How to Find Concentration from a Calibration-Line Equation
Learn how to calculate an unknown concentration from a calibration-line equation using the gradient and intercept. Work through examples involving blank corrections, successive dilutions, interfering substances, and measurement limits.
A colourimeter gives an unknown solution an absorbance of 0.185. Its calibration-line equation is \(A=0.0250c+0.010\), where \(c\) is measured in \(\mathrm{mg\,L^{-1}}\).
A tempting calculation is to divide 0.185 by 0.0250 and report \(7.40\ \mathrm{mg\,L^{-1}}\). But there’s a problem: the calibration line does not start at zero. The correct concentration is \(7.00\ \mathrm{mg\,L^{-1}}\).
To find concentration from a calibration-line equation, identify the measured signal, substitute it into the equation, and rearrange to make concentration the subject. Then check whether the signal needs blank correction, whether the sample was diluted, and whether the result falls within the calibration’s validated range.
All numerical data in the examples and questions below are illustrative.
01What does a calibration-line equation actually tell you?
Imagine analysing a solution containing a coloured chemical species. A colourimeter measures how strongly the solution absorbs light, but it doesn’t directly measure the concentration.
You need a way to connect the measured absorbance to a known concentration. That’s what a calibration line does.
How the calibration line is produced
The process starts with several standard solutions. Each standard contains a known concentration of the substance being analysed.
The standards are measured under the same conditions:
- The same instrument and wavelength are used.
- The same cuvette path length is maintained.
- The same solvent and relevant reagents are used.
- Conditions such as pH and reaction time are kept consistent.
Inside a colourimeter or spectrophotometer, light travels from a source through a wavelength selector, then through a cuvette containing the solution. A detector on the other side measures the transmitted light.
The instrument uses this measurement to calculate absorbance.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
Before looking at the results, make a prediction. If you increase the concentration of the absorbing species while keeping everything else constant, what should happen to absorbance?
Absorbance should generally increase. A more concentrated solution contains more absorbing particles in the light path, so less light reaches the detector.
Within a suitable concentration range, this relationship is approximately linear.
The chemist measures several standards and plots absorbance against concentration.
Consider these illustrative results:
| Concentration (\(\mathrm{mg\,L^{-1}}\)) | Blank-corrected absorbance |
|---|---|
| 2.00 | 0.060 |
| 4.00 | 0.110 |
| 6.00 | 0.160 |
| 8.00 | 0.210 |
| 10.0 | 0.260 |
The results follow a straight-line relationship:
\[ A_{\text{net}}=0.0250c+0.010 \]
This is the calibration-line equation.
The unknown concentration can now be calculated from a measured absorbance.
- A_net = 0.0250c + 0.010
- Vertical intercept = 0.010
- Range starts: 2.00 mg/L
- Range ends: 10.0 mg/L
- Unknown: A_net = 0.185; c = 7.00 mg/L
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
The important point is that the equation represents an experimentally established relationship. It is not a universal equation that applies to every substance or concentration.
For a broader explanation of how light absorption produces this relationship, see UV-Visible Spectrophotometry: Calibration and Concentration.
02How to rearrange the calibration-line equation
This is the calculation you need to master.
A typical calibration equation has the form:
\[ A=mc+b \]
Where:
- \(A\) is the absorbance of the sample.
- \(c\) is the concentration of the absorbing species.
- \(m\) is the gradient of the calibration line.
- \(b\) is the vertical intercept.
The gradient describes how much the absorbance changes when concentration increases by one unit.
The intercept represents the absorbance predicted by the fitted line when concentration is zero.
For example, if:
\[ A=0.0250c+0.010 \]
then:
- The gradient is \(0.0250\).
- The intercept is \(0.010\).
Notice that the intercept is not zero. You must account for it.
Rearranging for concentration
Start with:
\[ A=mc+b \]
Step 1: Subtract the intercept from both sides.
\[ A-b=mc \]
Step 2: Divide both sides by the gradient.
\[ \boxed{c=\frac{A-b}{m}} \]
This is the main formula for finding concentration from a calibration equation when absorbance is plotted on the vertical axis and concentration is plotted on the horizontal axis.
In plain English:
- Take the measured absorbance.
- Subtract the intercept.
- Divide by the gradient.
Why can’t you simply divide absorbance by the gradient?
Because the calibration line may not pass through the origin.
If the intercept is 0.010, the line predicts an absorbance of 0.010 even when the analyte concentration is zero.
You must remove this contribution represented by the fitted intercept before dividing by the gradient.
A useful prediction: what absorbance does the equation predict at \(c=0\)?
Substituting zero gives \(A=b\). That’s exactly why the intercept must be included when rearranging the equation.
Check which variable is on each axis
Here’s a common source of mistakes.
Students often memorise \(A=mc+b\) and assume every calibration equation must have this arrangement. That’s not necessarily true.
The equation depends on which variable is plotted on each axis.
| Calibration graph | Equation | How to find concentration |
|---|---|---|
| Absorbance against concentration | \(A=mc+b\) | \(c=(A-b)/m\) |
| Concentration against absorbance | \(c=mA+b\) | Substitute absorbance directly |
| Signal against concentration | \(S=mc+b\) | \(c=(S-b)/m\) |
Always read the axis labels and units before rearranging.
If concentration is already the subject of the equation, you don’t need to rearrange it.
Check the units of the gradient
Units provide another way to catch mistakes.
Suppose concentration is measured in \(\mathrm{mg\,L^{-1}}\), while absorbance has no units.
For:
\[ A=mc+b \]
the gradient must have units:
\[ m=\mathrm{L\,mg^{-1}} \]
This is because multiplying the gradient by concentration must produce a dimensionless absorbance.
If the gradient has units of \(\mathrm{mg\,L^{-1}}\) per absorbance unit, you may instead be looking at an equation where concentration is plotted against absorbance.
Don’t ignore the units. They help you identify what the equation is actually describing.
03Do you need to correct the absorbance for a blank?
Before substituting your reading into the calibration equation, establish whether the measurement is a raw absorbance or a blank-corrected absorbance.
A blank contains the solvent and relevant reagents but no target analyte.
It accounts for background absorption from materials other than the substance being measured.
Suppose you measure:
| Measurement | Absorbance |
|---|---|
| Unknown sample | 0.245 |
| Reagent blank | 0.045 |
If the instrument has not been zeroed using the blank, calculate:
\[ A_{\text{net}}=A_{\text{sample}}-A_{\text{blank}} \]
Therefore:
\[ A_{\text{net}}=0.245-0.045 \]
\[ A_{\text{net}}=0.200 \]
The corrected absorbance is 0.200.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
However, there’s an important distinction.
What if the instrument has already been zeroed?
If you place the reagent blank in the instrument and use its zero function, the instrument has already accounted for the background.
The displayed sample absorbance is then already corrected relative to that blank.
Do not subtract the original blank reading again.
What if the calibration equation already includes the blank?
The same principle applies when reading calibration equations.
You must know how the calibration line was constructed.
| Calibration information | What to do |
|---|---|
| Raw calibration equation using uncorrected absorbance | Use the raw sample reading with the raw calibration equation |
| Blank-corrected calibration equation | Use blank-corrected sample absorbance |
| Instrument already zeroed using the matching blank | Use the displayed sample reading directly |
The fitted intercept and the measured blank are not automatically the same thing.
A calibration line can retain a non-zero intercept even after blank correction. If the equation includes that intercept, you must still use it.
Contrasting case: when subtracting the blank gives the wrong answer
Suppose the calibration equation is:
\[ A_{\text{raw}}=0.0250c+0.045 \]
The equation was fitted using raw absorbance readings, so the background contribution is already represented in the intercept.
An unknown sample has a raw absorbance of 0.245.
The correct calculation is:
\[ \begin{aligned} c&=\frac{A_{\text{raw}}-0.045}{0.0250}\\ &=\frac{0.245-0.045}{0.0250}\\ &=8.00\ \mathrm{mg\,L^{-1}} \end{aligned} \]
Suppose a student also subtracts a separately measured blank of 0.045 before using the equation.
They would calculate:
\[ c=\frac{0.200-0.045}{0.0250} \]
\[ c=6.20\ \mathrm{mg\,L^{-1}} \]
This is incorrect. The same background contribution has been removed twice.
The rule is simple: match the sample signal to the calibration equation and account for each background contribution only once.
For more detail on choosing the correct blank and avoiding double subtraction, see How to Correct Absorbance Readings for a Blank.
04Worked examples: finding concentration from a calibration equation
Worked example: Calculate concentration directly from a calibration line
A coloured solution is analysed using a spectrophotometer.
The blank-corrected calibration equation is:
\[ A_{\text{net}}=0.0250c+0.010 \]
where \(c\) is measured in \(\mathrm{mg\,L^{-1}}\).
The instrument has already been zeroed using the appropriate reagent blank.
The unknown sample produces a displayed absorbance of 0.185.
The validated calibration range is \(2.00\)-\(10.0\ \mathrm{mg\,L^{-1}}\).
Calculate the concentration of the unknown solution.
Step 1
The instrument has already corrected for the blank, so:
\[ A_{\text{net}}=0.185 \]
Step 2
\[ A_{\text{net}}=0.0250c+0.010 \]
Subtract the intercept:
\[ c=\frac{A_{\text{net}}-0.010}{0.0250} \]
Step 3
\[ \begin{aligned} c&=\frac{0.185-0.010}{0.0250}\\ &=\frac{0.175}{0.0250}\\ &=7.00\ \mathrm{mg\,L^{-1}} \end{aligned} \]
Step 4
The calculated concentration is \(7.00\ \mathrm{mg\,L^{-1}}\).
This is between the validated limits of 2.00 and \(10.0\ \mathrm{mg\,L^{-1}}\).
The result is therefore within the stated quantitative range.
The unknown solution has a concentration of:
\[ \boxed{7.00\ \mathrm{mg\,L^{-1}}} \]
Notice that we didn’t subtract another blank reading. The instrument had already performed that correction.
Worked example: Calculate original concentration after dilution
A concentrated solution containing a coloured analyte is being analysed.
The spectrophotometer has not been zeroed using the reagent blank.
The original sample produces a raw absorbance of 2.25, while the reagent blank produces an absorbance of 0.050.
The calibration equation is:
\[ A_{\text{net}}=0.0250c+0.010 \]
where \(c\) is measured in \(\mathrm{mg\,L^{-1}}\).
The validated calibration range is \(2.00\)-\(10.0\ \mathrm{mg\,L^{-1}}\).
Because the original sample is too concentrated, a \(4.00\ \mathrm{mL}\) portion is transferred into a \(50.00\ \mathrm{mL}\) volumetric flask and diluted to the mark.
The diluted sample produces a raw absorbance of 0.235.
Calculate the concentration of the original solution.
Step 1
The instrument has not been zeroed, so the blank must be subtracted.
\[ \begin{aligned} A_{\text{net}}&=2.25-0.050\\ &=2.20 \end{aligned} \]
This is far above the absorbance corresponding to the calibration’s upper concentration limit.
The calibration only covers concentrations from 2.00 to \(10.0\ \mathrm{mg\,L^{-1}}\).
We cannot use the original reading to establish a reliable concentration by extrapolating the line.
Step 2
The diluted sample gives a raw absorbance of 0.235.
The matching blank still gives 0.050.
\[ \begin{aligned} A_{\text{net}}&=0.235-0.050\\ &=0.185 \end{aligned} \]
Step 3
Use the calibration equation:
\[ A_{\text{net}}=0.0250c+0.010 \]
Rearrange:
\[ c=\frac{A_{\text{net}}-0.010}{0.0250} \]
Substitute:
\[ \begin{aligned} c_{\text{diluted}}&=\frac{0.185-0.010}{0.0250}\\ &=\frac{0.175}{0.0250}\\ &=7.00\ \mathrm{mg\,L^{-1}} \end{aligned} \]
This concentration is within the validated calibration range.
Step 4
The original sample volume was \(4.00\ \mathrm{mL}\), and the final volume was \(50.00\ \mathrm{mL}\).
\[ DF=\frac{V_{\text{final}}}{V_{\text{aliquot}}} \]
Therefore:
\[ DF=\frac{50.00}{4.00}=12.5 \]
Step 5
The spectrophotometer measured the diluted solution, so we must reverse the dilution.
\[ c_{\text{original}}=c_{\text{diluted}}\times DF \]
\[ \begin{aligned} c_{\text{original}}&=7.00\times12.5\\ &=87.5\ \mathrm{mg\,L^{-1}} \end{aligned} \]
The original solution has a concentration of:
\[ \boxed{87.5\ \mathrm{mg\,L^{-1}}} \]
The important distinction is that the calibration range applies to the solution actually measured, not necessarily the original solution.
The original concentration can be much higher than the calibration’s upper limit, provided the sample is diluted appropriately before measurement.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
05When should you trust the calculated concentration?
A calculator will produce a numerical result even when the experiment doesn’t support it.
You need to check the measurement limits before reporting a concentration.
Interpolation versus extrapolation
Suppose a calibration is validated over the concentration range:
\[ 2.00\text{-}10.0\ \mathrm{mg\,L^{-1}} \]
There are three possible situations.
| Calculated concentration | What it means | Appropriate action |
|---|---|---|
| \(6.00\ \mathrm{mg\,L^{-1}}\) | Within the calibration range | Use the result, assuming the method is otherwise valid |
| \(15.0\ \mathrm{mg\,L^{-1}}\) | Above the calibration range | Dilute the sample and measure again |
| \(0.50\ \mathrm{mg\,L^{-1}}\) | Below the calibration range | Prepare a less diluted sample or use a more suitable method |
Using the calibration line between measured standards is called interpolation.
Extending the line beyond the measured range is called extrapolation.
Interpolation is generally more defensible because the relationship has been tested over that interval.
Extrapolation assumes the same relationship continues beyond the available experimental evidence. That may not be true.
What if the concentration is below the measurement limit?
There is an important difference between detecting a substance and measuring its concentration reliably.
A substance may be present even when its signal is too small for accurate quantification.
The lower limit of quantification is the lowest concentration at which the method can provide a sufficiently reliable quantitative result.
If a calculated concentration is below this limit, you should not report it as an established quantitative result.
You also cannot automatically conclude that the substance is absent.
The detection limit and the quantification limit are different. Only use numerical limits supplied or established for the method.
What if another substance also absorbs light?
A calibration equation assumes that the measured signal is related to the analyte being investigated.
But imagine your unknown sample contains two coloured substances, both of which absorb light at the selected wavelength.
The instrument measures their combined contribution.
A reagent blank can correct background from the solvent and reagents, but it cannot necessarily remove the contribution from another coloured substance already present in the unknown.
This means the calculated concentration could be higher than the actual concentration of the target analyte.
Diluting the sample does not automatically solve this problem. Both substances may become less concentrated together, producing a consistent but misleading result.
You must also consider whether the unknown and standards have comparable chemical conditions, including pH, wavelength, and reaction conditions.
A calibration equation is only useful when the sample behaves sufficiently like the standards.
06A reliable method to remember
When you encounter a calibration-line question in HSC Chemistry, work through these decisions in order.
Read the axes. Identify which variable represents the measured signal and which represents concentration.
Check the equation. If \(A=mc+b\), rearrange to \(c=(A-b)/m\). If concentration is already the subject, substitute the signal directly.
Check the blank. Determine whether the sample reading is raw or already corrected. Remove background contributions only once.
Calculate the measured concentration. Substitute the appropriate signal into the calibration equation and include units.
Check the measurement range. Make sure the measured solution falls within the validated calibration range.
Reverse any dilution. Multiply by the appropriate dilution factor for preparation steps not already accounted for by the calibration.
Evaluate the result. Consider whether interference, chemical changes, or unsuitable measurement conditions could make the numerical result unreliable.
The most common mistakes aren’t difficult algebra. They’re using the wrong signal, ignoring the intercept, subtracting a blank twice, or applying a dilution factor to a concentration that already accounts for that preparation step.
07Questions and solutions
Question 1
A solution containing a coloured compound is analysed using a colourimeter.
The calibration equation is:
\[ A_{\text{net}}=0.0400c+0.008 \]
where \(c\) is measured in \(\mathrm{mg\,L^{-1}}\).
The instrument has already been zeroed using the appropriate blank.
The unknown solution produces a displayed absorbance of 0.248.
The validated calibration range is \(1.00\)-\(8.00\ \mathrm{mg\,L^{-1}}\).
Calculate the concentration of the unknown solution.
Solution 1
The concentration is:
\[ \boxed{6.00\ \mathrm{mg\,L^{-1}}} \]
The calibration equation is:
\[ A_{\text{net}}=0.0400c+0.008 \]
Rearrange to make concentration the subject:
\[ c=\frac{A_{\text{net}}-0.008}{0.0400} \]
Substitute the measured absorbance:
\[ \begin{aligned} c&=\frac{0.248-0.008}{0.0400}\\ &=\frac{0.240}{0.0400}\\ &=6.00\ \mathrm{mg\,L^{-1}} \end{aligned} \]
The answer is within the validated range of 1.00-\(8.00\ \mathrm{mg\,L^{-1}}\).
There is no need to subtract another blank reading because the instrument has already been zeroed.
The key calculation is subtracting the intercept before dividing by the gradient.
Question 2
A calibration graph has concentration on its vertical axis and blank-corrected absorbance on its horizontal axis.
Its equation is:
\[ c=32.0A_{\text{net}}-0.40 \]
where \(c\) is measured in \(\mathrm{mg\,L^{-1}}\).
An unknown sample produces a raw absorbance of 0.410. The matching reagent blank produces an absorbance of 0.060.
The instrument has not been zeroed using the blank.
The validated concentration range is \(2.0\)-\(12.0\ \mathrm{mg\,L^{-1}}\).
Calculate the concentration of the unknown sample and explain why the calibration equation should not be rearranged into the usual absorbance-against-concentration form.
Solution 2
The concentration is:
\[ \boxed{10.8\ \mathrm{mg\,L^{-1}}} \]
The first step is to correct the sample absorbance.
\[ \begin{aligned} A_{\text{net}}&=A_{\text{sample}}-A_{\text{blank}}\\ &=0.410-0.060\\ &=0.350 \end{aligned} \]
Now inspect the calibration equation:
\[ c=32.0A_{\text{net}}-0.40 \]
Concentration is already the subject of the equation.
Therefore, substitute the corrected absorbance directly:
\[ \begin{aligned} c&=32.0(0.350)-0.40\\ &=11.2-0.40\\ &=10.8\ \mathrm{mg\,L^{-1}} \end{aligned} \]
The result falls within the validated range.
The tempting mistake is to assume every calibration equation must have absorbance on the left-hand side.
But the graph’s vertical axis is concentration, not absorbance. Its gradient describes the change in concentration per unit of absorbance.
The equation should therefore be used as supplied.
Question 3
A concentrated solution containing a coloured ion undergoes two successive dilutions before analysis.
The first preparation transfers \(5.00\ \mathrm{mL}\) of the original solution into a \(100.0\ \mathrm{mL}\) volumetric flask.
Next, \(10.0\ \mathrm{mL}\) of the first diluted solution is transferred into a \(25.0\ \mathrm{mL}\) volumetric flask and diluted to the mark.
The final diluted solution is measured using a spectrophotometer.
Its raw absorbance is 0.204, and the matching reagent blank has an absorbance of 0.040.
The calibration equation is:
\[ A_{\text{net}}=0.0300c+0.004 \]
where \(c\) is measured in \(\mathrm{mg\,L^{-1}}\).
The validated calibration range is \(1.00\)-\(8.00\ \mathrm{mg\,L^{-1}}\).
Calculate the concentration of the original solution.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
Solution 3
The original solution has a concentration of:
\[ \boxed{267\ \mathrm{mg\,L^{-1}}} \]
We need to correct the absorbance, calculate the final diluted concentration, and then reverse both dilutions.
Step 1: Correct the absorbance.
\[ \begin{aligned} A_{\text{net}}&=0.204-0.040\\ &=0.164 \end{aligned} \]
Step 2: Calculate the concentration of the final diluted solution.
Rearrange the calibration equation:
\[ c=\frac{A_{\text{net}}-0.004}{0.0300} \]
Substitute:
\[ \begin{aligned} c_{\text{final}}&=\frac{0.164-0.004}{0.0300}\\ &=\frac{0.160}{0.0300}\\ &=5.33\ \mathrm{mg\,L^{-1}} \end{aligned} \]
This is within the validated calibration range.
Step 3: Calculate the first dilution factor.
\[ F_1=\frac{100.0}{5.00} \]
\[ F_1=20.0 \]
Step 4: Calculate the second dilution factor.
\[ F_2=\frac{25.0}{10.0} \]
\[ F_2=2.50 \]
Step 5: Calculate the total dilution factor.
For successive dilutions, multiply the individual factors:
\[ \begin{aligned} F_{\text{total}}&=F_1\times F_2\\ &=20.0\times2.50\\ &=50.0 \end{aligned} \]
Step 6: Calculate the original concentration.
\[ \begin{aligned} c_{\text{original}}&=5.33\times50.0\\ &=266.5\ \mathrm{mg\,L^{-1}} \end{aligned} \]
To three significant figures:
\[ \boxed{c_{\text{original}}=267\ \mathrm{mg\,L^{-1}}} \]
A common mistake is to reverse only the final dilution.
Remember that the spectrophotometer measures the solution after both preparation steps. Both dilution factors must be included to recover the original concentration.
Question 4
A sample contains a target coloured compound, \(X\), and another coloured substance, \(Y\).
The calibration standards contain only \(X\).
The blank-corrected calibration equation is:
\[ A_{\text{net}}=0.0200c_X+0.005 \]
where \(c_X\) is measured in \(\mathrm{mg\,L^{-1}}\).
The validated concentration range is \(2.00\)-\(10.0\ \mathrm{mg\,L^{-1}}\).
The spectrophotometer has not been zeroed using the reagent blank.
The following measurements are obtained:
| Measurement | Raw absorbance |
|---|---|
| Original sample | 0.300 |
| Reagent blank | 0.025 |
| Sample after a 1:2 dilution | 0.165 |
An independent selectivity test establishes that substance \(Y\) contributes a net absorbance of 0.070 in the original sample and 0.035 in the diluted sample.
The diluted sample was prepared by transferring \(5.00\ \mathrm{mL}\) of the original sample into a \(10.00\ \mathrm{mL}\) volumetric flask.
Calculate the original concentration of \(X\).
Explain why simply applying the calibration equation to the diluted sample and reversing the dilution would produce an incorrect answer.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
Solution 4
The actual concentration of \(X\) is:
\[ \boxed{10.0\ \mathrm{mg\,L^{-1}}} \]
The important complication is that the measured signal contains contributions from both \(X\) and \(Y\).
The calibration equation was developed using \(X\) alone. We must therefore account for the contribution from \(Y\) before calculating the concentration of \(X\).
Step 1: Correct the diluted sample’s absorbance for the reagent blank.
\[ \begin{aligned} A_{\text{net}}&=0.165-0.025\\ &=0.140 \end{aligned} \]
Step 2: Remove the contribution from \(Y\).
The independent test establishes that \(Y\) contributes 0.035 absorbance units in the diluted sample.
Therefore, the absorbance represented by the target-only calibration, including its fitted intercept, is:
\[ \begin{aligned} A_X&=0.140-0.035\\ &=0.105 \end{aligned} \]
Step 3: Calculate the concentration of \(X\) in the diluted solution.
Use:
\[ A_X=0.0200c_X+0.005 \]
Rearrange:
\[ c_X=\frac{A_X-0.005}{0.0200} \]
Substitute:
\[ \begin{aligned} c_{X,\text{diluted}}&=\frac{0.105-0.005}{0.0200}\\ &=\frac{0.100}{0.0200}\\ &=5.00\ \mathrm{mg\,L^{-1}} \end{aligned} \]
Step 4: Reverse the dilution.
The dilution factor is:
\[ DF=\frac{10.00}{5.00}=2.00 \]
Therefore:
\[ \begin{aligned} c_{X,\text{original}}&=5.00\times2.00\\ &=10.0\ \mathrm{mg\,L^{-1}} \end{aligned} \]
The original concentration of \(X\) is:
\[ \boxed{10.0\ \mathrm{mg\,L^{-1}}} \]
Now consider the tempting alternative.
If we ignore \(Y\), the diluted sample would give:
\[ \begin{aligned} c_{\text{apparent}}&=\frac{0.140-0.005}{0.0200}\\ &=6.75\ \mathrm{mg\,L^{-1}} \end{aligned} \]
Reversing the dilution gives:
\[ 6.75\times2=13.5\ \mathrm{mg\,L^{-1}} \]
This is an apparent concentration of \(X\), not its true concentration.
In fact, the contribution from \(Y\) is also reduced by the same dilution factor. The two substances continue to contribute to the measured absorbance in the same way.
Consequently, repeating the dilution and obtaining consistent results would not prove that the calibration is selective for \(X\).
This is an important limitation of calibration equations: accurate arithmetic cannot separate substances that contribute to the same measured signal without additional analytical information.
Question 5
A laboratory uses a colour-forming reagent to measure the concentration of an analyte, \(X\).
Each calibration standard is prepared as follows:
- Transfer \(10.00\ \mathrm{mL}\) of a standard solution containing a known concentration of \(X\) into a volumetric flask.
- Add \(5.00\ \mathrm{mL}\) of colour-forming reagent.
- Dilute the mixture to a final volume of \(25.00\ \mathrm{mL}\).
- Measure the blank-corrected absorbance.
Importantly, the horizontal axis of the calibration graph is labelled:
Concentration of the standard solution before the colour-forming reagent is added.
The calibration equation is:
\[ A_{\text{net}}=0.0300c_{\text{pre}}+0.004 \]
where \(c_{\text{pre}}\) is measured in \(\mathrm{mg\,L^{-1}}\).
The validated calibration range is \(1.00\)-\(8.00\ \mathrm{mg\,L^{-1}}\), expressed as the concentration before reagent addition.
An unknown stock solution is prepared for analysis in two stages.
First, \(4.00\ \mathrm{mL}\) of the original stock is diluted to \(20.00\ \mathrm{mL}\).
Next, \(10.00\ \mathrm{mL}\) of this diluted solution undergoes exactly the same colour-development procedure as the calibration standards: \(5.00\ \mathrm{mL}\) of reagent is added, and the mixture is diluted to \(25.00\ \mathrm{mL}\).
The unknown sample gives a raw absorbance of 0.244, while the matching reagent blank gives 0.040. The instrument has not been zeroed.
Two students disagree about the final calculation:
- Student A calculates the concentration from the calibration equation, then multiplies by both \(25.00/10.00\) and \(20.00/4.00\).
- Student B calculates the concentration from the calibration equation and multiplies only by \(20.00/4.00\).
Determine the original concentration of the stock solution, evaluate both approaches, and calculate the actual concentration of \(X\) in the final reaction mixture.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
Solution 5
Student B is correct. The original stock concentration is:
\[ \boxed{33.3\ \mathrm{mg\,L^{-1}}} \]
The key is understanding exactly what concentration the calibration equation returns.
The calibration standards and the unknown both undergo the same colour-development preparation. More importantly, the calibration graph defines concentration before the reagent is added.
The reagent dilution is therefore already accounted for by the calibration relationship.
Step 1: Correct the sample absorbance.
The instrument has not been zeroed, so subtract the blank.
\[ \begin{aligned} A_{\text{net}}&=0.244-0.040\\ &=0.204 \end{aligned} \]
Step 2: Calculate the concentration using the calibration equation.
The supplied equation is:
\[ A_{\text{net}}=0.0300c_{\text{pre}}+0.004 \]
Rearrange:
\[ c_{\text{pre}}=\frac{A_{\text{net}}-0.004}{0.0300} \]
Substitute:
\[ \begin{aligned} c_{\text{pre}}&=\frac{0.204-0.004}{0.0300}\\ &=\frac{0.200}{0.0300}\\ &=6.67\ \mathrm{mg\,L^{-1}} \end{aligned} \]
This is the concentration of the solution before the colour-forming reagent was added.
It is within the validated calibration range.
Step 3: Identify which dilution still needs reversing.
The original stock was diluted by transferring \(4.00\ \mathrm{mL}\) into a final volume of \(20.00\ \mathrm{mL}\).
Therefore:
\[ DF=\frac{20.00}{4.00}=5.00 \]
The calibration equation has already accounted for the common colour-development preparation used by both standards and the unknown.
Only the initial stock dilution remains to be reversed.
Step 4: Calculate the original stock concentration.
\[ \begin{aligned} c_{\text{original}}&=6.67\times5.00\\ &=33.35\ \mathrm{mg\,L^{-1}} \end{aligned} \]
To three significant figures:
\[ \boxed{c_{\text{original}}=33.4\ \mathrm{mg\,L^{-1}}} \]
Student B’s method is correct.
Why Student A’s method fails
Student A multiplies the calibration result by an additional factor:
\[ \frac{25.00}{10.00}=2.50 \]
This factor represents the dilution caused by adding reagent and making the reaction mixture up to \(25.00\ \mathrm{mL}\).
However, the calibration equation already maps the measured absorbance back to the concentration before that reagent was added.
Multiplying by 2.50 again counts the same preparation step twice.
Student A would calculate:
\[ 6.67\times2.50\times5.00 \]
\[ =83.4\ \mathrm{mg\,L^{-1}} \]
This is incorrect.
What is the actual concentration in the final reaction mixture?
There is one more subtle distinction.
Although the calibration equation returns the concentration before reagent addition, the physical concentration inside the final reaction flask is lower.
The reaction mixture contains \(10.00\ \mathrm{mL}\) of the diluted sample in a total volume of \(25.00\ \mathrm{mL}\).
Therefore:
\[ \begin{aligned} c_{\text{reaction}}&=6.67\times\frac{10.00}{25.00}\\ &=2.67\ \mathrm{mg\,L^{-1}} \end{aligned} \]
If we start with this actual reaction-mixture concentration, we can correctly reverse both preparation steps:
\[ 2.67\times2.50\times5.00 \]
\[ =33.4\ \mathrm{mg\,L^{-1}} \]
This agrees with Student B’s result.
The distinction is between the concentration the calibration equation reports and the physical concentration in the final reaction mixture.
The equation’s horizontal-axis label tells you which one you have calculated. That information is just as important as the gradient and intercept.
Once you can interpret a calibration equation correctly, the next challenge is deciding whether the measured signal is genuinely specific to the substance you’re analysing. That’s where interference, sample preparation, and analytical reliability become the next important decisions.