Serial Dilution Uncertainty: How Measurement Errors Accumulate

Learn how pipette and volumetric-flask uncertainties affect concentration through repeated dilutions. Explore conservative uncertainty estimates, independent random errors, and systematic calibration errors through worked examples.

Four successive 1:10 dilutions turn a solution with a concentration of 0.1000 mol L⁻¹ into one with a nominal concentration of 0.00001000 mol L⁻¹. The calculation is straightforward. But how certain can you be that the final concentration is correct?

Every dilution involves measuring volumes. If the pipette delivers slightly too much solution, or the volumetric flask holds slightly more than its stated capacity, the calculated concentration will be affected. And because each dilution uses the previous solution, uncertainty can travel through the entire chain.

Predict what happens after four dilution steps. Does the relative uncertainty become four times larger, increase according to a square-root relationship, or stay the same?

The answer depends on the type of measurement uncertainty and, particularly, whether errors at different stages are independent.

Scope: This article goes beyond the required HSC Chemistry syllabus. It assumes knowledge from Year 11 Chemistry, 2025 syllabus, Quantitative chemistry, as its foundation. The year is a prerequisite entry point, not an assessment requirement for this extension. The 2025 syllabus is scheduled for implementation from 2028. Its practical investigation of the law of conservation of mass provides the measurement foundation for the ideas developed here.

Five volumetric flasks arranged on a laboratory bench show a 0.1000 mol L⁻¹ stock solution followed by four successive 1:10 dilutions. Pipettes illustrate transfers between flasks, with progressively lighter blue solutions.
After each 1:10 transfer, the receiving flask is made up to its calibration mark; the fading tint is illustrative, not quantitative.

Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.

01The measurement problem: when does a difference actually matter?

Consider a practical investigation into the law of conservation of mass.

A reaction produces a gas. A student places the reactants in an open vessel on a balance, records the initial mass, allows the reaction to occur, and records the final mass.

The final reading is lower.

Has mass disappeared?

It might seem reasonable to conclude that the reaction has lost mass. However, there is another possibility: the gas has escaped from the vessel.

The balance measures the material remaining on its weighing platform. It cannot measure gas that has already escaped into the surrounding air.

The original experimental arrangement therefore has a limitation. It does not control whether all the products remain within the system.

A revised experiment uses a closed reaction system.

The apparatus is arranged so that:

  1. The reactants are kept separate inside a sealed vessel.
  2. The entire sealed vessel is placed on a balance.
  3. The initial mass is recorded.
  4. The reaction is initiated without opening the vessel.
  5. The final mass of the same sealed system is recorded.

The prediction is that the total mass will remain unchanged because no matter enters or leaves the closed system.

In a real investigation, the two balance readings might not be identical. That does not automatically disprove conservation of mass. The difference must be considered alongside the uncertainty of the measurements.

For example, suppose a hypothetical investigation produces these readings:

MeasurementMass
Before reaction\(42.36 \pm 0.01\) g
After reaction\(42.35 \pm 0.01\) g

The measured mass difference is:

\[ \Delta m=42.35-42.36 \]

\[ \Delta m=-0.01\text{ g} \]

Using a conservative uncertainty model for subtraction, the absolute uncertainties are added:

\[ u(\Delta m)=0.01+0.01=0.02\text{ g} \]

Therefore:

\[ \Delta m=-0.01\pm0.02\text{ g} \]

The measured difference is smaller than the uncertainty. The result is consistent with conservation of mass.

This is an important experimental principle: a difference between two measurements is not necessarily evidence of a physical difference.

The same reasoning applies when measuring solution concentrations. A calculated concentration is not automatically exact simply because the calculator displays several decimal places.

02From one dilution to a chain of dilutions

A serial dilution involves preparing a diluted solution and then using that solution to prepare another dilution.

Picture a concentrated stock solution on the left. A pipette transfers a measured aliquot into a volumetric flask. Solvent is added until the solution reaches the flask’s calibration mark.

The flask is mixed thoroughly. A new aliquot is then transferred into another volumetric flask.

The process continues.

The key principle is that dilution changes concentration without changing the amount of dissolved solute in the transferred aliquot.

For one dilution:

\[ C_1V_1=C_2V_2 \]

where:

  • \(C_1\) is the initial concentration.
  • \(V_1\) is the volume of solution transferred.
  • \(C_2\) is the final concentration.
  • \(V_2\) is the total final solution volume.

Rearranging:

\[ C_2=C_1\frac{V_1}{V_2} \]

For this article, we will use slightly different notation to make repeated dilutions easier to follow.

Let:

  • \(C_0\) be the initial stock concentration.
  • \(C_i\) be the concentration after dilution stage \(i\).
  • \(V_{a,i}\) be the aliquot volume transferred at stage \(i\).
  • \(V_{f,i}\) be the total final volume at stage \(i\).
  • \(d_i\) be the dilution ratio at stage \(i\).
  • \(n\) be the total number of dilution stages.

The dilution ratio is:

\[ d_i=\frac{V_{a,i}}{V_{f,i}} \]

Therefore:

\[ C_i=C_{i-1}d_i \]

After several stages:

\[ C_n=C_0\prod_{i=1}^{n}d_i \]

The product symbol means that all the individual dilution ratios are multiplied together.

If every stage uses the same dilution ratio, the expression becomes:

\[ C_n=C_0d^n \]

For example, four successive 1:10 dilutions give:

\[ C_4=C_0(0.1)^4 \]

\[ C_4=0.0001C_0 \]

The final concentration is one ten-thousandth of the original concentration.

For a refresher on calculating the nominal concentration, see How to Calculate a Serial-Dilution Factor in HSC Chemistry.

One detail matters here: \(V_f\) is the total final solution volume, not the volume of solvent added. Confusing these quantities produces an incorrect dilution ratio.

03Why uncertainty travels through a serial dilution

Suppose a pipette is intended to transfer 10.00 mL.

Its stated measurement uncertainty is ±0.02 mL.

The actual volume might be slightly higher or lower than the nominal value. Consequently, the concentration in the receiving flask might also be slightly different from the calculated concentration.

Now consider the next dilution.

Its aliquot is taken from the first flask, which already has an uncertain concentration. The next dilution introduces additional volume-measurement uncertainty.

The third dilution inherits uncertainty from the second, and so on.

There are two different ideas to keep separate:

  • Absolute uncertainty: the uncertainty expressed in the same units as the measured quantity.
  • Relative uncertainty: the uncertainty expressed as a fraction of the measured quantity.

Relative uncertainty is particularly useful for serial dilutions because concentration is calculated by multiplying and dividing quantities.

1. Conservative uncertainty: adding percentage uncertainties

For a measured quantity \(x\), let \(\Delta x\) represent its absolute uncertainty.

Its relative uncertainty is:

\[ r_x=\frac{\Delta x}{x} \]

To express this as a percentage:

\[ \%\text{ uncertainty} = \frac{\Delta x}{x}\times100 \]

For multiplication and division, a commonly used school-level approximation is to add the percentage uncertainties.

For a single dilution:

\[ C_2=C_1\frac{V_a}{V_f} \]

The approximate relative uncertainty is therefore:

\[ r_{C_2}\approx r_{C_1}+r_{V_a}+r_{V_f} \]

Notice that the uncertainty in the final volume is also added.

This is because the final concentration depends on dividing by the final volume. A smaller-than-expected final volume increases the concentration, while a larger-than-expected volume decreases it.

For a sequence of dilutions, the conservative relative uncertainty becomes:

\[ \boxed{ r_{C_n}\approx r_{C_0}+ \sum_{i=1}^{n}(r_{V_{a,i}}+r_{V_{f,i}}) } \]

If every stage has the same volume uncertainties:

\[ \boxed{ r_{C_n}\approx r_{C_0}+n(r_{V_a}+r_{V_f}) } \]

This is a first-order approximation for small uncertainties.

It is useful when the stated ± values represent maximum uncertainty bounds and you want a conservative estimate. It assumes that the individual deviations could all combine in the direction that produces the largest overall deviation.

For example, if every dilution stage contributes a maximum relative uncertainty of 0.30%, three stages contribute approximately 0.90%.

The important point is that the initial stock uncertainty is also carried through the chain.

2. Independent random uncertainty: the square-root relationship

There is another way to combine uncertainties.

Suppose the uncertainty in each volume measurement represents an independent random variation, rather than a maximum tolerance.

Independent random variations can push the result in either direction. They do not necessarily all increase the final concentration or all decrease it.

For independent standard uncertainties, we combine their squares and take the square root. This is called the root-sum-of-squares method.

Let \(u(x)\) represent the standard uncertainty of a quantity \(x\).

Its relative standard uncertainty is:

\[ u_r(x)=\frac{u(x)}{x} \]

For one dilution:

\[ u_r(C_2)\approx \sqrt{ u_r(C_1)^2+ u_r(V_a)^2+ u_r(V_f)^2 } \]

For multiple independent dilution stages:

\[ \boxed{ u_r(C_n)\approx \sqrt{ u_r(C_0)^2+ \sum_{i=1}^{n} \left[ u_r(V_{a,i})^2+ u_r(V_{f,i})^2 \right] } } \]

If every stage has the same independent relative standard uncertainty in its dilution ratio, \(u_r(d)\), this simplifies to:

\[ \boxed{ u_r(C_n)\approx \sqrt{u_r(C_0)^2+nu_r(d)^2} } \]

Why the square root?

Independent random uncertainties combine through their variances. Their standard deviations do not simply add.

This is the same principle used in the formal law of propagation of uncertainty described by the National Institute of Standards and Technology (NIST). The full treatment also accounts for correlations between measurements.

An important restriction: you should not automatically use this method with instrument tolerance limits. The ± values must represent standard uncertainties, or be converted using a justified statistical assumption.

3. The important exception: repeated systematic errors

Here is where serial dilution uncertainty becomes more interesting.

Imagine using the same pipette for every dilution. The pipette consistently delivers slightly more than its stated volume because of a calibration offset.

Each transfer is affected by the same underlying bias.

The errors are now correlated. They are not independent random variations.

Think of a stopwatch that consistently runs slightly fast. Repeating a timing measurement does not remove the clock’s calibration problem. The readings may be very consistent, but they can still be systematically wrong.

A similar issue occurs with a pipette.

If the same relative calibration uncertainty affects every dilution ratio, its contribution grows approximately in proportion to the number of stages:

\[ u_{r,\mathrm{common}}(C_n) \approx n\,u_{r,\mathrm{cal}}(d) \]

Here, \(u_{r,\mathrm{cal}}(d)\) is the common relative standard uncertainty associated with the dilution ratio.

This is different from independent random uncertainty, which grows according to the square root of the number of stages.

Type of uncertaintyContribution after \(n\) identical stages
Conservative maximum boundsApproximately \(n\) times the per-stage relative bound
Independent random standard uncertainty\(\sqrt{n}\) times the per-stage relative standard uncertainty
Shared calibration uncertainty\(n\) times the common relative standard uncertainty

A known calibration bias should be corrected rather than simply included as an unknown uncertainty. The table describes uncertainty that remains after any known corrections.

The distinction between random and systematic uncertainty is essential. Treating every measurement error as independent can significantly underestimate the uncertainty in a serial dilution.

04Worked example: three successive dilutions with instrument tolerances

A student begins with a stock solution of concentration \(0.1000\ \mathrm{mol\,L^{-1}}\).

The student performs three successive dilutions. At each stage:

  • The pipette transfers \(10.00\pm0.02\) mL.
  • The volumetric flask has a final volume of \(100.0\pm0.1\) mL.

Assume the stock concentration uncertainty is negligible. Treat the stated volume uncertainties as maximum tolerance bounds.

Calculate the final concentration and its conservative relative uncertainty.

Step 1

\[ d=\frac{10.00}{100.0}=0.1000 \]

Each stage reduces the concentration to one-tenth of its previous value.

Step 2

There are three identical dilution stages:

\[ C_3=C_0d^3 \]

Substitute:

\[ C_3=0.1000(0.1000)^3 \]

\[ C_3=1.000\times10^{-4}\ \mathrm{mol\,L^{-1}} \]

Step 3

For the pipette:

\[ r_{V_a}=\frac{0.02}{10.00}=0.002 \]

\[ r_{V_a}=0.20\% \]

For the volumetric flask:

\[ r_{V_f}=\frac{0.1}{100.0}=0.001 \]

\[ r_{V_f}=0.10\% \]

Step 4

\[ r_d=0.20\%+0.10\% \]

\[ r_d=0.30\% \]

There are three stages, so:

\[ r_{C_3}=3(0.30\%)=0.90\% \]

Step 5

Convert 0.90% into a decimal:

\[ 0.90\%=0.0090 \]

Multiply by the final concentration:

\[ \Delta C_3=0.0090(1.000\times10^{-4}) \]

\[ \Delta C_3=9.0\times10^{-7}\ \mathrm{mol\,L^{-1}} \]

The final result is:

\[ \boxed{ C_3=(1.000\pm0.009)\times10^{-4} \ \mathrm{mol\,L^{-1}} } \]

The nominal concentration is \(1.000\times10^{-4}\ \mathrm{mol\,L^{-1}}\), but its conservative relative uncertainty is 0.90%.

The result is more informative than simply reporting the concentration to four significant figures.

05Worked example: four dilutions with independent standard uncertainties

Now consider a slightly different situation.

A stock solution has a concentration of:

\[ C_0=0.1000\pm0.0005\ \mathrm{mol\,L^{-1}} \]

Four successive 1:10 dilutions are performed.

Each dilution uses:

  • A pipette transferring 10.00 mL, with standard uncertainty 0.02 mL.
  • A volumetric flask with a final volume of 100.0 mL, with standard uncertainty 0.10 mL.

Assume these are independent random standard uncertainties, and any shared calibration uncertainty has already been corrected or is negligible.

Calculate the final concentration and its relative standard uncertainty.

Step 1

\[ C_4=C_0(0.1)^4 \]

\[ C_4=0.1000\times0.0001 \]

\[ C_4=1.000\times10^{-5}\ \mathrm{mol\,L^{-1}} \]

Step 2

\[ u_r(C_0)=\frac{0.0005}{0.1000} \]

\[ u_r(C_0)=0.005=0.50\% \]

The stock concentration already has a relative standard uncertainty of 0.50%.

This uncertainty is inherited by every subsequent dilution.

Step 3

\[ u_r(V_a)=\frac{0.02}{10.00} \]

\[ u_r(V_a)=0.20\% \]

For the volumetric flask:

\[ u_r(V_f)=\frac{0.10}{100.0} \]

\[ u_r(V_f)=0.10\% \]

Step 4

Use root-sum-of-squares:

\[ u_r(d)=\sqrt{(0.20)^2+(0.10)^2} \]

\[ u_r(d)=0.224\% \]

Step 5

\[ u_r(C_4)= \sqrt{ (0.50)^2+4(0.224)^2 } \]

\[ u_r(C_4)=\sqrt{0.450} \]

\[ u_r(C_4)=0.671\% \]

Step 6

Convert the relative uncertainty into a decimal:

\[ 0.671\%=0.00671 \]

Then:

\[ u(C_4)=0.00671(1.000\times10^{-5}) \]

\[ u(C_4)=6.71\times10^{-8}\ \mathrm{mol\,L^{-1}} \]

Therefore:

\[ \boxed{ C_4=(1.000\pm0.007)\times10^{-5} \ \mathrm{mol\,L^{-1}} } \]

The relative standard uncertainty is approximately 0.67%.

Notice that the answer differs from the previous worked example. This is not a contradiction.

The first example used maximum uncertainty bounds and a conservative addition method. The second used independent random standard uncertainties and root-sum-of-squares.

The correct method depends on what the supplied uncertainty values represent.

06When the simple model is not enough

The equations above describe how measurement uncertainty propagates through an idealised serial dilution.

They do not automatically account for every possible laboratory problem.

The model assumes:

  • Each solution is thoroughly mixed before an aliquot is removed.
  • The aliquot contains the same concentration as the solution in the flask.
  • No solute is lost during transfer.
  • No contamination occurs.
  • The solute does not react, evaporate, or become significantly adsorbed onto the container.
  • The stated volume uncertainties are appropriate for the equipment and conditions.
  • Relative uncertainties are sufficiently small for the first-order approximations to be useful.

These assumptions matter.

For example, an incompletely mixed solution can produce an aliquot with a concentration different from the nominal concentration in the flask. This is not simply another pipette-volume uncertainty.

Similarly, a pipette can be precise without being accurate. If it consistently delivers the same incorrect volume, repeated transfers may have little random variation while retaining a systematic bias.

Manual pipetting variation is a recognised issue in analytical laboratories. Guan and colleagues’ 2023 review discusses sources of variation, methods of assessing pipetting performance, and the way errors can compound in multistep procedures.

For more on the practical consequences of this problem, read Why Small Pipette Errors Grow in Serial Dilutions.

The extension developed here goes one step further: instead of only recognising that small errors accumulate, we calculate how much uncertainty they contribute and distinguish between independent and shared errors.

07Questions and solutions

Question 1

A stock solution has a concentration of \(0.0800\ \mathrm{mol\,L^{-1}}\).

A student performs two successive 1:10 dilutions. At each stage:

  • The pipette transfers \(5.00\pm0.02\) mL.
  • The volumetric flask has a final volume of \(50.0\pm0.1\) mL.

Assume the stock concentration uncertainty is negligible and the stated volume uncertainties are maximum tolerance bounds.

Calculate:

a. The final concentration.

b. The conservative maximum relative uncertainty.

c. The absolute uncertainty in the final concentration.

Solution 1

The final concentration is \(8.00\times10^{-4}\ \mathrm{mol\,L^{-1}}\), with a conservative maximum relative uncertainty of 1.2%.

Step 1: Calculate the final concentration.

Each stage has a dilution ratio of:

\[ d=\frac{5.00}{50.0}=0.100 \]

After two stages:

\[ C_2=0.0800(0.100)^2 \]

\[ \boxed{C_2=8.00\times10^{-4}\ \mathrm{mol\,L^{-1}}} \]

Step 2: Calculate the relative uncertainty of the pipette.

\[ r_{V_a}=\frac{0.02}{5.00}\times100 \]

\[ r_{V_a}=0.40\% \]

Step 3: Calculate the relative uncertainty of the flask.

\[ r_{V_f}=\frac{0.1}{50.0}\times100 \]

\[ r_{V_f}=0.20\% \]

Step 4: Calculate the uncertainty per stage.

\[ r_d=0.40\%+0.20\% \]

\[ r_d=0.60\% \]

There are two stages:

\[ r_{C_2}=2(0.60\%) \]

\[ \boxed{r_{C_2}=1.2\%} \]

Step 5: Calculate the absolute uncertainty.

\[ \Delta C_2=0.012(8.00\times10^{-4}) \]

\[ \Delta C_2=9.6\times10^{-6}\ \mathrm{mol\,L^{-1}} \]

Therefore:

\[ \boxed{ C_2=(8.00\pm0.10)\times10^{-4} \ \mathrm{mol\,L^{-1}} } \]

The concentration becomes one-hundredth of its initial value, but the relative uncertainty increases because both dilution stages introduce additional volume-measurement uncertainty.

Question 2

A student prepares a solution through two successive dilutions.

The initial stock concentration is:

\[ C_0=0.250\pm0.002\ \mathrm{mol\,L^{-1}} \]

The first dilution uses:

  • An aliquot of \(4.00\pm0.02\) mL.
  • A final volume of \(100.0\pm0.1\) mL.

The second dilution uses:

  • An aliquot of \(10.00\pm0.02\) mL.
  • A final volume of \(50.0\pm0.1\) mL.

All stated uncertainties are maximum tolerance bounds.

Calculate:

a. The nominal final concentration.

b. The conservative maximum relative uncertainty.

c. The absolute uncertainty in the final concentration.

Explain why the uncertainty in the original stock must still be included.

Solution 2

The nominal final concentration is \(2.00\times10^{-3}\ \mathrm{mol\,L^{-1}}\), with a conservative maximum relative uncertainty of 1.8%.

Step 1: Calculate the first dilution ratio.

\[ d_1=\frac{4.00}{100.0}=0.0400 \]

Step 2: Calculate the second dilution ratio.

\[ d_2=\frac{10.00}{50.0}=0.200 \]

Step 3: Calculate the final concentration.

\[ C_2=C_0d_1d_2 \]

\[ C_2=0.250(0.0400)(0.200) \]

\[ \boxed{C_2=0.00200\ \mathrm{mol\,L^{-1}}} \]

Step 4: Calculate the relative uncertainty of the stock.

\[ r_{C_0}=\frac{0.002}{0.250}\times100 \]

\[ r_{C_0}=0.80\% \]

Step 5: Calculate the uncertainty contributed by the first dilution.

For its aliquot:

\[ r_{V_{a,1}}=\frac{0.02}{4.00}\times100 =0.50\% \]

For its final volume:

\[ r_{V_{f,1}}=\frac{0.1}{100.0}\times100 =0.10\% \]

Therefore:

\[ r_{d_1}=0.50+0.10=0.60\% \]

Step 6: Calculate the uncertainty contributed by the second dilution.

For its aliquot:

\[ r_{V_{a,2}}=\frac{0.02}{10.00}\times100 =0.20\% \]

For its final volume:

\[ r_{V_{f,2}}=\frac{0.1}{50.0}\times100 =0.20\% \]

Therefore:

\[ r_{d_2}=0.20+0.20=0.40\% \]

Step 7: Calculate the total relative uncertainty.

Include the original stock uncertainty:

\[ r_{C_2}=0.80+0.60+0.40 \]

\[ \boxed{r_{C_2}=1.80\%} \]

Step 8: Calculate the absolute uncertainty.

\[ \Delta C_2=0.018(0.00200) \]

\[ \Delta C_2=0.000036\ \mathrm{mol\,L^{-1}} \]

Therefore:

\[ \boxed{ C_2=(2.00\pm0.04)\times10^{-3} \ \mathrm{mol\,L^{-1}} } \]

The original stock uncertainty must be included because every diluted solution depends on the concentration of that stock.

Dilution reduces the absolute uncertainty inherited from the stock, but it does not remove its relative uncertainty. The additional volume measurements then contribute further uncertainty.

Question 3

A student begins with a stock solution:

\[ C_0=0.0500\pm0.0005\ \mathrm{mol\,L^{-1}} \]

They perform four successive 1:10 dilutions.

The combined maximum relative uncertainty in each dilution ratio is 0.50%.

The student argues:

“The final concentration is 10,000 times smaller than the stock concentration. Therefore, the relative uncertainty inherited from the stock must also be 10,000 times smaller.”

Evaluate this argument.

Calculate:

a. The nominal final concentration.

b. The conservative maximum relative uncertainty.

c. The absolute uncertainty in the final concentration.

Your explanation must distinguish between the inherited stock uncertainty and the additional uncertainty introduced by the dilution stages.

Solution 3

The student’s argument is incorrect. The inherited relative uncertainty from the stock remains 1.0%, while the four dilution stages introduce another 2.0%, giving a total conservative relative uncertainty of 3.0%.

Step 1: Calculate the initial stock’s relative uncertainty.

\[ r_{C_0}=\frac{0.0005}{0.0500}\times100 \]

\[ r_{C_0}=1.0\% \]

Step 2: Calculate the final concentration.

Four successive 1:10 dilutions give:

\[ C_4=C_0(0.1)^4 \]

\[ C_4=0.0500(0.0001) \]

\[ \boxed{C_4=5.00\times10^{-6}\ \mathrm{mol\,L^{-1}}} \]

Step 3: Calculate the uncertainty added by the dilution stages.

Each stage contributes 0.50%.

Therefore:

\[ r_{\mathrm{dilution}}=4(0.50\%) \]

\[ r_{\mathrm{dilution}}=2.0\% \]

Step 4: Calculate the total relative uncertainty.

\[ r_{C_4}=1.0\%+2.0\% \]

\[ \boxed{r_{C_4}=3.0\%} \]

Step 5: Calculate the absolute uncertainty.

\[ \Delta C_4=0.030(5.00\times10^{-6}) \]

\[ \boxed{ \Delta C_4=1.50\times10^{-7}\ \mathrm{mol\,L^{-1}} } \]

The final result is:

\[ \boxed{ C_4=(5.00\pm0.15)\times10^{-6} \ \mathrm{mol\,L^{-1}} } \]

The student’s mistake is confusing absolute uncertainty with relative uncertainty.

The stock’s absolute uncertainty is reduced by the same dilution factor as its concentration:

\[ \frac{0.0005}{10\,000} =5.0\times10^{-8}\ \mathrm{mol\,L^{-1}} \]

However, its relative uncertainty remains 1.0%.

The four dilution stages then contribute an additional 2.0% relative uncertainty.

The concentration becomes smaller, but that does not automatically make the result proportionally more precise.

Question 4

A laboratory performs three successive nominal 1:10 dilutions using the same pipette and the same volumetric flask.

The stock concentration is:

\[ C_0=0.1000\ \mathrm{mol\,L^{-1}} \]

The nominal equipment specifications are:

  • Pipette: 10.00 mL.
  • Volumetric flask: 100.0 mL.

A subsequent calibration check reveals:

  • The pipette actually delivers 10.03 mL when set to 10.00 mL.
  • The volumetric flask actually contains 100.2 mL when filled to its stated calibration mark.

Assume these calibration offsets are known and apply consistently at every stage. There is no loss of solute.

Calculate the actual final concentration.

Explain why adding the two percentage deviations as positive uncertainties would produce an incorrect result.

Solution 4

The actual final concentration is approximately \(1.0030\times10^{-4}\ \mathrm{mol\,L^{-1}}\), which is about 0.30% higher than the nominal result.

The important detail is that the two calibration offsets have opposite effects on the dilution ratio.

Step 1: Calculate the nominal dilution ratio.

\[ d_{\mathrm{nominal}}=\frac{10.00}{100.0} \]

\[ d_{\mathrm{nominal}}=0.1000 \]

Step 2: Calculate the actual dilution ratio.

The actual aliquot is 10.03 mL, and the actual final volume is 100.2 mL.

\[ d_{\mathrm{actual}}=\frac{10.03}{100.2} \]

\[ d_{\mathrm{actual}}\approx0.1000998 \]

The actual dilution ratio is slightly larger than the nominal ratio.

Step 3: Calculate the nominal final concentration.

\[ C_{3,\mathrm{nominal}} =0.1000(0.1000)^3 \]

\[ C_{3,\mathrm{nominal}} =1.000\times10^{-4}\ \mathrm{mol\,L^{-1}} \]

Step 4: Calculate the actual final concentration.

\[ C_{3,\mathrm{actual}} =0.1000\left(\frac{10.03}{100.2}\right)^3 \]

\[ C_{3,\mathrm{actual}} \approx1.002997\times10^{-4} \ \mathrm{mol\,L^{-1}} \]

Therefore:

\[ \boxed{ C_{3,\mathrm{actual}} \approx1.0030\times10^{-4} \ \mathrm{mol\,L^{-1}} } \]

Step 5: Explain the competing effects.

The pipette delivers 0.30% more liquid than intended.

This increases the amount of dissolved solute transferred into the next flask, increasing the resulting concentration.

The volumetric flask holds 0.20% more liquid than intended.

This increases the final solution volume, decreasing the resulting concentration.

For small deviations, the approximate net relative change in the dilution ratio is:

\[ +0.30\%-0.20\%=+0.10\% \]

Because the same equipment is used three times, the resulting concentration is approximately 0.30% higher than the nominal value.

The tempting approach is to add 0.30% and 0.20% as positive uncertainties. That would ignore the direction of each calibration offset.

This question also illustrates why a known systematic error should be corrected rather than treated as an unspecified random uncertainty.

Question 5

A laboratory prepares a dilute solution using four successive 1:10 dilutions.

The stock concentration has a relative standard uncertainty of 0.30%.

The same pipette and volumetric flask are used for all four dilution stages.

For each dilution ratio, the laboratory identifies two uncertainty components:

  • Repeatability uncertainty: 0.20% relative standard uncertainty at each stage. These random variations are independent between stages.
  • Calibration uncertainty: 0.15% relative standard uncertainty. This uncertainty is common to all four stages because the same calibrated equipment is used repeatedly.

Assume:

  • The stock concentration uncertainty is independent of the equipment uncertainties.
  • The repeatability and calibration components are independent of each other.
  • The calibration component is one shared uncertainty, not four independent uncertainties.
  • All relative uncertainties are small enough for first-order propagation.

Answer the following:

a. Calculate the final relative standard uncertainty.

b. Calculate what the final relative standard uncertainty would be if the 0.15% calibration component were genuinely independent at each stage instead.

c. The laboratory repeats the entire four-stage dilution three times using the same stock solution and the same equipment, then averages the three final concentrations. Assuming the repeatability variations are independent between chains, while the stock and calibration uncertainties remain common, calculate the relative standard uncertainty of the average.

Explain why repeating the experiment does not eliminate every source of uncertainty.

Solution 5

The final relative standard uncertainty for one dilution chain is approximately 0.78%. The shared calibration uncertainty must be multiplied by four, not combined as though it were four independent random variations.

Step 1: Identify the stock concentration contribution.

The stock has a relative standard uncertainty of:

\[ u_r(C_0)=0.30\% \]

This uncertainty is inherited by the entire dilution chain.

Step 2: Calculate the repeatability contribution.

There are four independent repeatability components, each with a relative standard uncertainty of 0.20%.

Combine them using root-sum-of-squares:

\[ u_{r,\mathrm{repeat}} =\sqrt{4(0.20)^2} \]

\[ \boxed{u_{r,\mathrm{repeat}}=0.40\%} \]

Step 3: Calculate the shared calibration contribution.

The same calibration uncertainty affects all four dilution stages.

Because the same dilution ratio is effectively applied four times, the common component contributes:

\[ u_{r,\mathrm{cal}}=4(0.15\%) \]

\[ \boxed{u_{r,\mathrm{cal}}=0.60\%} \]

This is the central difficulty in the question.

A common calibration uncertainty does not behave like four independent repeatability variations.

Step 4: Calculate the final relative standard uncertainty.

The stock, repeatability, and common calibration components are independent of one another.

Therefore:

\[ u_r(C_4)= \sqrt{ (0.30)^2+ (0.40)^2+ (0.60)^2 } \]

\[ u_r(C_4)=\sqrt{0.61} \]

\[ \boxed{u_r(C_4)=0.78\%} \]

This is the correct result for part (a).

Step 5: Recalculate under the independent calibration assumption.

Suppose the calibration uncertainty were genuinely independent at each stage.

It would then contribute:

\[ u_{r,\mathrm{cal}} =\sqrt{4(0.15)^2} \]

\[ u_{r,\mathrm{cal}}=0.30\% \]

The final relative standard uncertainty would become:

\[ u_r(C_4)= \sqrt{ (0.30)^2+ (0.40)^2+ (0.30)^2 } \]

\[ u_r(C_4)=\sqrt{0.34} \]

\[ \boxed{u_r(C_4)=0.58\%} \]

This is the answer to part (b).

The difference between 0.78% and 0.58% comes entirely from the assumption about correlation.

Treating a shared calibration uncertainty as independent at every stage underestimates the final uncertainty.

Step 6: Consider averaging three independently prepared chains.

Averaging independent random variations reduces their standard uncertainty by a factor of \(\sqrt{3}\).

Therefore, the repeatability contribution becomes:

\[ u_{r,\mathrm{repeat,mean}} =\frac{0.40}{\sqrt{3}} \]

\[ u_{r,\mathrm{repeat,mean}} \approx0.231\% \]

However, the stock concentration uncertainty remains 0.30%, and the shared calibration uncertainty remains 0.60%.

Both affect all three chains.

The relative standard uncertainty of the average is:

\[ u_{r,\mathrm{mean}} = \sqrt{ (0.30)^2+ (0.60)^2+ (0.231)^2 } \]

\[ u_{r,\mathrm{mean}} \approx\sqrt{0.503} \]

\[ \boxed{u_{r,\mathrm{mean}}\approx0.71\%} \]

Repeating the experiment has reduced the independent repeatability contribution, but it has not removed the shared calibration uncertainty or the uncertainty in the original stock.

This is why taking more measurements does not automatically solve every measurement problem.

The experiment can become more precise through repetition while retaining a systematic limitation.

08What to do with the uncertainty next

The most useful habit is to identify what kind of uncertainty you are dealing with before choosing an equation.

When calculating a serial dilution:

  1. Calculate the nominal concentration using the dilution ratios.
  2. Identify whether the supplied ± values are maximum bounds or standard uncertainties.
  3. Include the uncertainty inherited from the original stock.
  4. Identify whether measurement errors are independent or share a calibration effect.
  5. Calculate the final relative uncertainty, then convert it into an absolute uncertainty if required.

The deeper lesson is that a calculated result and the reliability of that result are two different things.

This also connects directly to the conservation-of-mass practical. When comparing initial and final measurements, you need to consider the uncertainty of the measurement process before deciding whether an apparent difference represents a real physical change.

The next useful extension is carrying uncertainty into analytical measurements, such as colourimetry and calibration curves, where the uncertainty in a serially diluted standard can affect every subsequent concentration calculated from the experimental data.