How to Explain Overshooting an Endpoint and Its Effect on the Answer

Imagine you pipette 25.00 mL of an unknown hydrochloric acid solution into a conical flask and titrate it with 0.1000 mol L⁻¹ sodium hydroxide.

Imagine you pipette 25.00 mL of an unknown hydrochloric acid solution into a conical flask and titrate it with 0.1000 mol L⁻¹ sodium hydroxide.

The first faint pink colour appears after 24.60 mL of NaOH has been delivered. You keep going, however, and finish at 25.40 mL. What happens to the calculated concentration of the hydrochloric acid?

It is overestimated. The extra 0.80 mL of NaOH was added after the acid had already been neutralised. It does not react with additional hydrochloric acid, but your calculation treats it as though it did.

This is the central rule for explaining endpoint overshooting: when the known solution is in the burette and the unknown solution is in the flask, an overshot endpoint produces a calculated concentration that is too high.

There is an important exception. If the unknown solution is in the burette instead, overshooting makes its calculated concentration too low. The difference comes from how the measured volume enters the calculation.

01What actually happens when you overshoot?

Follow the titration through the apparatus

A standard acid-base titration uses three main pieces of volumetric glassware.

  • A volumetric pipette transfers an accurately measured volume of the unknown solution into a conical flask.
  • A burette delivers a measured volume of the known solution, called the titrant.
  • The conical flask holds the reacting solutions and indicator.

The burette is clamped vertically above the flask. Its tap controls the flow of titrant through the narrow tip and into the flask. As the titrant enters, the flask is continuously swirled to mix the solutions.

The burette measures how much titrant leaves it. It does not directly measure how much titrant actually reacts.

To calculate the volume delivered, subtract the initial burette reading from the final reading:

\[ V_{\mathrm{titre}}=V_{\mathrm{final}}-V_{\mathrm{initial}} \]

For example, if the initial reading is 0.15 mL and the final reading is 24.75 mL:

\[ V_{\mathrm{titre}}=24.75-0.15 \]

\[ V_{\mathrm{titre}}=24.60\ \mathrm{mL} \]

Remember that burette readings increase downwards. You do not need to begin at exactly 0.00 mL.

Endpoint versus equivalence point

These two terms are related, but they do not mean exactly the same thing.

The equivalence point is the point at which the reactants have combined in the exact mole ratio given by the balanced chemical equation.

The endpoint is the experimentally observed point at which the indicator changes colour.

For a suitable indicator, its colour-change range is chosen to fall within the steep section of the titration curve. This makes the observed endpoint very close to the equivalence point.

Why does this distinction matter?

Imagine titrating hydrochloric acid with sodium hydroxide using phenolphthalein. The solution starts colourless. As NaOH is added, the pH increases, and the indicator eventually changes to pink.

The first faint pink colour that persists after swirling signals the endpoint.

If you continue adding NaOH, the solution becomes increasingly pink. The extra hydroxide ions remain in solution because the hydrochloric acid has already been consumed.

The important distinction is:

  • The equivalence point describes the chemical reaction.
  • The endpoint describes what you observe.
  • Overshooting means continuing to add titrant beyond the chosen endpoint.

An indicator endpoint may differ slightly from the exact equivalence point even when the titration is performed correctly. That small indicator-related difference is not the same as deliberately adding excess titrant after observing the endpoint.

02Why does overshooting change the calculated concentration?

Consider the reaction between hydrochloric acid and sodium hydroxide:

\[ \ce{HCl + NaOH -> NaCl + H2O} \]

The mole ratio is 1:1.

This means one mole of NaOH reacts with one mole of HCl.

The amount of sodium hydroxide delivered is calculated using:

\[ n=cV \]

where:

  • \(n\) is the amount of substance in mol
  • \(c\) is the concentration in \(\mathrm{mol\,L^{-1}}\)
  • \(V\) is the volume in L

At the equivalence point:

\[ n(\ce{HCl})=n(\ce{NaOH}) \]

Therefore, if the known NaOH is in the burette and the unknown HCl is in the flask:

\[ c(\ce{HCl})= \frac{c(\ce{NaOH})V(\ce{NaOH})} {V(\ce{HCl})} \]

Notice where the titre appears: in the numerator.

If you overshoot, the recorded titre increases. This makes the calculated number of moles of NaOH larger.

The calculation then assumes that more HCl must have been present to react with that extra NaOH. Consequently, the calculated HCl concentration increases.

The problem is that the additional NaOH never neutralised that extra acid. It was simply left in excess.

The direction-of-error rule

The effect depends on which solution is unknown.

Experimental arrangementEffect of overshootingCalculated unknown concentration
Known titrant in burette, unknown analyte in flaskRecorded titre too largeToo high
Unknown titrant in burette, known analyte in flaskRecorded titre too largeToo low

Why does the second situation reverse the result?

When the unknown solution is in the burette, the calculation uses the known amount of substance in the flask and divides it by the measured volume of unknown solution.

For a 1:1 reaction:

\[ c_{\mathrm{unknown}}= \frac{c_{\mathrm{known}}V_{\mathrm{known}}} {V_{\mathrm{titre}}} \]

Now the titre is in the denominator.

Increasing the measured volume decreases the calculated concentration.

This is why you should never write that overshooting always makes the calculated concentration too high. First identify which solution is unknown and where it is located.

Worked example: Calculate the error after overshooting

A student transfers 25.00 mL of hydrochloric acid into a conical flask. The flask is titrated with 0.1000 mol L⁻¹ NaOH.

The initial burette reading is 0.12 mL. The correct endpoint occurs at a final reading of 23.62 mL, but the student continues adding NaOH until the final reading reaches 24.22 mL.

Assuming a 1:1 reaction, calculate the correct hydrochloric acid concentration and the concentration the student would calculate using the overshot titre.

Step 1

\[ V_{\mathrm{titre}}=23.62-0.12 \]

\[ V_{\mathrm{titre}}=23.50\ \mathrm{mL} \]

Step 2

Convert the volume into litres:

\[ V=0.02350\ \mathrm{L} \]

Substitute into \(n=cV\):

\[ n(\ce{NaOH})=0.1000\times0.02350 \]

\[ n(\ce{NaOH})=0.002350\ \mathrm{mol} \]

Step 3

The balanced equation has a 1:1 mole ratio, so:

\[ n(\ce{HCl})=0.002350\ \mathrm{mol} \]

The volume of hydrochloric acid is 0.02500 L.

\[ c(\ce{HCl})=\frac{0.002350}{0.02500} \]

\[ \boxed{c(\ce{HCl})=0.09400\ \mathrm{mol\,L^{-1}}} \]

Step 4

\[ V_{\mathrm{titre}}=24.22-0.12 \]

\[ V_{\mathrm{titre}}=24.10\ \mathrm{mL} \]

The student has delivered an additional 0.60 mL of NaOH.

Step 5

\[ n(\ce{NaOH})=0.1000\times0.02410 \]

\[ n(\ce{NaOH})=0.002410\ \mathrm{mol} \]

Using the 1:1 mole ratio:

\[ c(\ce{HCl})=\frac{0.002410}{0.02500} \]

\[ \boxed{c(\ce{HCl})=0.09640\ \mathrm{mol\,L^{-1}}} \]

The student calculates 0.09640 mol L⁻¹ instead of the correct 0.09400 mol L⁻¹.

The result is approximately 2.55% too high.

The extra NaOH has not changed the original hydrochloric acid concentration. It has only caused the student to overestimate the amount of acid present.

Worked example: What if the unknown solution is in the burette?

A student places 25.00 mL of 0.1200 mol L⁻¹ sodium hydroxide into a conical flask. An unknown sulfuric acid solution is placed in the burette.

The reaction is:

\[ \ce{H2SO4 + 2NaOH -> Na2SO4 + 2H2O} \]

The correct endpoint requires 10.00 mL of sulfuric acid. However, the student delivers 10.50 mL before stopping.

Calculate the concentration the student would report using the overshot volume.

Step 1

\[ n(\ce{NaOH})=0.1200\times0.02500 \]

\[ n(\ce{NaOH})=0.003000\ \mathrm{mol} \]

Step 2

From the balanced equation:

\[ 1\ \mathrm{mol}\ \ce{H2SO4} : 2\ \mathrm{mol}\ \ce{NaOH} \]

Therefore:

\[ n(\ce{H2SO4})=\frac{0.003000}{2} \]

\[ n(\ce{H2SO4})=0.001500\ \mathrm{mol} \]

Step 3

The volume delivered is 10.50 mL, or 0.01050 L.

\[ c(\ce{H2SO4})= \frac{0.001500}{0.01050} \]

\[ \boxed{c(\ce{H2SO4})=0.1429\ \mathrm{mol\,L^{-1}}} \]

Step 4

The correct titre was 10.00 mL. Using this volume would give:

\[ c(\ce{H2SO4})= \frac{0.001500}{0.01000} \]

\[ c(\ce{H2SO4})=0.1500\ \mathrm{mol\,L^{-1}} \]

The student reports 0.1429 mol L⁻¹ instead of 0.1500 mol L⁻¹.

The calculated concentration is too low because the unknown acid is in the burette, so its measured volume is in the denominator.

The chemistry has not changed. Only the volume used in the calculation is incorrect.

03How apparatus conditioning affects the result

Endpoint overshooting is not the only way to obtain an incorrect titre. Poor preparation of volumetric glassware can produce similar errors.

Conditioning means rinsing an apparatus with the solution it will contain or deliver, after cleaning it with deionised water.

The purpose is to prevent residual water or another solution from changing the concentration of the measured sample.

Burette conditioning

The burette should be rinsed with deionised water and then with small portions of the titrant.

Why?

Imagine a burette containing a small amount of water before you add the sodium hydroxide solution. The remaining water dilutes some of the NaOH.

The actual concentration of the titrant is now lower than the stated concentration.

More solution must be delivered to reach the endpoint. However, if you use the original stated concentration in your calculation, you overestimate the number of moles of NaOH delivered.

For the usual arrangement, with unknown acid in the flask, this produces a calculated acid concentration that is too high.

Pipette conditioning

A volumetric pipette should also be rinsed with deionised water and then with small portions of the solution it will transfer.

Suppose a pipette still contains water when you fill it with an unknown hydrochloric acid solution.

Some of the measured volume is now water rather than hydrochloric acid. The number of moles of acid delivered into the flask is lower than expected.

Less NaOH is required to reach the endpoint, so the measured titre is smaller.

If you still calculate the concentration assuming the full marked volume of the pipette contained hydrochloric acid, you underestimate its concentration.

A standard volumetric pipette is calibrated to deliver its stated volume when allowed to drain correctly. Let it drain under gravity and touch its tip against the inside wall of the receiving flask. Do not blow out the remaining liquid unless the pipette is specifically designed as a blow-out type.

Conical flask conditioning

This is where students often make the opposite mistake.

A conical flask should be clean, but it does not need to be rinsed with the analyte solution before use.

A small amount of deionised water remaining inside the flask does not change the number of moles of analyte transferred by the pipette.

It changes the volume and concentration of the mixture, but not the amount of substance.

Therefore, provided the water is pure and does not react with the sample, it does not change the stoichiometric titre.

This is also why washing down the inside walls of the flask with deionised water during a titration is acceptable. It returns splashed reactants to the main solution without adding more moles of either reactant.

Quick comparison of apparatus errors

Assume the standard arrangement: known NaOH in the burette and unknown acid in the flask.

Experimental mistakeEffect on measured titreEffect on calculated acid concentration
Water left in buretteIncreasesToo high
Water left in pipetteDecreasesToo low
Deionised water left in flaskNo significant changeNo significant change
Air bubble in burette tip at the startApparent titre increasesToo high
Pipette not allowed to drain fullyDecreasesToo low
A calibrated pipette is incorrectly blown outIncreasesToo high

An air bubble in the burette tip deserves particular attention. If the tip initially contains air, some of the measured liquid must first fill the tip rather than enter the flask.

The burette reading still records that volume as delivered. Consequently, the recorded titre is larger than the amount that actually reached the reaction mixture.

The general lesson is to trace what happens to the measured moles, not simply memorise whether a particular mistake makes the answer higher or lower.

04How to avoid overshooting the endpoint

A good titration is not about pouring the solution as quickly as possible. It is about controlling the addition when the reaction approaches completion.

The first titration is usually a rough trial. Its purpose is to estimate the volume needed to reach the endpoint.

It is not normally included in the final average.

For subsequent accurate trials:

  1. Use a fresh, accurately pipetted aliquot of the analyte.
  2. Add titrant relatively quickly while the endpoint is still some distance away.
  3. As you approach the expected endpoint, slow down and add one drop at a time.
  4. Swirl the flask after each addition so the solutions mix completely.
  5. If the indicator colour begins to persist, use smaller additions.
  6. Stop at the first appropriate persistent colour change for the chosen indicator.
  7. Record the final burette reading immediately.

For phenolphthalein when titrating an acid with a strong base, the endpoint is the first faint pink colour that remains after swirling. A deep pink solution suggests that too much NaOH may have been added.

Do not wait for a strong colour change. You want to identify the endpoint, not prove that you can empty half the burette into the flask.

What should you do if you overshoot?

If you know that you have overshot an accurate trial, do not include it in your average of concordant titres.

Repeat the titration using a fresh aliquot and approach the endpoint more carefully.

You cannot simply subtract an estimated amount of excess titrant from the final burette reading. The actual equivalence point is not known precisely from the overshot result.

Also remember that concordant titres do not automatically prove accuracy.

If you consistently add too much titrant in every trial, your titres might agree closely with one another while all being systematically too large.

This is the difference between:

  • Precision: how closely repeated measurements agree.
  • Accuracy: how close a measurement is to the correct value.

A consistent technique helps with both, but repeating the same mistake does not remove it.

05A reliable method for explaining the direction of error

When an exam question asks you to explain how overshooting an endpoint affects a calculated concentration, use this sequence.

Step 1: Identify the unknown solution.

Is the unknown solution in the conical flask or the burette?

Step 2: Identify what the overshoot changes.

The recorded titre is larger than the volume required to reach the endpoint.

Step 3: Follow the calculation.

If the unknown is in the flask, the titre is generally used to calculate its moles. A larger titre gives a larger calculated amount and concentration.

If the unknown is in the burette, its concentration is calculated by dividing the known amount of substance by the larger recorded volume. The result is too low.

Step 4: Explain the chemistry.

State that the additional titrant is in excess because the analyte has already reached the endpoint. It is incorrectly counted as though it had reacted with additional analyte.

A complete exam response might read:

“Because the endpoint was overshot, the recorded titre is greater than the volume required for complete neutralisation. The calculation therefore overestimates the moles of titrant that reacted with the analyte. Since the known titrant is in the burette and the unknown acid is in the flask, the calculated acid concentration is too high.”

That explanation connects the experimental mistake to the mathematical error and then to the final answer.

06Questions and solutions

The following questions move from straightforward calculations to more complicated situations involving apparatus errors, indicator selection, and errors that interact across multiple titrations.

All numerical data are illustrative.

Question 1

A student transfers 20.00 mL of an unknown monoprotic acid into a conical flask. The solution is titrated with 0.1200 mol L⁻¹ NaOH.

The correct endpoint occurs at a titre of 16.40 mL. The student overshoots by 0.20 mL.

Calculate the concentration of the acid that the student would report if they used the overshot titre. State whether the result is too high or too low.

Solution 1

The student calculates an acid concentration of \(0.09960\ \mathrm{mol\,L^{-1}}\), which is too high.

The reaction has a 1:1 mole ratio, so the calculated amount of acid equals the amount of NaOH delivered.

Step 1: Calculate the overshot titre.

\[ V_{\mathrm{titre}}=16.40+0.20 \]

\[ V_{\mathrm{titre}}=16.60\ \mathrm{mL} \]

Convert to litres:

\[ V=0.01660\ \mathrm{L} \]

Step 2: Calculate the moles of NaOH.

\[ n(\ce{NaOH})=cV \]

\[ n(\ce{NaOH})=0.1200\times0.01660 \]

\[ n(\ce{NaOH})=0.001992\ \mathrm{mol} \]

Step 3: Calculate the acid concentration.

The mole ratio is 1:1, so:

\[ n(\ce{HA})=0.001992\ \mathrm{mol} \]

The acid volume is 0.02000 L.

\[ c(\ce{HA})=\frac{0.001992}{0.02000} \]

\[ \boxed{c(\ce{HA})=0.09960\ \mathrm{mol\,L^{-1}}} \]

Using the correct titre of 16.40 mL would give:

\[ c(\ce{HA})= \frac{0.1200\times0.01640}{0.02000} \]

\[ c(\ce{HA})=0.09840\ \mathrm{mol\,L^{-1}} \]

The reported result is too high because the student has counted the excess NaOH as though it reacted with additional acid.

Question 2

A student uses a 25.00 mL volumetric pipette to transfer an unknown hydrochloric acid solution into a conical flask.

The acid concentration is actually 0.0800 mol L⁻¹.

The pipette has not been rinsed with the acid after being washed with deionised water. When filled to the calibration mark, it contains 24.80 mL of hydrochloric acid and 0.20 mL of residual water.

The conical flask also contains 30.0 mL of deionised water.

The acid is titrated with 0.1000 mol L⁻¹ NaOH.

Calculate the expected titre and the concentration the student would calculate if they assumed that the full 25.00 mL pipette volume contained acid.

Explain why the water in the conical flask has a different effect from the water in the pipette.

Solution 2

The expected titre is 19.84 mL, and the student calculates an acid concentration of \(0.07936\ \mathrm{mol\,L^{-1}}\), which is too low.

The residual water inside the pipette reduces the amount of acid transferred. The water already in the conical flask does not change the number of moles transferred.

Step 1: Calculate the actual moles of hydrochloric acid delivered.

Only 24.80 mL of the pipette contents is acid.

Convert this to litres:

\[ V=0.02480\ \mathrm{L} \]

\[ n(\ce{HCl})=0.0800\times0.02480 \]

\[ n(\ce{HCl})=0.001984\ \mathrm{mol} \]

Step 2: Calculate the required moles of NaOH.

The reaction is 1:1:

\[ \ce{HCl + NaOH -> NaCl + H2O} \]

Therefore:

\[ n(\ce{NaOH})=0.001984\ \mathrm{mol} \]

Step 3: Calculate the expected titre.

\[ V=\frac{n}{c} \]

\[ V=\frac{0.001984}{0.1000} \]

\[ \boxed{V=0.01984\ \mathrm{L}=19.84\ \mathrm{mL}} \]

Step 4: Calculate the concentration using the student’s incorrect assumption.

The student assumes the entire 25.00 mL volume contains hydrochloric acid.

\[ c(\ce{HCl})= \frac{0.001984}{0.02500} \]

\[ \boxed{c(\ce{HCl})=0.07936\ \mathrm{mol\,L^{-1}}} \]

The result is too low.

Step 5: Explain the difference between the two sources of water.

The water inside the pipette mixes with the acid before delivery. It reduces the amount of acid contained in the measured aliquot.

The water already in the conical flask simply dilutes the acid after the pipette has delivered it. It does not change the number of moles of hydrochloric acid present.

Therefore, the pipette error affects the titre, while the additional water in the flask does not change the stoichiometric amount of NaOH required.

Question 3

A student titrates 25.00 mL of 0.1000 mol L⁻¹ sodium hydroxide with an unknown sulfuric acid solution in the burette.

The reaction is:

\[ \ce{H2SO4 + 2NaOH -> Na2SO4 + 2H2O} \]

The equivalence point occurs at 25.00 mL of sulfuric acid. The chosen indicator has a transition that produces its first persistent endpoint at 25.02 mL.

However, the student continues adding acid until 25.42 mL has been delivered.

Answer the following:

a) How much of the difference between the equivalence volume and the final recorded volume is due to the indicator endpoint, and how much is due to the student’s overshoot?

b) Calculate the sulfuric acid concentration using the final recorded volume.

c) Explain why it would be incorrect to claim that the indicator itself caused the entire error.

Solution 3

The indicator-related difference is 0.02 mL, while the avoidable overshoot is 0.40 mL. The calculated sulfuric acid concentration is \(0.09796\ \mathrm{mol\,L^{-1}}\), which is too low.

The indicator’s endpoint does not have to coincide perfectly with the equivalence point. The student’s additional 0.40 mL, however, was delivered after the correct observed endpoint.

Part a: Separate the two volume differences.

The difference between the equivalence point and the indicator endpoint is:

\[ 25.02-25.00=0.02\ \mathrm{mL} \]

This is the indicator-related difference.

The additional volume delivered after the indicator endpoint is:

\[ 25.42-25.02=0.40\ \mathrm{mL} \]

This is the student’s avoidable overshoot.

The total difference from the equivalence volume is:

\[ 0.02+0.40=0.42\ \mathrm{mL} \]

Part b: Calculate the sulfuric acid concentration.

First calculate the moles of sodium hydroxide in the flask:

\[ n(\ce{NaOH})=0.1000\times0.02500 \]

\[ n(\ce{NaOH})=0.002500\ \mathrm{mol} \]

From the balanced equation, two moles of NaOH react with one mole of sulfuric acid.

Therefore:

\[ n(\ce{H2SO4})=\frac{0.002500}{2} \]

\[ n(\ce{H2SO4})=0.001250\ \mathrm{mol} \]

Use the final recorded volume of 25.42 mL:

\[ V=0.02542\ \mathrm{L} \]

\[ c(\ce{H2SO4})= \frac{0.001250}{0.02542} \]

\[ \boxed{c(\ce{H2SO4})=0.04918\ \mathrm{mol\,L^{-1}}} \]

The correct concentration, using the equivalence volume, is:

\[ c(\ce{H2SO4})= \frac{0.001250}{0.02500} \]

\[ =0.05000\ \mathrm{mol\,L^{-1}} \]

The student’s calculated result is too low.

Part c: Explain the indicator’s role.

The indicator was selected so that its transition occurred close to the equivalence point. Its small 0.02 mL difference is a feature of the chosen indicator and the experimental conditions, not the student’s additional 0.40 mL.

The student should have stopped at the first persistent indicator endpoint.

Blaming the entire error on the indicator ignores the fact that the student continued adding titrant after the endpoint had already been observed.

Question 4

A student titrates 25.00 mL of an unknown monoprotic acid with 0.1000 mol L⁻¹ NaOH.

Three trials produce the following illustrative data:

TrialInitial reading (mL)Final reading (mL)Titre (mL)
10.1022.4422.34
20.1522.5322.38
30.2023.3223.12

The first two trials reach the intended faint-pink endpoint. In the third trial, the student continues adding NaOH and produces a much darker pink colour.

The laboratory’s specified concordance tolerance is 0.10 mL.

a) Identify the concordant titres.

b) Calculate the acid concentration using the appropriate titres.

c) Explain why averaging all three titres would give a misleading result.

d) Would repeating the same overshooting technique several times necessarily improve the accuracy of the experiment?

Solution 4

The concordant titres are 22.34 mL and 22.38 mL. Their average gives an acid concentration of \(0.08944\ \mathrm{mol\,L^{-1}}\).

The third titre should not be included because the student has identified a specific experimental mistake: overshooting the endpoint.

Part a: Identify the concordant titres.

Compare the first two:

\[ 22.38-22.34=0.04\ \mathrm{mL} \]

Their difference is less than the specified 0.10 mL tolerance.

Therefore, they are concordant.

The third titre is substantially larger and is known to have been overshot.

Part b: Calculate the average titre and concentration.

\[ V_{\mathrm{average}}= \frac{22.34+22.38}{2} \]

\[ V_{\mathrm{average}}=22.36\ \mathrm{mL} \]

Convert to litres:

\[ V=0.02236\ \mathrm{L} \]

The acid and NaOH react in a 1:1 mole ratio.

\[ n(\ce{NaOH})=0.1000\times0.02236 \]

\[ n(\ce{NaOH})=0.002236\ \mathrm{mol} \]

Therefore:

\[ c(\ce{HA})=\frac{0.002236}{0.02500} \]

\[ \boxed{c(\ce{HA})=0.08944\ \mathrm{mol\,L^{-1}}} \]

Part c: Explain why all three titres should not be averaged.

If all three values were averaged:

\[ V_{\mathrm{average}}= \frac{22.34+22.38+23.12}{3} \]

\[ V_{\mathrm{average}}=22.61\ \mathrm{mL} \]

This would produce a calculated concentration of approximately:

\[ c(\ce{HA})=0.09045\ \mathrm{mol\,L^{-1}} \]

This is higher than the result obtained using the concordant titres.

The third trial does not represent a properly identified endpoint. Including it would allow the excess NaOH to increase the average titre and therefore increase the calculated acid concentration.

Part d: Explain the effect of repeating the same mistake.

No. Repeating the same overshooting technique can produce very similar titres while maintaining a systematic error.

This would improve precision, because the results agree with one another, but not necessarily accuracy.

A good experiment needs both repeatable measurements and a consistent, correctly identified endpoint.

Question 5

A laboratory performs two titrations using the same sodium hydroxide solution.

In the first titration, the sodium hydroxide is standardised using 25.00 mL of 0.1000 mol L⁻¹ hydrochloric acid in a conical flask.

The sodium hydroxide is in the burette. The true equivalence volume is 25.00 mL, but the operator overshoots and records 25.50 mL.

The operator calculates the sodium hydroxide concentration using this recorded volume.

In the second titration, the same sodium hydroxide solution is used to determine the concentration of an unknown monoprotic acid.

A 25.00 mL aliquot of the unknown acid is placed in the flask. Its actual concentration is 0.08000 mol L⁻¹.

With correctly standardised sodium hydroxide, the true equivalence volume would be 20.00 mL. However, the operator again overshoots, this time recording a titre of 20.40 mL.

The operator uses the incorrectly standardised sodium hydroxide concentration from the first titration to calculate the unknown acid concentration.

Assume:

  • Both reactions have a 1:1 mole ratio.
  • The same sodium hydroxide solution is used in both titrations.
  • Its actual concentration does not change between experiments.
  • The only experimental errors are the two stated endpoint overshoots.
  • All other measurements are exact.

a) Calculate the sodium hydroxide concentration obtained from the first titration.

b) Calculate the unknown acid concentration obtained from the second titration.

c) Explain whether the final calculated concentration is too high, too low, or exactly correct.

d) Explain why obtaining the correct final answer does not demonstrate that either titration was performed accurately.

Solution 5

The final calculated unknown acid concentration is exactly \(0.08000\ \mathrm{mol\,L^{-1}}\), even though both titrations have been overshot.

This is an accidental cancellation of two errors. The first titration underestimates the sodium hydroxide concentration, while the second overestimates the volume of sodium hydroxide used.

Part a: Calculate the incorrectly standardised sodium hydroxide concentration.

The first reaction is:

\[ \ce{HCl + NaOH -> NaCl + H2O} \]

The hydrochloric acid concentration and volume are known.

Calculate its moles:

\[ n(\ce{HCl})=0.1000\times0.02500 \]

\[ n(\ce{HCl})=0.002500\ \mathrm{mol} \]

The reaction has a 1:1 mole ratio, so:

\[ n(\ce{NaOH})=0.002500\ \mathrm{mol} \]

However, the operator uses the overshot titre of 25.50 mL.

\[ V=0.02550\ \mathrm{L} \]

Therefore:

\[ c(\ce{NaOH})= \frac{0.002500}{0.02550} \]

\[ \boxed{c(\ce{NaOH})=0.09804\ \mathrm{mol\,L^{-1}}} \]

The actual concentration is 0.1000 mol L⁻¹, so the standardised concentration is underestimated.

This occurs because the known amount of hydrochloric acid is divided by an incorrectly large sodium hydroxide volume.

Part b: Calculate the unknown acid concentration.

The second titration uses the incorrectly calculated sodium hydroxide concentration:

\[ c(\ce{NaOH})=0.09804\ \mathrm{mol\,L^{-1}} \]

The operator records a titre of 20.40 mL:

\[ V=0.02040\ \mathrm{L} \]

Calculate the apparent moles of NaOH:

\[ n(\ce{NaOH})=0.0980392\times0.02040 \]

\[ n(\ce{NaOH})=0.002000\ \mathrm{mol} \]

The acid reacts with NaOH in a 1:1 mole ratio.

Therefore:

\[ n(\ce{HA})=0.002000\ \mathrm{mol} \]

The unknown acid volume is 0.02500 L.

\[ c(\ce{HA})=\frac{0.002000}{0.02500} \]

\[ \boxed{c(\ce{HA})=0.08000\ \mathrm{mol\,L^{-1}}} \]

Part c: Explain the final result.

The calculated concentration is exactly correct.

This happens because the two errors act in opposite directions.

During standardisation, the overshot titre makes the calculated sodium hydroxide concentration too low.

During the second titration, the overshot titre makes the calculated number of moles of sodium hydroxide too high.

The numerical errors cancel:

\[ \frac{25.00}{25.50}\times \frac{20.40}{20.00}=1.000 \]

The first factor represents the reduction in the calculated sodium hydroxide concentration. The second represents the increase in the recorded titre during the second titration.

Their product is exactly one.

Part d: Explain why the correct result does not prove experimental accuracy.

The final result is correct because two incorrect measurements happen to compensate for each other.

Neither titration has been performed accurately. The first produces an underestimated standard concentration, and the second produces an overestimated titre.

If either overshoot were slightly different, the errors would no longer cancel exactly.

This is why a result that agrees with an expected value does not automatically prove that the experimental method was reliable.

It also shows why you must assess each stage of a multi-step experiment independently. A correct final number can conceal significant experimental errors along the way.

07What to remember

When explaining an overshot titration endpoint, focus on the relationship between the measured volume, the mole calculation, and the unknown concentration.

  • If the known titrant is in the burette and the unknown analyte is in the flask, overshooting generally makes the calculated concentration too high.
  • If the unknown titrant is in the burette and the known analyte is in the flask, overshooting makes the calculated concentration too low.
  • Water left in a burette dilutes the titrant, while water left in a pipette reduces the amount of analyte transferred.
  • Pure deionised water in a conical flask changes the dilution, not the number of moles.
  • Concordant titres demonstrate precision, not necessarily accuracy.

The next useful step is understanding how these errors affect standardisation and back titration calculations. In both, you must track which quantities are measured, which are known, and which are being inferred at each stage of the experiment.