- The number of photons incident on a detector.
- The probability that an incident photon produces a count.
- The statistical fluctuations in repeated measurements.
- The difference between counting detection events and resolving individual photon numbers.
These distinctions become particularly useful in quantum optics, astronomical observations, and low-light imaging.
A natural next step is to investigate what happens when a photon interacts with matter and transfers energy and momentum. The related topic of Compton scattering: photon momentum and wavelength shifts develops that interaction from the same photon model.
Photon counting tells us how many detection events occur. Compton scattering helps explain what an individual photon can transfer during an interaction.