Why Small Pipette Errors Grow in Serial Dilutions
A small pipette error can produce a much larger concentration error across repeated dilutions. Learn why systematic errors compound, how one-off errors behave, and how mixing solutions changes the calculation, with worked HSC Chemistry examples.
A pipette is used to transfer 10.00 mL of solution into a 100.0 mL volumetric flask. The student repeats this dilution six times, each time transferring solution from the previous flask. However, the pipette actually delivers 10.10 mL every time. The error is only 1%. How far off could the final concentration be?
The final concentration will be approximately **6.15% higher than intended**.
This happens because each dilution starts with the solution produced by the previous dilution. If the pipette consistently delivers too much solution, every new flask receives slightly more solute than expected. The concentration error is carried forward and compounds across the series.
The important distinction is between a one-off measurement error and a repeated systematic error. One incorrect transfer produces an error that remains in subsequent dilutions. A pipette that consistently over-delivers produces a larger error at every successive stage.
## Why a small pipette error multiplies
### Follow the solution through the dilution series
Imagine a stock solution containing dissolved potassium nitrate, \(\ce{KNO3}\).
A student:
- Transfers 10.00 mL of stock solution into a 100.0 mL volumetric flask.
- Adds distilled water until the bottom of the meniscus reaches the calibration mark.
- Mixes the solution thoroughly.
- Transfers 10.00 mL from this flask into a second 100.0 mL volumetric flask.
- Repeats the process using each newly prepared solution.
Each transfer takes only a portion of the previous solution. The amount of solute in that transferred portion is then spread through a larger final volume.
[IMAGE: A scientifically accurate serial-dilution apparatus showing a stock-solution flask followed by six labelled 100.0 mL volumetric flasks. Show a pipette transferring solution sequentially from one flask to the next, with distilled water added to each calibration mark. Annotate the intended 10.00 mL transfer and the actual 10.10 mL delivery in a contrasting inset. Show that each flask is thoroughly mixed before the next transfer.]
For an ordinary dilution, the amount of solute in the transferred aliquot remains constant while water is added, provided no chemical reaction or physical loss occurs.
We can describe this using:
\[
c=\frac{n}{V}
\]
where:
- \(c\) is molar concentration, in \(\mathrm{mol\,L^{-1}}\).
- \(n\) is the amount of solute, in mol.
- \(V\) is the total solution volume, in L.
Rearranging gives \(n=cV\). This tells us that the amount of solute in a particular sample depends on its concentration and volume.
For a single dilution, conservation of the solute amount gives:
\[
c_{\mathrm{before}}V_{\mathrm{aliquot}}
=
c_{\mathrm{after}}V_{\mathrm{final}}
\]
Therefore:
\[
c_{\mathrm{after}}
=
c_{\mathrm{before}}
\frac{V_{\mathrm{aliquot}}}{V_{\mathrm{final}}}
\]
The aliquot is the measured portion transferred from the original solution. The final volume is the total volume after adding solvent, not the volume of water added.
This is the relationship behind serial dilution. For the broader calculation method, see [Solution dilution calculations for HSC Chemistry](https://tutorgum.com/learn/resources/hsc-chemistry-solution-dilution-calculations/).
### What happens when the pipette delivers too much?
Return to the six-stage example.
The intended transfer is 10.00 mL, but the pipette actually delivers 10.10 mL. Each flask is still made up to exactly 100.0 mL.
The intended dilution factor is:
\[
DF=\frac{100.0}{10.00}=10.00
\]
The actual dilution factor is:
\[
DF_{\mathrm{actual}}
=
\frac{100.0}{10.10}
=9.901
\]
The actual dilution is slightly weaker than intended. More of the original solution has entered the flask, so its concentration is higher.
The concentration after one stage is therefore:
\[
c_{\mathrm{actual}}
=
c_{\mathrm{before}}\frac{10.10}{100.0}
\]
Compare this with the intended result:
\[
c_{\mathrm{intended}}
=
c_{\mathrm{before}}\frac{10.00}{100.0}
\]
Taking the ratio:
\[
\frac{c_{\mathrm{actual}}}{c_{\mathrm{intended}}}
=
\frac{10.10}{10.00}
=1.01
\]
The new solution is 1% more concentrated than intended.
Here is the important part: the next dilution begins with this already over-concentrated solution. If the same pipette error occurs again, the next flask is another 1% higher relative to its intended concentration.
[IMAGE: A comparison graph showing two six-stage dilution chains. Plot intended concentration and concentration with a repeated 1% pipette over-delivery, using relative concentration rather than extremely small absolute values. Label the intended chain as 1.000 at every stage relative to its target and the actual-to-target ratios as 1.000, 1.010, 1.0201, 1.0303, 1.0406, 1.0510, and 1.0615. Make clear that these are calculated illustrative values, not experimental measurements.]
### The mathematics of repeated error
Define the relative pipette error as:
\[
e=
\frac{V_{\mathrm{actual}}-V_{\mathrm{intended}}}
{V_{\mathrm{intended}}}
\]
A positive value means the pipette over-delivers. A negative value means it under-delivers.
For the example:
\[
e=\frac{10.10-10.00}{10.00}
=0.0100
\]
Therefore, the relative concentration after one incorrectly measured dilution is multiplied by \(1+e=1.01\).
If the same relative error occurs at every stage, then after \(N\) stages:
\[
\boxed{
\frac{c_{\mathrm{actual}}}{c_{\mathrm{intended}}}
=(1+e)^N
}
\]
For six stages:
\[
\begin{aligned}
\frac{c_{\mathrm{actual}}}{c_{\mathrm{intended}}}
&=(1.01)^6\\
&=1.0615
\end{aligned}
\]
The percentage error is:
\[
\begin{aligned}
\%\text{ error}
&=(1.0615-1)\times100\\
&=6.15\%
\end{aligned}
\]
A 1% error repeated six times produces a final concentration error of approximately 6.15%.
Notice that the errors do not simply add to give exactly 6%. Each new error acts on a concentration that is already slightly different from its intended value. This is why the factors multiply.
### A one-off error does not compound in the same way
There is an important qualification.
Suppose the pipette delivers 1% too much solution during only the third dilution. The first two transfers and all subsequent transfers are accurate.
The third flask will be 1% more concentrated than intended. The fourth flask will also be 1% more concentrated than intended, and so will every later flask.
The error is carried forward, but it is not multiplied again.
Why? Every subsequent accurate dilution reduces both the actual concentration and the intended concentration by exactly the same proportion. Their relative difference stays at 1%.
For errors that vary between stages, the general relationship is:
\[
\frac{c_{\mathrm{actual}}}{c_{\mathrm{intended}}}
=
(1+e_1)(1+e_2)\cdots(1+e_N)
\]
Here, \(e_i\) is the relative pipette error at stage \(i\), and \(N\) is the number of dilution stages.
This gives us three different situations:
| Type of error | Effect on the final concentration |
|---|---|
| One isolated 1% over-delivery | Final concentration is 1% too high |
| Repeated 1% over-delivery | Final concentration is multiplied by \(1.01^N\) |
| Random over- and under-delivery | Errors may partly cancel, but the final result becomes less predictable |
A systematic error is particularly troublesome because it consistently pushes the result in the same direction.
For small errors, you can estimate the overall percentage error by multiplying the percentage error per stage by the number of stages. This is an approximation, not the exact calculation.
For example, six repeated 1% errors give an approximate error of 6%, while the exact result is 6.15%.
## How could you test whether the pipette is causing the error?
Suppose a laboratory prepares two serial dilution chains from the same stock solution.
One chain uses a calibrated pipette. The other uses a pipette suspected of delivering slightly too much liquid.
The apparatus follows the same sequence in both cases:
1. A pipette transfers an aliquot from the stock flask.
2. The aliquot enters a volumetric flask.
3. Distilled water is added to the calibration mark.
4. The flask is mixed thoroughly.
5. Another aliquot is transferred into the next flask.
The final concentrations can then be compared using a suitable analytical measurement.
[IMAGE: Experimental comparison of two parallel serial-dilution chains on a laboratory bench. Show identical stock flasks, pipettes, and successive volumetric flasks. One chain uses a calibrated pipette, while the other uses a pipette with a small systematic over-delivery. Show the transfer sequence and final samples being collected for concentration measurement.]
### What should the experiment show?
If the suspected pipette consistently delivers 1% too much, the calculated prediction is that its six-stage dilution chain will finish 6.15% more concentrated than the accurately prepared chain.
These are illustrative calculated results, not measurements from an actual experiment.
| Dilution stage | Actual concentration relative to intended concentration |
|---|---:|
| Stock solution | 1.0000 |
| 1 | 1.0100 |
| 2 | 1.0201 |
| 3 | 1.0303 |
| 4 | 1.0406 |
| 5 | 1.0510 |
| 6 | 1.0615 |
The difference becomes progressively larger because each new transfer acts on the concentration produced by the previous stage.
### How could the experimental design be improved?
Measuring only the final flask would tell us that something went wrong, but not necessarily where the error occurred.
The final difference could also come from:
- Incorrect final volumes in the volumetric flasks.
- Incomplete mixing before taking the next aliquot.
- Liquid remaining in the transfer equipment.
- Contamination or physical loss of solute.
- Errors in the instrument used to measure concentration.
A better investigation would independently check the pipette's delivered volume.
One way to do this is to deliver water into a tared weighing vessel on a balance. The mass of the delivered water can be converted into volume using its density:
\[
V=\frac{m}{\rho}
\]
Here, \(V\) is the delivered volume, \(m\) is the mass of water, and \(\rho\) is the density of water at the measurement temperature.
For a rough school-level estimate, water can be treated as having a density of approximately \(1.00\ \mathrm{g\,mL^{-1}}\). More accurate calibration requires the appropriate temperature-dependent density.
Repeat the measurement several times.
- If the average delivered volume is consistently too high, this suggests a systematic bias.
- If individual measurements vary above and below the target, this suggests random variation.
Checking the pipette independently helps separate its error from problems with the volumetric flasks or mixing procedure.
The central experimental conclusion is that a small, repeated measurement bias can produce a much larger difference in the final concentration. However, the calculated model must be tested against the actual equipment and experimental conditions.
## A dilution is not the same as mixing two solutions
There is another condition that matters when applying conservation of solute.
A simple dilution starts with one solution and adds solvent. The solvent does not contribute any additional amount of the solute being measured.
But what happens if you combine two solutions that both contain the same solute?
You now have two sources of solute. You cannot calculate the final concentration by following only one of them.
[IMAGE: Two-panel chemistry illustration. Left panel: one flask containing a coloured solute solution is transferred into a larger volumetric flask and diluted with pure water. Right panel: two separate flasks containing the same dissolved solute at different concentrations are poured into one receiving flask. Show particles from both sources contributing to the final mixture. Clearly distinguish solvent addition from combining two solute-containing solutions.]
### Comparing the two situations
| Situation | What happens to the solute amount? | Calculation |
|---|---|---|
| One solution diluted with pure water | Amount in the transferred aliquot stays constant | \(c_1V_1=c_2V_2\) |
| Two solutions containing the same solute are mixed | Amounts from both solutions are added | \(n_{\mathrm{total}}=n_1+n_2\) |
| Two different, nonreacting solutes are mixed | Each solute amount is tracked separately | Calculate each concentration independently |
| Solutions react when mixed | The amounts of reacting species change | Use reaction stoichiometry first |
For two solutions containing the same solute, assuming no reaction or physical loss:
\[
n_{\mathrm{total}}=c_AV_A+c_BV_B
\]
The final concentration is:
\[
c_{\mathrm{final}}
=
\frac{c_AV_A+c_BV_B}{V_{\mathrm{final}}}
\]
Here, \(c_A\) and \(c_B\) are the initial concentrations of the two solutions, \(V_A\) and \(V_B\) are their transferred volumes, and \(V_{\mathrm{final}}\) is the total final volume.
Notice that you add the amounts of solute, not the concentrations.
### What if a reaction occurs?
Conservation of the amount of a solute is not a universal rule.
It applies only when the substance being tracked is not consumed, produced, or physically removed.
For example, mixing hydrochloric acid and sodium hydroxide causes a neutralisation reaction:
\[
\ce{H3O+ + OH- -> 2H2O}
\]
The hydronium ions are consumed to produce water. Their original amount is therefore not conserved.
You must calculate the reaction using mole ratios before calculating the concentration of any remaining hydronium ions.
The same principle applies if a reaction produces a precipitate or changes one dissolved species into another.
This is why you should identify what is happening before applying a dilution equation. A pipette error can affect the amount transferred, but a chemical reaction can change the identity and amount of the substance being tracked.
## Worked examples
### Worked example: A single pipette error in a two-stage dilution
A student prepares a potassium nitrate solution with an initial concentration of \(0.200\ \mathrm{mol\,L^{-1}}\).
They:
- Transfer 10.00 mL into a 100.0 mL volumetric flask using an accurate pipette.
- Transfer a second aliquot, intended to be 5.00 mL, into a 50.0 mL volumetric flask.
- Discover that the second transfer actually delivered 5.05 mL.
Assume both final volumes are exact, the solutions are thoroughly mixed, and no solute is lost or reacts.
Calculate the intended final concentration, the actual final concentration, and the percentage error.
[IMAGE: Two-stage potassium nitrate dilution apparatus. Show a stock flask labelled 0.200 mol L^-1, a first volumetric flask labelled 100.0 mL, and a second labelled 50.0 mL. Annotate the first transfer as 10.00 mL and the second as an intended 5.00 mL but actual 5.05 mL. Do not show calculated concentrations.]
**Step 1**: Calculate the concentration after the first dilution.
The first transfer is accurate, so:
\[
\begin{aligned}
c_1&=c_0\frac{V_1}{V_2}\\
&=0.200\times\frac{10.00}{100.0}\\
&=0.0200\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 2**: Calculate the intended final concentration.
If the second transfer were exactly 5.00 mL:
\[
\begin{aligned}
c_{\mathrm{intended}}
&=0.0200\times\frac{5.00}{50.0}\\
&=0.00200\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 3**: Calculate the actual final concentration.
The pipette delivers 5.05 mL rather than 5.00 mL.
\[
\begin{aligned}
c_{\mathrm{actual}}
&=0.0200\times\frac{5.05}{50.0}\\
&=0.00202\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 4**: Calculate the percentage error.
Use:
\[
\%\text{ error}
=
\frac{c_{\mathrm{actual}}-c_{\mathrm{intended}}}
{c_{\mathrm{intended}}}
\times100
\]
Substituting:
\[
\begin{aligned}
\%\text{ error}
&=\frac{0.00202-0.00200}{0.00200}\times100\\
&=1.00\%
\end{aligned}
\]
The actual concentration is \(0.00202\ \mathrm{mol\,L^{-1}}\), which is 1.00% higher than intended.
The key observation is that this was a one-off error. The first dilution was accurate, and only the second transfer was incorrect. There is no reason to raise 1.01 to the power of two.
### Worked example: Mixing two solutions when one pipette over-delivers
A student combines two glucose solutions:
- 25.0 mL of \(0.0800\ \mathrm{mol\,L^{-1}}\) glucose solution.
- 15.0 mL of \(0.120\ \mathrm{mol\,L^{-1}}\) glucose solution.
The pipette used for the first solution over-delivers by 2.00%. The second transfer is accurate.
The student mixes both solutions and adds distilled water until the final volume reaches 100.0 mL.
Calculate the actual final glucose concentration and its percentage error compared with the intended concentration.
Assume no reaction or physical loss.
[IMAGE: Two separate glucose solution flasks feeding into a single 100.0 mL volumetric flask. Label the first source 0.0800 mol L^-1 and intended transfer 25.0 mL, with a 2.00% pipette over-delivery annotation. Label the second source 0.120 mol L^-1 and accurate transfer 15.0 mL. Show distilled water being added to the final calibration mark.]
**Step 1**: Calculate the actual volume transferred from the first solution.
A 2.00% over-delivery means the pipette transfers 102.00% of its intended volume.
\[
\begin{aligned}
V_A&=25.0\times1.0200\\
&=25.5\ \mathrm{mL}\\
&=0.0255\ \mathrm{L}
\end{aligned}
\]
**Step 2**: Calculate the amount of glucose contributed by the first solution.
Using \(n=cV\):
\[
\begin{aligned}
n_A&=0.0800\times0.0255\\
&=0.00204\ \mathrm{mol}
\end{aligned}
\]
**Step 3**: Calculate the amount of glucose contributed by the second solution.
The second transfer is accurate:
\[
15.0\ \mathrm{mL}=0.0150\ \mathrm{L}
\]
Therefore:
\[
\begin{aligned}
n_B&=0.120\times0.0150\\
&=0.00180\ \mathrm{mol}
\end{aligned}
\]
**Step 4**: Add the amounts of glucose.
Both solutions contain glucose, and no reaction occurs.
\[
\begin{aligned}
n_{\mathrm{total}}
&=n_A+n_B\\
&=0.00204+0.00180\\
&=0.00384\ \mathrm{mol}
\end{aligned}
\]
**Step 5**: Calculate the actual final concentration.
The final volume is 100.0 mL, or 0.1000 L.
\[
\begin{aligned}
c_{\mathrm{actual}}
&=\frac{0.00384}{0.1000}\\
&=0.0384\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 6**: Compare this with the intended concentration.
If the first pipette had delivered exactly 25.0 mL:
\[
n_{A,\mathrm{intended}}
=
0.0800\times0.0250
=
0.00200\ \mathrm{mol}
\]
The intended total amount is:
\[
n_{\mathrm{intended}}
=
0.00200+0.00180
=
0.00380\ \mathrm{mol}
\]
Therefore:
\[
c_{\mathrm{intended}}
=
\frac{0.00380}{0.1000}
=
0.0380\ \mathrm{mol\,L^{-1}}
\]
The percentage error is:
\[
\begin{aligned}
\%\text{ error}
&=\frac{0.0384-0.0380}{0.0380}\times100\\
&=1.05\%
\end{aligned}
\]
The final concentration is \(0.0384\ \mathrm{mol\,L^{-1}}\), approximately 1.05% higher than intended.
Notice that the final error is not 2.00%. Only one of the two glucose contributions was affected by the pipette error. Mixing the solutions changes how strongly that error affects the final result.
## Questions and solutions
Assume all solutions are homogeneous and all stated final volumes are exact. Unless otherwise stated, no solute is lost and no chemical reaction occurs.
### Question 1
A student prepares a diluted potassium nitrate solution using a stock solution of concentration \(0.120\ \mathrm{mol\,L^{-1}}\).
They intend to transfer 20.00 mL of stock solution into a 100.0 mL volumetric flask.
However, the pipette actually delivers 19.80 mL.
Calculate:
a) The intended final concentration.
b) The actual final concentration.
c) The percentage error, including whether the result is higher or lower than intended.
[IMAGE: A single volumetric pipette transferring an intended 20.00 mL aliquot from a stock flask into a 100.0 mL volumetric flask. Annotate the actual delivery as 19.80 mL. Include the stock concentration of 0.120 mol L^-1. Do not show the final concentration.]
### Solution 1
The actual final concentration is \(0.02376\ \mathrm{mol\,L^{-1}}\), which is 1.00% lower than intended.
**Step 1**: Calculate the intended final concentration.
\[
\begin{aligned}
c_{\mathrm{intended}}
&=0.120\times\frac{20.00}{100.0}\\
&=0.0240\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 2**: Calculate the actual final concentration.
The pipette delivers only 19.80 mL.
\[
\begin{aligned}
c_{\mathrm{actual}}
&=0.120\times\frac{19.80}{100.0}\\
&=0.02376\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 3**: Calculate the percentage error.
\[
\begin{aligned}
\%\text{ error}
&=\frac{0.02376-0.0240}{0.0240}\times100\\
&=-1.00\%
\end{aligned}
\]
The negative sign indicates that the actual concentration is lower than intended.
This makes physical sense. Less stock solution entered the flask, so less potassium nitrate was transferred into the same final volume.
### Question 2
A student performs six successive dilutions using a potassium nitrate stock solution with an initial concentration of \(0.250\ \mathrm{mol\,L^{-1}}\).
At every stage:
- The intended aliquot is 5.00 mL.
- The final solution volume is 50.0 mL.
- The same pipette is used.
The pipette consistently delivers 5.04 mL rather than 5.00 mL.
Calculate:
a) The intended final concentration after six dilutions.
b) The actual final concentration.
c) The percentage error.
Explain why multiplying 0.8% by six gives an approximation rather than the exact percentage error.
[IMAGE: Six-stage serial dilution chain using six 50.0 mL volumetric flasks. Show a transfer from each flask to the next, with every transfer labelled 5.04 mL actual instead of 5.00 mL intended. Show the initial stock concentration as 0.250 mol L^-1. Do not show the final calculated concentration.]
### Solution 2
The actual final concentration is approximately \(2.62\times10^{-7}\ \mathrm{mol\,L^{-1}}\), which is 4.90% higher than intended.
**Step 1**: Calculate the intended concentration after one dilution.
\[
c_1
=
0.250\times\frac{5.00}{50.0}
=
0.0250\ \mathrm{mol\,L^{-1}}
\]
Each dilution reduces the concentration to one-tenth of its previous value.
**Step 2**: Calculate the intended final concentration after six stages.
\[
\begin{aligned}
c_{\mathrm{intended}}
&=0.250\left(\frac{5.00}{50.0}\right)^6\\
&=0.250(0.100)^6\\
&=2.50\times10^{-7}\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 3**: Calculate the actual concentration.
Each transfer is 5.04 mL rather than 5.00 mL.
\[
\begin{aligned}
c_{\mathrm{actual}}
&=0.250\left(\frac{5.04}{50.0}\right)^6\\
&=0.250(0.1008)^6\\
&=2.6224\times10^{-7}\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 4**: Calculate the percentage error.
It is useful to compare the actual and intended concentrations directly:
\[
\begin{aligned}
\%\text{ error}
&=\left[(1.008)^6-1\right]\times100\\
&=4.90\%
\end{aligned}
\]
The final concentration is approximately 4.90% higher than intended.
Why isn't the answer exactly 4.8%?
Multiplying \(0.8\%\) by six assumes each error acts on the original intended concentration. In reality, each error acts on the concentration already affected by the previous errors.
The exact calculation therefore uses \(1.008^6\), not \(1+6(0.008)\).
### Question 3
A student prepares four successive dilutions from a stock solution of concentration \(0.400\ \mathrm{mol\,L^{-1}}\).
Each dilution is intended to use:
- A 10.00 mL aliquot.
- A final volume of 100.0 mL.
The pipette is accurate for the first, second, and fourth transfers. However, during the third transfer, it delivers only 9.80 mL.
Calculate the actual concentration after each stage and the final percentage error compared with the intended concentration.
A student argues that because there are four dilution stages, the 2.00% error should become an 8.00% error. Explain why this reasoning is incorrect.
[IMAGE: Four sequential 100.0 mL volumetric flasks connected by pipette transfers. Label the first, second, and fourth aliquots as 10.00 mL. Label only the third aliquot as 9.80 mL actual instead of 10.00 mL intended. Show the original concentration as 0.400 mol L^-1.]
### Solution 3
The final concentration is \(3.92\times10^{-5}\ \mathrm{mol\,L^{-1}}\), which is 2.00% lower than intended, not 8.00% lower.
**Step 1**: Calculate the concentration after the first two accurate dilutions.
First dilution:
\[
\begin{aligned}
c_1&=0.400\times\frac{10.00}{100.0}\\
&=0.0400\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
Second dilution:
\[
\begin{aligned}
c_2&=0.0400\times\frac{10.00}{100.0}\\
&=0.00400\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
The first two stages are accurate, so there is no concentration error yet.
**Step 2**: Calculate the concentration after the third dilution.
The pipette delivers 9.80 mL instead of 10.00 mL.
\[
\begin{aligned}
c_3&=0.00400\times\frac{9.80}{100.0}\\
&=0.000392\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
The third solution is 2.00% less concentrated than intended.
**Step 3**: Calculate the concentration after the fourth dilution.
The fourth transfer is accurate.
\[
\begin{aligned}
c_4&=0.000392\times\frac{10.00}{100.0}\\
&=0.0000392\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
The fourth dilution reduces the concentration of both the actual and intended solutions by the same factor.
**Step 4**: Compare the final concentration with the intended result.
The intended final concentration is:
\[
\begin{aligned}
c_{\mathrm{intended}}
&=0.400(0.100)^4\\
&=0.0000400\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
Therefore:
\[
\begin{aligned}
\%\text{ error}
&=\frac{0.0000392-0.0000400}{0.0000400}
\times100\\
&=-2.00\%
\end{aligned}
\]
The final solution remains 2.00% below its intended concentration.
The tempting approach is to multiply the error by all four stages. However, the pipette is only inaccurate at the third stage. The first two stages introduce no error, and the fourth accurately preserves the existing relative difference.
### Question 4
A student combines three solutions in one volumetric flask:
- 20.0 mL of \(0.100\ \mathrm{mol\,L^{-1}}\) glucose solution. The pipette over-delivers by 1.00%.
- 30.0 mL of \(0.0600\ \mathrm{mol\,L^{-1}}\) glucose solution. This transfer is accurate.
- 10.0 mL of \(0.0500\ \mathrm{mol\,L^{-1}}\) sucrose solution. This transfer is accurate.
The student adds distilled water until the final volume reaches 100.0 mL.
Assuming glucose and sucrose do not react, calculate:
a) The final glucose concentration.
b) The final sucrose concentration.
c) Why is it incorrect to add the total moles of glucose and sucrose together and treat them as one solute?
[IMAGE: Three source flasks feeding into one 100.0 mL volumetric flask. Show two glucose sources labelled 0.100 mol L^-1 and 0.0600 mol L^-1, and a sucrose source labelled 0.0500 mol L^-1. Label the glucose transfers as intended 20.0 mL with 1.00% over-delivery, accurate 30.0 mL, and accurate 10.0 mL sucrose. Show the final flask being made up to its 100.0 mL mark. Do not include calculated concentrations.]
### Solution 4
The final glucose concentration is \(0.0382\ \mathrm{mol\,L^{-1}}\), and the final sucrose concentration is \(0.00500\ \mathrm{mol\,L^{-1}}\).
The two glucose contributions must be added together, but sucrose must be tracked separately.
**Step 1**: Calculate the actual volume of the first glucose solution transferred.
\[
\begin{aligned}
V_A&=20.0\times1.0100\\
&=20.2\ \mathrm{mL}\\
&=0.0202\ \mathrm{L}
\end{aligned}
\]
**Step 2**: Calculate the total amount of glucose.
The first glucose solution contributes:
\[
\begin{aligned}
n_A&=0.100\times0.0202\\
&=0.00202\ \mathrm{mol}
\end{aligned}
\]
The second glucose solution contributes:
\[
\begin{aligned}
n_B&=0.0600\times0.0300\\
&=0.00180\ \mathrm{mol}
\end{aligned}
\]
Add both contributions:
\[
\begin{aligned}
n_{\mathrm{glucose}}
&=0.00202+0.00180\\
&=0.00382\ \mathrm{mol}
\end{aligned}
\]
**Step 3**: Calculate the final glucose concentration.
\[
\begin{aligned}
c_{\mathrm{glucose}}
&=\frac{0.00382}{0.1000}\\
&=0.0382\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 4**: Calculate the amount of sucrose.
\[
\begin{aligned}
n_{\mathrm{sucrose}}
&=0.0500\times0.0100\\
&=0.000500\ \mathrm{mol}
\end{aligned}
\]
**Step 5**: Calculate the final sucrose concentration.
\[
\begin{aligned}
c_{\mathrm{sucrose}}
&=\frac{0.000500}{0.1000}\\
&=0.00500\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
The two substances must be tracked separately because glucose remains glucose and sucrose remains sucrose. Their amounts cannot be combined to produce a meaningful concentration of a single chemical species.
The important distinction is that both glucose solutions contribute to the final glucose amount, whereas the sucrose solution contributes to a separate amount of sucrose.
### Question 5
A laboratory prepares two branches of a dilution experiment using a stock dye solution with a concentration of \(0.200\ \mathrm{mol\,L^{-1}}\).
The branches are prepared as follows.
**Branch A:**
- Transfer 10.00 mL of stock solution into a 100.0 mL volumetric flask.
- Transfer 10.00 mL of the first diluted solution into another 100.0 mL volumetric flask.
- Use the same suspect pipette for both transfers.
The pipette has an unknown systematic relative error, \(e\). It delivers the same percentage above or below its intended volume during both transfers.
**Branch B:**
- Transfer 2.00 mL of the original stock solution into a 100.0 mL volumetric flask.
- Use an accurately calibrated pipette.
The student then transfers 25.0 mL from Branch A and 25.0 mL from Branch B into the same container. The mixture is diluted to a final volume of 100.0 mL.
The final dye concentration is measured as:
\[
0.0015202\ \mathrm{mol\,L^{-1}}
\]
Assume:
- Both branches are homogeneous before aliquots are removed.
- No reaction or physical loss occurs.
- All final flask volumes are exact.
- The two branches contain the same solute.
- The measured final concentration is exact for this question.
Calculate the percentage error of the suspect pipette.
A student argues that because the final concentration is only about 1.35% higher than the expected concentration, the pipette must over-deliver by approximately 1.35% at each transfer.
Explain why this conclusion is incorrect.
[IMAGE: A branched serial-dilution apparatus. Start with one stock dye flask labelled 0.200 mol L^-1. Branch A passes through two successive 100.0 mL volumetric flasks, with each intended transfer labelled 10.00 mL and both using the same suspect pipette. Branch B passes through one 100.0 mL flask using an accurate 2.00 mL transfer. Show 25.0 mL from each branch entering a common mixing vessel, followed by dilution to a final volume of 100.0 mL. Show the measured final concentration of 0.0015202 mol L^-1 as supplied data. Do not show the unknown pipette error or its solution.]
### Solution 5
The suspect pipette over-delivers by **2.00% at each transfer**.
The final mixture is only 1.35% more concentrated than intended because it contains solute from two branches, and only one branch is affected by the suspect pipette.
**Step 1**: Express the concentration of Branch A in terms of the unknown error.
Let \(e\) be the relative pipette error expressed as a decimal.
For each transfer, the actual volume is:
\[
V_{\mathrm{actual}}=10.00(1+e)
\]
The first dilution produces:
\[
c_A'
=
0.200\frac{10.00(1+e)}{100.0}
\]
The second dilution uses the same suspect pipette, so:
\[
\begin{aligned}
c_A
&=c_A'\frac{10.00(1+e)}{100.0}\\
&=0.200(0.100)^2(1+e)^2\\
&=0.00200(1+e)^2
\end{aligned}
\]
Therefore, the concentration of Branch A depends on the square of the repeated pipette error.
**Step 2**: Calculate the concentration of Branch B.
Branch B uses an accurate pipette:
\[
\begin{aligned}
c_B
&=0.200\times\frac{2.00}{100.0}\\
&=0.00400\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
**Step 3**: Calculate the final concentration after mixing.
The mixture contains equal volumes from both branches.
The amount of dye transferred from each branch is:
\[
n_A=c_A\times0.0250
\]
\[
n_B=c_B\times0.0250
\]
The total amount is:
\[
n_{\mathrm{total}}
=
0.0250(c_A+c_B)
\]
The final volume is 0.1000 L.
Therefore:
\[
\begin{aligned}
c_{\mathrm{final}}
&=\frac{0.0250(c_A+c_B)}{0.1000}\\
&=0.250(c_A+c_B)
\end{aligned}
\]
Substitute the expressions for both branches:
\[
0.0015202
=
0.250\left[
0.00200(1+e)^2+0.00400
\right]
\]
**Step 4**: Rearrange to find the unknown error.
Divide both sides by 0.250:
\[
0.0060808
=
0.00200(1+e)^2+0.00400
\]
Subtract 0.00400:
\[
0.0020808=0.00200(1+e)^2
\]
Divide by 0.00200:
\[
(1+e)^2=1.0404
\]
Taking the positive square root is appropriate because the delivered volume must be positive:
\[
1+e=\sqrt{1.0404}=1.020
\]
Therefore:
\[
e=0.0200
\]
Converting to a percentage:
\[
e=2.00\%
\]
The suspect pipette over-delivers by 2.00% on each transfer.
**Step 5**: Explain why the final mixture is only 1.35% too concentrated.
First, calculate the intended concentration of Branch A:
\[
c_{A,\mathrm{intended}}=0.00200\ \mathrm{mol\,L^{-1}}
\]
Branch B is:
\[
c_B=0.00400\ \mathrm{mol\,L^{-1}}
\]
The intended final concentration is:
\[
\begin{aligned}
c_{\mathrm{intended}}
&=0.250(0.00200+0.00400)\\
&=0.001500\ \mathrm{mol\,L^{-1}}
\end{aligned}
\]
Compare this with the measured concentration:
\[
\begin{aligned}
\%\text{ error}
&=\frac{0.0015202-0.001500}{0.001500}
\times100\\
&=1.35\%
\end{aligned}
\]
The final mixture is 1.35% more concentrated than intended, even though the suspect pipette over-delivers by 2.00% on each of its two transfers.
The tempting mistake is to assume that the percentage error in the final mixture must equal the percentage error of the pipette.
That ignores how the mixture was prepared.
Branch A contains the accumulated error from two transfers. Branch B contains no pipette error. When both branches are mixed, the accurate contribution from Branch B reduces the relative effect of the error in Branch A.
This is the conceptual difference between following one solution through successive dilutions and combining two solutions that have separate preparation histories.
## Where to go next
When calculating serial dilutions, first identify whether you are following one homogeneous solution through successive transfers or combining multiple solutions.
For a single solution, follow the aliquot and account for every dilution factor. For a mixture, calculate the amount contributed by each source before dividing by the final volume. If a reaction occurs, perform the stoichiometric calculation first.
To extend these calculations to longer dilution chains and overall dilution factors, see [How to Calculate a Serial-Dilution Factor in HSC Chemistry](https://tutorgum.com/learn/resources/hsc-chemistry-how-to-calculate-a-serial-dilution-factor/).