Internal Standards: Using Response Ratios in Chemical Analysis
Learn how internal standards improve quantitative chemical analysis by comparing analyte and reference signals. Explore response ratios, calibration curves, worked examples, and the limitations of the method.
Two injections of the same chemical sample produce different results. The first gives an analyte peak area of 18,000, while the second gives 27,000. Does the second sample contain 50% more analyte?
Not necessarily. The instrument may simply have received 50% more sample during the second injection. If another compound in the sample also produces a peak 50% larger, the ratio between the two peak areas has not changed. That ratio can tell us more about the actual composition of the sample than either peak area alone.
This is the central idea behind internal standards: adding a known amount of a second compound to an unknown sample, then comparing the detector responses of the two compounds to reduce measurement variability.
Internal standardisation is an extension beyond the required Year 11 Quantitative chemistry syllabus content. It is not an HSC assessment requirement. Year 11 quantitative chemistry, including the particle model of ideal gases, provides the starting knowledge for understanding the method. The NSW Chemistry 11-12 Syllabus (2025) is scheduled for implementation for Year 11 from 2028.
01Why measuring a sample can be inconsistent
Imagine analysing a gas mixture using a technique called gas chromatography. The mixture contains several gases, including the substance we want to measure.
The instrument separates the components and produces a chromatogram. Each substance appears as a peak, and the area under that peak is related to the amount of substance detected.
Due to technical limitations, this figure may display incorrectly or contain inaccuracies. Please check key details against your course materials.
Ideally, injecting the same sample twice should produce identical peak areas. In practice, several things can interfere:
- Slight differences in the volume injected.
- Small sample losses during preparation.
- Changes in detector sensitivity.
- Differences in how the sample is transported through the instrument.
- Changes in the sample matrix, meaning the other substances present alongside the analyte.
Suppose the first injection delivers 1.0 microlitre of sample, but the second delivers only 0.8 microlitres.
The second injection contains approximately 20% less of every component, assuming the sample is homogeneous and both injections have the same composition.
The detector will therefore produce smaller peaks, even though the original sample has not changed.
This creates a problem: how can we distinguish a genuinely lower analyte concentration from a smaller amount of sample reaching the detector?
An internal standard provides a reference point.
02Start with the particle model of gases
Before introducing the analytical method, consider what happens to the particles in a gas sample.
The ideal gas particle model describes gas particles as:
- Moving continuously and randomly.
- Having negligible volume compared with the space between them.
- Experiencing negligible intermolecular attractions.
- Undergoing elastic collisions with each other and the container walls.
The ideal gas equation connects the macroscopic properties of the gas:
\[ PV=nR_gT \]
Where:
- \(P\) is the absolute pressure in pascals (Pa).
- \(V\) is the gas volume in cubic metres (m³).
- \(n\) is the amount of gas in moles (mol).
- \(R_g\) is the universal gas constant, approximately \(8.314\ \mathrm{J\,mol^{-1}\,K^{-1}}\).
- \(T\) is the absolute temperature in kelvin (K).
Rearranging:
\[ n=\frac{PV}{R_gT} \]
At constant temperature and pressure, the number of moles in a gas sample is directly proportional to its volume.
Now imagine a gas mixture containing two components:
- Analyte A, which we want to measure.
- Compound B, which we will use as an internal standard.
If the gas is homogeneous, both components are distributed throughout the sample.
A smaller sample therefore contains fewer particles of A and fewer particles of B. However, the ratio of their amounts remains unchanged.
Think of a well-mixed bag of red and blue beads. Taking a smaller scoop reduces the number of both colours, but the red-to-blue ratio stays the same.
The analogy has one limitation: a chemical detector does not necessarily produce the same signal for equal amounts of different compounds. We will account for that shortly.
This is the physical idea that makes internal standardisation useful.
If both compounds experience the same change in sample delivery, comparing their signals can remove the effect of that change.
The ideal gas model is a useful starting point, not an exact description of every real gas. Real gases can deviate from ideal behaviour, particularly at high pressures and low temperatures. Internal standardisation also applies well beyond gas analysis, including liquid chromatography and mass spectrometry.
03What is an internal standard?
An internal standard is a compound added in a known amount to a sample to help determine the amount of an unknown substance.
The compound being measured is called the analyte.
The added reference compound is called the internal standard.
For a successful analysis, the internal standard should:
- Be different from the analyte so that the detector can distinguish their signals.
- Have sufficiently similar physical and chemical behaviour to the analyte.
- Be absent from the original sample, or have a known background contribution.
- Remain chemically stable during sample preparation and analysis.
- Produce a measurable signal that does not overlap with the analyte or other sample components.
The same known amount of internal standard is added to every calibration standard and unknown sample.
IUPAC describes an internal standard as a compound added to a sample in a known concentration to assist with qualitative identification or quantitative determination (IUPAC Gold Book).
Why must the internal standard be chemically similar to the analyte?
Because we want both compounds to respond similarly to changes in the experimental conditions.
For example, if a small amount of both compounds is lost during sample preparation, their relative amounts should remain approximately unchanged.
However, similarity alone is not enough. Their detector signals must also be distinguishable.
A stable-isotope-labelled version of the analyte is often an excellent internal standard in mass spectrometry. It behaves very similarly to the original compound but can be distinguished using its different mass.
This idea leads into the more specialised technique of isotope dilution analysis.
04How the response ratio works
A chromatogram displays detector response against time.
As different substances pass through the detector, they produce peaks at different retention times. The retention time helps identify a compound, while the peak area is used to estimate how much of it is present.
For quantitative analysis, peak area is generally preferred because it accounts for the entire detector response across the peak.
Consider two compounds:
- Analyte A.
- Internal standard IS.
Let:
- \(A_A\) = peak area of the analyte.
- \(A_{IS}\) = peak area of the internal standard.
We define the response ratio as:
\[ r=\frac{A_A}{A_{IS}} \]
This ratio is the starting point of internal standardisation.
Why does dividing the signals help?
Suppose both compounds are injected together.
Their detector responses depend on the amounts injected and the sensitivity of the detector to each compound.
We can represent this using:
\[ A_A=gk_An_A \]
\[ A_{IS}=gk_{IS}n_{IS} \]
Where:
- \(g\) is a common factor representing sample delivery and instrument sensitivity.
- \(k_A\) is the detector sensitivity towards the analyte.
- \(k_{IS}\) is the detector sensitivity towards the internal standard.
- \(n_A\) is the amount of analyte injected, in moles.
- \(n_{IS}\) is the amount of internal standard injected, in moles.
Dividing the two expressions gives:
\[ \begin{aligned} r&=\frac{A_A}{A_{IS}}\\ &=\frac{gk_An_A}{gk_{IS}n_{IS}}\\ &=\frac{k_A}{k_{IS}} \times\frac{n_A}{n_{IS}} \end{aligned} \]
Notice that the common factor \(g\) disappears.
This is important.
If the injection volume decreases, both peak areas decrease by approximately the same proportion. Their ratio remains unchanged, provided both compounds respond linearly and experience the same proportional change.
However, there is another complication.
The detector might be more sensitive to the analyte than to the internal standard.
For example, equal amounts of two compounds might produce peak areas of 20,000 and 30,000. Their signals are not directly comparable without accounting for this difference.
We therefore need a relative response factor.
The relative response factor
We define the relative response factor, \(F\), as:
\[ F= \frac{ (A_A/A_{IS})_{\mathrm{standard}} }{ (n_A/n_{IS})_{\mathrm{standard}} } \]
This definition compares the detector response ratio with the known amount ratio in a calibration standard.
The factor is dimensionless when both amounts are expressed in the same units.
Rearranging gives the central equation:
\[ r=F\times\frac{n_A}{n_{IS}} \]
For solutions prepared to the same final volume, the mole ratio is equal to the concentration ratio:
\[ \boxed{ r=F\times\frac{c_A}{c_{IS}} } \]
Where:
- \(c_A\) is the analyte concentration.
- \(c_{IS}\) is the internal standard concentration.
Rearranging to calculate the unknown analyte concentration:
\[ \boxed{ c_A=\frac{r}{F}\times c_{IS} } \]
This equation is the foundation of quantitative analysis using an internal standard.
Be aware that different analytical methods sometimes define response factors using the reciprocal convention. Always check the definition being used. Here, \(F\) is explicitly defined as the signal ratio divided by the amount or concentration ratio.
05How do we determine the response factor?
We need a calibration standard containing known concentrations of both the analyte and the internal standard.
This is similar to preparing a standard solution for ordinary quantitative analysis. If you need to revise the preparation process, see How to Prepare a Standard Solution for HSC Chemistry.
Suppose a standard contains:
- Analyte concentration: \(4.00\ \mathrm{mg/L}\).
- Internal standard concentration: \(2.00\ \mathrm{mg/L}\).
The detector produces:
- Analyte peak area: 32,000.
- Internal standard peak area: 20,000.
The measured response ratio is:
\[ r=\frac{32000}{20000}=1.60 \]
The known concentration ratio is:
\[ \frac{c_A}{c_{IS}}=\frac{4.00}{2.00}=2.00 \]
Therefore:
\[ F=\frac{1.60}{2.00}=0.800 \]
We now have the relative response factor needed to analyse an unknown sample using the same analytical method.
A single calibration standard can provide an estimate of the response factor, but a series of calibration standards is generally more reliable.
Using a calibration curve
Instead of relying on one standard, we can prepare several standards containing different known analyte concentrations.
Each standard receives the same known concentration of internal standard.
We then calculate the response ratio for each standard and plot:
- Horizontal axis: analyte concentration, \(c_A\).
- Vertical axis: response ratio, \(r\).
If the detector response is linear over the concentration range, the relationship can be represented by:
\[ r=mc_A+b \]
Where:
- \(m\) is the gradient of the calibration line.
- \(b\) is the vertical intercept.
- \(c_A\) is the known analyte concentration.
For an unknown sample, we measure its response ratio and rearrange:
\[ \boxed{ c_A=\frac{r-b}{m} } \]
The intercept should not automatically be assumed to be zero. It should be determined from the calibration data and the behaviour of the analytical method.
The National Institute of Standards and Technology discusses the use of internal standard calibration, external standard calibration, and different calibration models in chromatographic analysis (NIST: Principles of Quantitation).
06When does internal standardisation work?
Internal standardisation is particularly useful when an experimental error affects the analyte and internal standard in approximately the same way.
| Source of variation | Can an internal standard help? |
|---|---|
| Slightly different injection volumes | Yes, if both compounds are affected proportionally |
| Common changes in detector sensitivity | Yes, if both signals change proportionally |
| Sample dilution after adding the internal standard | Yes, if both compounds undergo the same dilution |
| Sample loss during preparation | Potentially, if the internal standard was added beforehand and behaves similarly |
| Analyte lost before the internal standard is added | No |
| Different extraction efficiencies | Not necessarily |
| An interfering compound overlapping one peak | No, not without resolving or correcting the interference |
| Detector saturation or non-linear response | Not reliably |
The important distinction is between a common change and a selective change.
A common change affects both compounds in approximately the same proportion. A selective change affects one compound more than the other.
Internal standardisation is designed to compensate for the first type.
It cannot automatically correct the second.
The timing of the internal standard matters
Imagine extracting an analyte from a complex sample.
Some analyte may be lost during the extraction process. If the internal standard is added before extraction, it has an opportunity to experience the same losses.
If it is added after extraction, it cannot tell us how much analyte was lost earlier.
For the method to correct extraction losses, the analyte and internal standard must have sufficiently similar recovery.
This is one reason why a chemically similar internal standard is important.
Matrix effects can still cause problems
The sample matrix is everything in the sample other than the analyte and the internal standard.
For example, a blood sample contains many dissolved substances. These can affect how efficiently compounds enter an analytical instrument or how strongly the detector responds.
Suppose the sample matrix reduces the analyte signal by 20%.
If it also reduces the internal standard signal by 20%, the response ratio remains approximately unchanged.
But if the analyte signal decreases by 20% while the internal standard signal decreases by only 5%, the ratio changes.
The method will produce a biased result.
An analytical review by Wilschefski and Baxter explains how internal standards are used to address matrix effects and instrument drift in inductively coupled plasma mass spectrometry. It also discusses why choosing an unsuitable internal standard can produce inaccurate results (Clinical Biochemist Reviews, 2019).
The key lesson is that an internal standard is not a magic correction for every experimental error. Its usefulness depends on how closely it follows the analyte throughout the measurement process.
07Worked examples
Worked example: Determine an unknown concentration using a single calibration standard
A calibration standard contains an analyte at \(3.00\ \mathrm{mg/L}\) and an internal standard at \(1.50\ \mathrm{mg/L}\). Their measured peak areas are 27,000 and 18,000, respectively.
An unknown sample is prepared with the same internal standard concentration of \(1.50\ \mathrm{mg/L}\). Its analyte and internal standard peak areas are 18,000 and 24,000.
Calculate the analyte concentration in the unknown sample.
Step 1
\[ r_{\mathrm{standard}} =\frac{27000}{18000} =1.50 \]
Step 2
\[ \frac{c_A}{c_{IS}} =\frac{3.00}{1.50} =2.00 \]
Step 3
\[ F=\frac{1.50}{2.00} =0.750 \]
Step 4
\[ r_{\mathrm{unknown}} =\frac{18000}{24000} =0.750 \]
Step 5
Using:
\[ c_A=\frac{r}{F}\times c_{IS} \]
Substitute:
\[ c_A=\frac{0.750}{0.750}\times1.50 \]
\[ \boxed{c_A=1.50\ \mathrm{mg/L}} \]
The unknown sample contains \(1.50\ \mathrm{mg/L}\) of analyte.
Notice that the unknown has a different response ratio from the calibration standard, even though the internal standard concentration is unchanged.
The response factor allows us to account for the detector’s different sensitivities to the two compounds.
Worked example: Use a calibration curve and account for injection variability
Four calibration standards are prepared, each containing an internal standard concentration of \(1.00\ \mathrm{mg/L}\).
Their measured response ratios are:
| Analyte concentration (mg/L) | Response ratio |
|---|---|
| 0.50 | 0.22 |
| 1.00 | 0.42 |
| 1.50 | 0.62 |
| 2.00 | 0.82 |
An unknown sample is prepared with an internal standard concentration of \(1.00\ \mathrm{mg/L}\).
Its peak areas are:
- Analyte: 16,500.
- Internal standard: 25,000.
Calculate the analyte concentration. Then determine whether the result changes if a second injection produces half the peak areas of the first injection.
Step 1
The data follow a linear relationship:
\[ r=0.40c_A+0.02 \]
The gradient is \(0.40\), and the intercept is \(0.02\).
Step 2
\[ r=\frac{16500}{25000} \]
\[ r=0.660 \]
Step 3
\[ 0.660=0.40c_A+0.02 \]
Rearrange:
\[ c_A=\frac{0.660-0.02}{0.40} \]
\[ \boxed{c_A=1.60\ \mathrm{mg/L}} \]
Step 4
The second injection produces:
- Analyte peak area: 8,250.
- Internal standard peak area: 12,500.
Its response ratio is:
\[ r=\frac{8250}{12500}=0.660 \]
The response ratio is identical.
Therefore, the calculated concentration remains:
\[ \boxed{c_A=1.60\ \mathrm{mg/L}} \]
The second injection contains smaller absolute signals, but the relative response has not changed.
This example demonstrates why an internal standard can improve the reliability of a quantitative measurement when the injection volume varies.
It also highlights an important assumption: the detector must remain within its linear working range. A ratio cannot automatically correct for signals affected by saturation or significant background noise.
08Questions and solutions
Question 1
A calibration standard contains an analyte concentration of \(3.00\ \mathrm{mg/L}\) and an internal standard concentration of \(1.50\ \mathrm{mg/L}\).
The measured peak areas are:
- Analyte: 27,000.
- Internal standard: 18,000.
An unknown sample contains the same internal standard concentration, \(1.50\ \mathrm{mg/L}\), and produces peak areas of:
- Analyte: 18,000.
- Internal standard: 24,000.
Calculate the analyte concentration in the unknown sample.
Solution 1
The analyte concentration is \(1.50\ \mathrm{mg/L}\).
First, calculate the response ratio of the calibration standard:
\[ r_{\mathrm{standard}}=\frac{27000}{18000}=1.50 \]
Calculate the concentration ratio:
\[ \frac{c_A}{c_{IS}}=\frac{3.00}{1.50}=2.00 \]
Determine the relative response factor:
\[ F=\frac{1.50}{2.00}=0.750 \]
Now calculate the response ratio of the unknown:
\[ r_{\mathrm{unknown}}=\frac{18000}{24000}=0.750 \]
Apply the internal standard equation:
\[ c_A=\frac{r}{F}\times c_{IS} \]
\[ c_A=\frac{0.750}{0.750}\times1.50 \]
\[ \boxed{c_A=1.50\ \mathrm{mg/L}} \]
The unknown has half the analyte concentration of the calibration standard, despite containing the same internal standard concentration.
The calculation assumes a linear detector response, a valid response factor, and no significant interference.
Question 2
A laboratory analyses an unknown sample using an internal standard.
The calibration equation is:
\[ r=0.50c_A+0.010 \]
where \(c_A\) is measured in \(\mathrm{mg/L}\).
Two injections of the same prepared sample produce the following results:
| Measurement | Injection 1 | Injection 2 |
|---|---|---|
| Analyte peak area | 30,000 | 18,000 |
| Internal standard peak area | 25,000 | 15,000 |
(a) Calculate the analyte concentration using each injection.
(b) Explain why the raw analyte peak areas differ substantially even though both injections came from the same prepared sample.
(c) A student suggests averaging the two analyte peak areas before calculating the concentration. Explain why using the response ratios is more appropriate.
Solution 2
Both injections give an analyte concentration of \(2.38\ \mathrm{mg/L}\).
The key is to calculate the response ratio separately for each injection.
(a) Calculate the concentration from each injection.
For injection 1:
\[ r_1=\frac{30000}{25000}=1.20 \]
Substitute into the calibration equation:
\[ 1.20=0.50c_A+0.010 \]
\[ c_A=\frac{1.20-0.010}{0.50} \]
\[ \boxed{c_A=2.38\ \mathrm{mg/L}} \]
For injection 2:
\[ r_2=\frac{18000}{15000}=1.20 \]
Therefore:
\[ c_A=\frac{1.20-0.010}{0.50} \]
\[ \boxed{c_A=2.38\ \mathrm{mg/L}} \]
Both injections produce the same concentration.
(b) Explain the difference in peak areas.
Injection 2 has smaller peak areas because less sample was delivered to the detector, or because of another common proportional change in signal.
The analyte peak area decreased by 40%.
The internal standard peak area also decreased by 40%.
Because both signals changed by the same proportion, their ratio remained constant.
The sample composition has not changed.
(c) Explain why averaging the analyte peak areas is inappropriate.
Averaging the raw peak areas would ignore the variation in sample delivery.
The detector response depends on both the amount of analyte present and the amount of sample reaching the detector.
The internal standard provides a reference for the second factor.
Calculating the ratio removes the common proportional variation, allowing the analyte concentration to be determined more reliably.
This is the main advantage of internal standardisation over relying on the analyte signal alone.
Question 3
A laboratory measures an analyte concentration of \(5.00\ \mathrm{mg/L}\) before a sample extraction procedure.
The extraction procedure recovers 70.0% of the analyte. Assume the final sample volume remains unchanged.
An internal standard is available at a concentration of \(2.00\ \mathrm{mg/L}\), and the relative response factor is \(F=0.800\).
Assume the internal standard also has an extraction recovery of 70.0% when added before extraction.
Compare two procedures:
- Procedure A: Add the internal standard before extraction.
- Procedure B: Add the internal standard after extraction.
(a) Calculate the response ratio expected for Procedure A.
(b) Calculate the response ratio expected for Procedure B.
(c) Determine the analyte concentration that would be inferred using the known internal standard concentration of \(2.00\ \mathrm{mg/L}\) in each procedure.
(d) Explain why the timing of the internal standard addition changes the result.
Solution 3
Procedure A correctly recovers the original analyte concentration of \(5.00\ \mathrm{mg/L}\), whereas Procedure B reports only \(3.50\ \mathrm{mg/L}\).
The difference occurs because only Procedure A allows the internal standard to experience the same extraction loss as the analyte.
(a) Calculate the response ratio for Procedure A.
The internal standard is added before extraction.
Both compounds experience a recovery of 70.0%.
The recovered analyte concentration is:
\[ c_A=0.700\times5.00 \]
\[ c_A=3.50\ \mathrm{mg/L} \]
The recovered internal standard concentration is:
\[ c_{IS}=0.700\times2.00 \]
\[ c_{IS}=1.40\ \mathrm{mg/L} \]
The response ratio is:
\[ r=F\times\frac{c_A}{c_{IS}} \]
\[ r=0.800\times\frac{3.50}{1.40} \]
\[ \boxed{r=2.00} \]
(b) Calculate the response ratio for Procedure B.
The internal standard is added after extraction.
The analyte has already experienced its 70.0% extraction recovery:
\[ c_A=3.50\ \mathrm{mg/L} \]
However, the internal standard is not subjected to the extraction loss:
\[ c_{IS}=2.00\ \mathrm{mg/L} \]
Therefore:
\[ r=0.800\times\frac{3.50}{2.00} \]
\[ \boxed{r=1.40} \]
(c) Calculate the inferred analyte concentrations.
For Procedure A, the known internal standard concentration at the point of addition was \(2.00\ \mathrm{mg/L}\).
The equal extraction recoveries mean the original analyte-to-internal-standard ratio is preserved.
Therefore:
\[ c_A=\frac{2.00}{0.800}\times2.00 \]
\[ \boxed{c_A=5.00\ \mathrm{mg/L}} \]
For Procedure B, the internal standard was added after extraction. It cannot account for the analyte already lost.
Therefore:
\[ c_A=\frac{1.40}{0.800}\times2.00 \]
\[ \boxed{c_A=3.50\ \mathrm{mg/L}} \]
(d) Explain the effect of addition timing.
In Procedure A, the analyte and internal standard experience the same proportional loss. Their original ratio is preserved, so the known amount of internal standard can be used to reconstruct the original analyte concentration.
In Procedure B, only the analyte experiences the extraction loss. The internal standard is added too late to account for it.
The response ratio therefore reflects the reduced analyte concentration in the extract rather than the original sample.
An internal standard can only correct for processes that it actually experiences.
Question 4
A laboratory is validating two possible internal standards, X and Y, for measuring an analyte A.
A validation sample contains:
- Analyte concentration: \(3.00\ \mathrm{mg/L}\).
- Internal standard concentration: \(2.00\ \mathrm{mg/L}\).
The relative response factor for each internal standard is \(0.800\) in a clean calibration matrix.
In the actual sample matrix:
- The analyte signal is suppressed to 80% of its clean-matrix value.
- Internal standard X is suppressed to 80% of its clean-matrix value.
- Internal standard Y is suppressed to 95% of its clean-matrix value.
Assume both internal standards are stable, absent from the original sample, and produce fully resolved peaks. Each has its own valid clean-matrix calibration.
(a) Calculate the clean-matrix response ratio.
(b) Calculate the response ratio obtained using internal standard X in the sample matrix.
(c) Calculate the response ratio obtained using internal standard Y in the sample matrix.
(d) Calculate the analyte concentration that would be reported using each internal standard and explain which is more suitable.
Solution 4
Internal standard X is more suitable because it experiences the same proportional signal suppression as the analyte. It produces the correct analyte concentration of \(3.00\ \mathrm{mg/L}\).
Internal standard Y produces an underestimated concentration because its signal is suppressed less than the analyte signal.
(a) Calculate the clean-matrix response ratio.
Use:
\[ r=F\times\frac{c_A}{c_{IS}} \]
Substitute:
\[ r=0.800\times\frac{3.00}{2.00} \]
\[ \boxed{r=1.20} \]
This is the expected ratio before matrix effects are introduced.
(b) Calculate the response ratio using X.
Both the analyte and internal standard X experience 80% signal retention.
The ratio therefore remains unchanged:
\[ r_X=1.20\times\frac{0.80}{0.80} \]
\[ \boxed{r_X=1.20} \]
The inferred concentration is:
\[ c_A=\frac{1.20}{0.800}\times2.00 \]
\[ \boxed{c_A=3.00\ \mathrm{mg/L}} \]
Internal standard X successfully compensates for the matrix effect.
(c) Calculate the response ratio using Y.
The analyte retains 80% of its original signal, while Y retains 95%.
Therefore:
\[ r_Y=1.20\times\frac{0.80}{0.95} \]
\[ \boxed{r_Y=1.01} \]
(d) Calculate the inferred concentration using Y.
\[ c_A=\frac{1.0105}{0.800}\times2.00 \]
\[ \boxed{c_A=2.53\ \mathrm{mg/L}} \]
This is lower than the actual concentration of \(3.00\ \mathrm{mg/L}\).
The reason is that the analyte signal has been suppressed more strongly than the internal standard signal. Dividing the two signals does not completely remove the matrix effect.
This question demonstrates an important principle: an internal standard should be selected based on how it behaves under the actual analytical conditions, not simply because it produces a convenient signal.
Question 5
A chemist adds \(10.0\ \mathrm{\mu mol}\) of an internal standard B to an unknown sample before extraction.
Independent validation has established that B has an extraction recovery of 80.0%.
After extraction, the sample is analysed using chromatography.
The measured peak areas are:
- Analyte A: 18,000.
- Internal standard B: 20,000.
The relative response factor is:
\[ F=0.900 \]
Assume the detector response is linear, there are no interfering peaks, and any variation in sample injection affects both compounds proportionally.
A student calculates the original analyte amount using:
\[ n_A=\frac{r}{F}\times n_{IS} \]
and reports \(10.0\ \mathrm{\mu mol}\).
However, the extraction recovery of analyte A has not been independently established.
(a) Calculate the amount of internal standard recovered after extraction.
(b) Calculate the amount of analyte recovered after extraction.
(c) Determine the original amount of analyte if its extraction recovery was 80.0%.
(d) Determine the original amount of analyte if its extraction recovery was instead 50.0%.
(e) Is the student’s answer uniquely justified by the experimental data? Explain why two different original analyte amounts can produce the same measured response ratio, and state what additional information is required to determine the original amount reliably.
Solution 5
The amount of analyte recovered is \(8.00\ \mathrm{\mu mol}\), but the original amount cannot be uniquely determined without knowing the analyte’s extraction recovery.
This is a more subtle limitation of internal standardisation: correcting for common signal variation does not automatically establish that two different compounds have identical extraction recoveries.
(a) Calculate the recovered internal standard amount.
The initial amount added was:
\[ n_{IS,\mathrm{initial}}=10.0\ \mathrm{\mu mol} \]
The extraction recovery was 80.0%.
Therefore:
\[ n_{IS,\mathrm{recovered}} =0.800\times10.0 \]
\[ \boxed{n_{IS,\mathrm{recovered}}=8.00\ \mathrm{\mu mol}} \]
(b) Calculate the recovered analyte amount.
First, calculate the measured response ratio:
\[ r=\frac{18000}{20000}=0.900 \]
The relative response factor is also \(0.900\).
Therefore:
\[ \frac{n_{A,\mathrm{recovered}}} {n_{IS,\mathrm{recovered}}} =\frac{r}{F} \]
\[ =\frac{0.900}{0.900}=1.00 \]
The recovered amounts of analyte and internal standard are equal.
Since the internal standard recovery amount is \(8.00\ \mathrm{\mu mol}\):
\[ \boxed{n_{A,\mathrm{recovered}}=8.00\ \mathrm{\mu mol}} \]
(c) Calculate the original analyte amount at 80.0% recovery.
If the analyte recovery is also 80.0%:
\[ n_{A,\mathrm{initial}} =\frac{8.00}{0.800} \]
\[ \boxed{n_{A,\mathrm{initial}}=10.0\ \mathrm{\mu mol}} \]
In this situation, the analyte and internal standard experience equal proportional losses.
The student’s result is correct.
(d) Calculate the original analyte amount at 50.0% recovery.
If the analyte recovery is only 50.0%:
\[ n_{A,\mathrm{initial}} =\frac{8.00}{0.500} \]
\[ \boxed{n_{A,\mathrm{initial}}=16.0\ \mathrm{\mu mol}} \]
The measured response ratio is still 0.900.
The detector sees the same recovered analyte and internal standard amounts as before, even though the original analyte concentration was different.
(e) Evaluate the student’s conclusion.
The student’s answer is not uniquely justified.
The experiment establishes that the recovered amounts of A and B are equal. It does not establish that their original amounts were equal.
The general relationship is:
\[ r=F\times \frac{ n_{A,\mathrm{initial}}\eta_A }{ n_{IS,\mathrm{initial}}\eta_{IS} } \]
Where:
- \(\eta_A\) is the fractional extraction recovery of the analyte.
- \(\eta_{IS}\) is the fractional extraction recovery of the internal standard.
- The other symbols have their previously defined meanings.
Rearranging:
\[ n_{A,\mathrm{initial}} = \frac{r}{F} n_{IS,\mathrm{initial}} \frac{\eta_{IS}}{\eta_A} \]
The student’s calculation assumes:
\[ \eta_A=\eta_{IS} \]
But this equality has not been established.
Two different original analyte amounts can therefore produce exactly the same detector response ratio.
To determine the original amount reliably, the relative extraction recovery must be validated. A suitable internal standard should behave sufficiently similarly to the analyte during the entire preparation process. A stable-isotope-labelled analogue can be particularly useful because it often has very similar chemical behaviour.
The deeper lesson is that an internal standard corrects shared experimental variation, not every source of chemical loss.
09Where to take this concept next
Internal standardisation gives us a way to compare an unknown compound against a known reference while reducing the influence of common experimental variation.
The next useful distinction is between an ordinary internal standard and an isotopically labelled version of the analyte.
In isotope dilution analysis, the labelled compound acts as a reference, and the measured isotope ratio can be used to determine the original analyte quantity. It is a more specialised application of the same underlying idea.
Understanding when the two compounds behave similarly, when they do not, and what the detector actually measures is what makes quantitative analytical chemistry reliable.